An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Use of Pedicle Screws for VEPTR Fixation to the Spine Richard E. McCarthy, M.D. Professor, Departments of Orthopaedics and Neurosurgery University of Arkansas for Medical Sciences Arkansas Children’s Hospital Frances L. McCullough, MNSc, RNP Arkansas Children’s Hospital

  2. Introduction • The original design of the VEPTR used supralaminar hooks as anchors to the spine. • One difficulty with rib to spine VEPTR has been complications related to the distal anchor hook to the spine (loss of fixation, migration) • A technique of using reinforced pedicle screws in place of the hook as fixation of the spine has been developed in an attempt to eliminate this occurrence.

  3. Methods • Charts and radiographs were reviewed on all patients who had placement of rib to spine VEPTR devices utilizing screw fixation to the spine. • The technique for the pedicle screw fixation was done with pedicle screws reinforced with fiberwire around the spinous process or lamina or with a supplemental hook above.

  4. Results • Eight patients • Diagnoses: • Congenital-4 • Neuromuscular-1 • Syndromic-2 • Cyanotic heart disease-1 • Average age at placement: 6+5 years • (range 1+6 to 17+10 years) • Average length of follow-up: 2+7 years • (range 6 weeks to 5+8 years) • Average number of lengthenings per patient: 4 • (range 0-12)

  5. Complications • Asymptomatic migration of the screw distally which was replaced at the next scheduled lengthening- 1 patient • Infection-2 patients • Upper anchor pullout- 1 patient

  6. Conclusion • Pedicle screw reinforcement offers safe and secure distal fixation at two year follow-up for rib to spine VEPTR placement.