Low pressure system

Low pressure system. Presented by : Dr Rashmi Moderator: Dr Gian Chauhan. Starts downstream of flow control devices Pressures: Slightly above atmospheric pressure (Max upto 6-8psig) & variable Components: Flowmeters Hypoxia prevention safety devices Unidirectional valves

apria
Download Presentation

Low pressure system

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Low pressure system Presented by : Dr Rashmi Moderator: Dr GianChauhan

  2. Starts downstream of flow control devices • Pressures: Slightly above atmospheric pressure (Max upto 6-8psig) & variable • Components: • Flowmeters • Hypoxia prevention safety devices • Unidirectional valves • Pressure relief devices • Common gas outlet • Vaporizers & Mounting devices including backbar

  3. Flowmeters: (aka flow indicators, flow tubes, rotameters) • Types: Mechanical & Eletronic • Mechanical flowmeters: Based on principle that flow past a resistance is proportional to pressure • Variable orifice type tubes (Thorpe tube), internally tapered narrowest at bottom • Indicator: free to move inside the tube • Gas passes through an annular opening b/w the indicator and the tube towards the outlet • Wt. of bobbin balances the upward force by the gas pressure

  4. Rate of flow depends on: • Pressure drop across the constriction (frictional resistance caused to the gas, equals wt. of float/ area), remains constant throughout (Constant pressure flowmeters) • Size of the annular opening • Physical properties of the gas • (for Low flows (Laminar flow) flow is dependent on viscosity; Hagen Poiseuille equation (Tubular orifice) • (High flows(Turbulent flow) flow is dependent on density; Graham’s law (orificial orifice)

  5. Calibrated at atmospheric pressure (760 mmHg) & room temperature (20 Degree) • If pressure decrease as in high altitude, more gas will flow that what the indicator specified and in hyperbaric chamber, flow will be less than what is indicated. • Temperature may alter viscosity and density of gas but effect on gas flows is less significant.

  6. Flowmeter Assembly: • Tube , Scale , Indicator, Stop at the top • Lights • Coding (color, touch, name) • Plastic shield (for protection)

  7. Tubes: • Glass tubes • Single taper (different tubes for low and high flows) • Double taper (single tube with gradations for fine and coarse flows)

  8. Stop (at the top): • Prevents bobbin from plugging the outlet (Damage, hidden bobbin) • If falls into the tube on the bobbin will increase the wt. & so show falsely low reading • Scale: • Flow indicator (marked on or immediately adjacent to tube) • L/ min (lower than 1 L as ml/min or Decimals of L) • Lights/ fluorescent background: helps in darkened rooms

  9. Flowmeter Tube Arrangement: • Tubes for various gases arranged side by side, meet at the top into a common manifold (mixing chamber) • Arranged in series or tandem • One flow control valve for one gas (single/ double tube) • Normal direction of flow: bottom to top, then left to right

  10. Flowmeter arrangement

  11. Auxiliary Oxygen Flowmeters: • Self contained flowmeter with its own flow control valve, flow indicator, outlet • Short tube, max of 10 L/min flow • Usually mounted on the left side of the machine • Can be used to supply oxygen even without turning the machine on

  12. Problems with Flowmeters: • Inaccuracy: • Inseparable unit (mixing of components not allowed) • Calibrated for a particular gas (other gas cannot be used) • Greatest accuracy in the middle half of the tube (single taper provides better accuracy) • Indicator problems: (worn, distorted, sudden burst of gas pushing bobbin to top, stop may dislodge- rest on bobbin)

  13. Leaks: • leak upstream of the bobbin will result in low FGF but will not be indicated by the bobbin • If cylinder/ yoke plug not in place, open flow control valve will l/t leakage • Using wrong flowmeter: • Changing the position of the gases (arrangement) l/t errors as anaesthetists are used to a particular arrangement • Usually with Nitrous and air

  14. Electronic Flowmeters: • Use electronic flow sensors and conventional flow control valves • Slow measured is represented digitally or by a simulated flowmeter • Clockwise movement opens the gas flow • Advantage: Information is available in a form which can be directly sent to data management system

  15. Hypoxia Prevention Safety Devices Mandatory Minimum Oxygen Flow: (50 to 250 mL/minute), preset by the manufacturer (altered on customer specification), activated when master switch is turned on • Minimum Oxygen Ratio: Prevents delivery of mixture of oxygen and nitrous oxide having an oxygen concentration below 21% oxygen • Mechanical Linkage: 14-tooth sprocket on the nitrous oxide flow control valve and a 29-tooth sprocket on the oxygen flow control valve (maintain a min. of 25% oxygen), if we increase nitrous beyond this it engages the Oxygen valve. • aka ‘Link 25’ system, drawback: links only 2 gases

  16. Mechanical linkage

  17. Electronic linkage: electronic proportioning valve controls the oxygen concentration in the fresh gas, supply a minimum of 25% oxygen • Alarms: To alert the operator that % of Oxygen has fallen in the gas mixture

