JSM -5310 使用方法

JSM -5310 使用方法. 精密分析室. 2010/11/25. 目次. 始 動. 試料交換. 観察. 写真撮影. 装置停止. 終了. 始動. 1.配電盤のスイッチを ON 通常ONになっています 2 .冷却水を流す 3.POWERキーを START まで 回す 4.LCDの排気シーケンス が   HT  REDYになるまで 20分 まつ. 目次に戻る. 試料交換. 1.フィラメントを反時計方向一杯に回す 2.HTをOFF 3.VACUM VENTをON 4.鏡筒内が大気圧に なった ら ステージを引き出す

albany
Download Presentation

JSM -5310 使用方法

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. JSM-5310使用方法 精密分析室 2010/11/25

  2. 目次 始動 試料交換 観察 写真撮影 装置停止 終了

  3. 始動 1.配電盤のスイッチをON 通常ONになっています 2.冷却水を流す 3.POWERキーをSTART まで回す 4.LCDの排気シーケンスが   HT REDYになるまで20分 まつ 目次に戻る

  4. 試料交換 1.フィラメントを反時計方向一杯に回す 2.HTをOFF 3.VACUM VENTをON 4.鏡筒内が大気圧になった らステージを引き出す 5.試料交換し試料室を押し込む 6.VACUM EVACをON 7.LCDにHT READYが表示されれば観察可能 目次に戻る

  5. 観察 1.GUN BIASはAUTO 2.IMAGE SEI(ランプ消灯) 3.加速電圧を設定し、HTをON 4.SCANNING MODE LSP1 5.フィラメントをゆっくり加熱する 目次に戻る

  6. 写真撮影 PHOTOを押し撮影するとデジカメに自動的に露出される 画像は上下が幾分カットされます 撮影データはSDカードをコピーするか、PCに取り込んで使用する 目次に戻る

  7. 装置停止 FILAMENT  反時計方向いっぱいに回す HT OFF POWERキー OFF 配電盤スイッチ OFF 通常切らなくても良い 10~15分後に冷却水を止める 使用後は検鏡記録をつけてください 目次に戻る 2010.5.18