Chapter 18 Electrosurgery Generators

Chapter 18 Electrosurgery Generators. 徐國恩 2014/9/15. 18-3 Electrosurgery Machines. RF generator:300 ~ 3000 kHz Active electrode: 截 面積很小( a few mm 2 ), 形成一 probe 供外科醫生操作 Passive electrode(patient plate): 大面積(>100 cm 2) ), 放置在病患臀部或大腿下. Operating principle: 相同的電流流經不同截面積的電極,造成電流密度不同而產生不同的加熱效果

torie
Download Presentation

Chapter 18 Electrosurgery Generators

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Chapter 18Electrosurgery Generators 徐國恩 2014/9/15

  2. 18-3 Electrosurgery Machines • RF generator:300 ~ 3000 kHz • Active electrode:截面積很小(a few mm2),形成一probe供外科醫生操作 • Passive electrode(patient plate):大面積(>100 cm2)),放置在病患臀部或大腿下

  3. Operating principle:相同的電流流經不同截面積的電極,造成電流密度不同而產生不同的加熱效果 • Application:切除不良組織以及燒結出血的血管

  4. 18-4 Electrosurgery Circuits • Spard Gap Machine • T1:昇壓變壓器,使之足以使空氣離子化,一般為2000 ~ 3000 V • C1/L1:Tank circuit • L1/L2:Coupler • RFC1,RFC2:To prevent RF energy from getting into the power supply.

  5. Cut Coagulation • 一般不加 60 Hz filter 的 cut wave

  6. 18-4-1 Solid-state electrosurgery generator circuit

  7. 18-5 Electrosurgery Safety • Inappropriate points on the body:patient plate 放置處遇到凸骨、凹痕、皺紋,或是 patient plate 本身的彎曲不規則,都會產生「熱點」而引起燒傷 • Inadvertent ground path :另外在病患身上形成接地迴路,使得電流不經由 patient plate 流出,造成燒傷 • Microscopic holes in the surgeon’s gloves • Lipuids spilled into the machine 18-3

  8. 18-6 Testing Electrosurgery Units 使用thermocouple RF ammeter: • 優點: • 本身是一 rms-reading meter,可直接反應出 output level • 不需要像 peak-reading meter 要針對輸出波形作校正 • 缺點: • 較其他儀表昂貴 • 其輸出是非線性,小功率部份判讀失準,所以甚至需要不同範圍的兩個thermocouple ammeter

  9. 使用 peak-reading meter: • Voltage divider

  10. Toroid current transformer • 普遍應用在低功率電療儀器的測試,如微燒結及眼部或內視鏡電療

  11. Chapter 19Care and Feeding of Battery-Operated Medical Equipment

  12. 19-3 Introduction Batteries are used for : • Protability • Defibrillator • Patient monitors • ECG & blood pressure • Patient safety • Cardiac output computer

  13. 19-4 Cell or Batteries? • The cell is the most basic element in a battery and sets the minimum voltage for that sort of device • Additional voltage is gained by connecting the cells in series, and extra current is available by connecting them in parallel • We would refer to multiple-cell entities as batteries • But in common usage, all cells and batteries are called batterise

  14. 19-5 NiCd Cells and Batteries • Most commonly used • 1.2 V terminal voltage • Sustain a charge-discharge cycle life of 1000 times before becoming unusable(the battery drops below 80% of its oribinal specified value

  15. 19-6 Battery Capacity • The capacity of a battery is measured in ampere-hours (A-H) (i.e., the product of current load , in amperes, and the time required to reach the designated discharge state) • NiCd batteries can deliver short-duration currents of 50A or more, that is why they are used in medical defibrillators and some medium-powered portable radio transmitters

  16. 19-7 Battery-charging Protocols • The general charging rule is:Charge at 1/10 ampere-hour rating for 14 hours • Fast charging should not be done unless the battery maker recommends it and be a little cautious

  17. Problems : • Sitting:The solution for this type of problem is a trickle charge at a rate between A-H/30 and A-H/50 • Temperature effect

  18. 19-8 NiCd Battery Memory • Memory means that a battery will not allow deep discharge after repeated shallow discharges (premature failure) • A NiCd battery with memory problems can some times be reformed by repeatedly fully charging it and them immediately deep discharging it

  19. 19-9 Battery Maintenance • For most equipment the manufacturer recommends that the batteries be periodically discharged and then recharged: • 1.Fully charge the battery or cell • 2.Discharge it fully with a current of A-H/10 for 8~9 hours for multicell batteries and 10 hours for single cells (because of polarity reversal) • 3.Recharge the battery at the A-H/10 rate for 14 to 16 hours

  20. Do not leave the battery in a discharged condition for a long period, it may develop interelement shorts • Revitalization • Revitalized batteries and cells should not be regarded as reliable and should only be used for shout terms, under emergency conditions. Good engineering practice requires replacement with a new cell rather than salvaging the defective cell

  21. 19-10 Charging NiCd Batteries There are two basic forms of charger for NiCd batteries: • Constant current (CI) • The transformer secondary voltage hsould be 2.5 times (or more) the battery voltage • A resistor limits the output current under short-circuit conditions to A-H/10 charging rate

  22. Electronic constant current charger based on three-terminal voltage regulator

  23. Constant voltage (CV)

  24. 19-12 Other Batteries 19-12-1 Lead-acid batteries • 優點:reliable and easily available • 缺點:very heavy and dangerous • 用途:high-power portable applications (automobile battery is often preferred) • 保養: • Water level in each of the cells must be checked periodically (weekly) • Clean the cap or replace it • Warning:Lead-acid batteries produce hydrogen as a normal byproduct

  25. 19-12-2 Carbon-zinc and alkaline dry cells • 一般的用途(閃光燈、收音機、……) • 缺點: • 不可充電,用完就丟 • 端電壓在使用時下降很快

  26. 19-12-3 Mercury dry cells • 優點:在電池壽命結束前,端電壓幾乎維持定值 • 用途:儀器校正用 • 保養:如同NiCd batteries、carbon-zinc, alkaline,水銀電池在不使用的時候最好放在冷的環境中(ex.冰箱),增加電池存放時間

  27. 19-12-4 Gel-cell batteries

  28. 19-12-5 Lithium cells • 優點:壽命長 • 用途:電腦、手錶、數位儲存記憶單元 • Warning: • 注意過充或過熱的問題 • 不可攜帶到大氣壓力較小之處,如國際航線班機