  18. Vaporizer Mounting Systems: (back bar) Permanent mounting (Cage mount): • tools are required to remove or install a vaporizer on the anesthesia machine, back bar is a component of vaporizer • Adv: Less physical damage to vaporizers and fewer leaks • Disadv: May not have enough mounting locations to accommodate all of the vaporizers, malfunctioning vaporizer cannot easily be exchanged • Penlon “off-line” system: A modified cage mount arrangement, Flexible back bar hoses with connections which can be separated and connected to a vaporizer (which will then be included in the backbar)

  19. Penlon “Back entry” system: • Vaporizer is attached to the back bar by a fixing bolt protruding through the middle of the rear panel of the machine • Protruding pegs and seals on the back bar automatically locate and seal the inlet and outlet connections to vaporizer • Fixing bolt is tightened using a cylinder key • 3 entry systems can be mounted in series • Each system has a cover (blanking plate),should be fitted properly when a vaporizer is not connected (or else leak)

  20. Detachable mounting: Are standard on most new anesthesia machines, allow the vaporizer to be mounted & removed without use of tools. • Select-a-tecsystem: A pair of port valves for each vaporizer, Vaporizer has special mounting brackets with 2 plungers (spindles) which fit over the port valves. Seal created by wt. of vaporizer and O-ring around the post valve. Locking lever on the back of vaporizer.

  21. Before mounting the vaporizer, control dial should be in off position • Vaporizer is fitted onto the mounting system and locked in position • When the vaporizer is turned on, the two plungers move downward, opening the valve ports and connecting the vaporizer into the fresh gas stream • When the vaporizer is turned off, the gases may still flow through the head of the vaporizer (even when not locked properly)

  22. Drager Mounting System: • Vapor 2000 vaporizer must be in the “T” (travel) position before it can be unlocked from the machine • Teats to be performed before operation of vaporizers: • Looking at the level of each vaporizer (at same level) • Attempt to lift each off the manifold without unlocking it (shouldn’t be able to) • It should be possible to turn ON only one vaporizer at a time • Perform leak tests

  23. Advantages of detachable mounting: • Machine can have fewer mounting locations, allowing a more compact machine • Vaporizers can be easily removed and replaced, even during a case • If malignant hyperthermia is a potential problem, the vaporizers can be removed: better results than if the vaporizers remain in the off position

  24. Disadvantages: • Partial or complete obstruction to gas flow from problems with the mounting system • Leaks may occur • an absent or damaged O-ring • leaving the locking lever in the unlocked position • something is pushed under the vaporizer enough to lift it slightly off of the O-ring • Differences among vaporizers and interlocks from different manufacturers can pose problems of compatibility

  25. Interlocking Devices: • Prevent more than one vaporizer from being turned on at a time • Checking the interlock device should be done • For Datex-Ohmeda vaporizers, operating the dial release activates two extension rods that prevent operation of any other vaporizer installed on the manifold • Older versions of the Select-a-tec system: provide mounting positions for three vaporizers, if only two vaporizers are fitted, the center position must be occupied. If the center position is not occupied, the interlock is ineffective • Newer ones: interlocking irrespective of center position occupancy

  26. Unidirectional valves (check valves): • Backpressure from: • Breathing system during IPPV • Oxygen flush • Partial switching from open to closed circuit • Can affect flowmeter readings and conc. of volatile agents • Located b/w the vaporizer and common gas outlet, upstream of where the oxygen flush joins the fresh gas flow • Disadvantage: while checking for leaks we cannot detect a leak upstream of the check valve

  27. Pressure relief device: • Near common gas outlet to protect the machine from high pressures • Opens to atmosphere • Disadv: May limit the ability of anaesthesia machine to provide adequate pressure for jet ventilation (separate ports recommended for jet ventilation) • Low pressure piping: • Has large number of connections so prone for breakages and leaks

  28. Common Gas Outlet: • Delivers the gas mixture to the breathing system • Common site for disconnections so machine standards mandate that it should be difficult to accidentally disengage the system • Have a 15mm female slip joint fitting (accepts a tracheal tube connector) and a co-axial 22mm male connector • Fresh gas supply tube of the breathing system connects to this outlet

  29. Some machines may have 2 such outlets (may l/t confusion, gas may be directed to the wrong outlet) According to standards there should be only one CGO. • Some may not have even one (have internal breathing system connections) • This outlet should not be used to administer supplemental oxygen (delays use of circuit if emergency arises, also if accidentally vaporizer remains on inhalational agent will be delivered unknowingly)- prefer auxiliary Oxygen flowmeter

  30. Tests for Low pressure system: • Check vaporizer: Level of agent, cap to be tight, should not be lifted off. • Leak test: performed without the basal flow so machine master switch to be turned off. To be performed with vaporizer off first and then on. • Positive pressure test • Oxygen flush test • Pressure gauge test • Negative pressure test

  31. thankyou