E N D
1. External Quality Assessment (EQA)for AFB Microscopy
2. Area 2
a microscope table, (if there is no electricity, this table should
be placed directly in front of a window) Area 2
a microscope table, (if there is no electricity, this table should
be placed directly in front of a window)
3. Labeling of Specimens
It is critical to label a specimen properly to avoid confusion
and delay.
A request form must accompany each specimen. The
information on the form must exactly match the information
on the container. Always label the container on the side, never
on the lid.
Using an indelible marking pen, write the name of the TB
suspect and the date of collection. Labeling of Specimens
It is critical to label a specimen properly to avoid confusion
and delay.
A request form must accompany each specimen. The
information on the form must exactly match the information
on the container. Always label the container on the side, never
on the lid.
Using an indelible marking pen, write the name of the TB
suspect and the date of collection.
4. Washing Hands
Always wash hands with soap and water after
handling specimens and containers. Washing Hands
Always wash hands with soap and water after
handling specimens and containers.
5. The Laboratory Register
The WHO or IUATLD laboratory register is strongly recommended as a guideline. The format of the register should
never be altered.
The Laboratory Register
The WHO or IUATLD laboratory register is strongly recommended as a guideline. The format of the register should
never be altered.
6. If the slide is not frosted, engrave the laboratory serial number
using a diamond pointed stylus.
Never use a grease pencil, since markings could wash off
during the staining process.
If the slide is not frosted, engrave the laboratory serial number
using a diamond pointed stylus.
Never use a grease pencil, since markings could wash off
during the staining process.
7. Using a continuous rotation, cover an oval area approximately
2 cm in length on the slide. Put only one smear on each slide. Using a continuous rotation, cover an oval area approximately
2 cm in length on the slide. Put only one smear on each slide.
8. Gently wash the stain from each slide with a stream of cold
water until all free stain has washed away. Always rinse gently
so the smear is not washed off of the slide.
Individually tilt each slide to drain off any excess rinse water.
This prevents puddles remaining on the slide which can dilute
the next reagent.
Gently wash the stain from each slide with a stream of cold
water until all free stain has washed away. Always rinse gently
so the smear is not washed off of the slide.
Individually tilt each slide to drain off any excess rinse water.
This prevents puddles remaining on the slide which can dilute
the next reagent.
9. Arrange the following supplies next to the microscope: a
bottle of immersion oil, organic solvent, lens paper, laboratory
register and a slide box for storage.
Use the 40X objective to focus and determine a suitable
reading area of the slide.
Place a drop of immersion oil on the stained
smear. Let the drop fall freely onto the slide.
Never touch the slide with the oil applicator. This
could lead to carry over of AFB in the immersion oil to
the next slide.
Turn the nosepiece to bring the 100X objective into place.
Now, gently lower the 100X objective.
It should barely touch the oil. Never allow the lens to touch
the slide. This can damage the lens and possibly break the
slide.
While looking through the eye-piece, adjust the immersion
lens slowly and focus until the image on the smear appears.
To fine focus, turn the fine adjustment knob carefully.
Arrange the following supplies next to the microscope: a
bottle of immersion oil, organic solvent, lens paper, laboratory
register and a slide box for storage.
Use the 40X objective to focus and determine a suitable
reading area of the slide.
Place a drop of immersion oil on the stained
smear. Let the drop fall freely onto the slide.
Never touch the slide with the oil applicator. This
could lead to carry over of AFB in the immersion oil to
the next slide.
Turn the nosepiece to bring the 100X objective into place.
Now, gently lower the 100X objective.
It should barely touch the oil. Never allow the lens to touch
the slide. This can damage the lens and possibly break the
slide.
While looking through the eye-piece, adjust the immersion
lens slowly and focus until the image on the smear appears.
To fine focus, turn the fine adjustment knob carefully.
11. External Quality Assessment Early warning-system for problems
Measure of laboratory quality
Valuable benchmarking tool (standardization and traceability)
Indicator of where to direct improvement efforts
Monitor of changes in technology and testing practices (evaluation component)
13. Take Home Message The primary responsibility for National TB Reference Laboratories is supporting AFB microscopy networks through implementation of EQA and training
14. International GuidelinesEQA – AFB Microscopy External Quality Assessment (EQA) Components
onsite evaluation
rechecking
proficiency testing
15. Key Features of EQA guidelines Resource analyses and phased approach for EQA
Identify laboratories with performance problems---not intended to identify individual errors or validate patient diagnosis
Emphasizes the need for onsite evaluation of laboratories using standard indicators
Recognizes panel/proficiency testing as a method of EQA
16. Key Features of EQA Guidelines Rechecking Emphasis on blinded random sample
Emphasis on the smallest possible sample size that provides some information on test performance (sensitivity)
Provides information to measure performance---based on the number and types of errors
17. EQA -- AFB MicroscopyOn-site Evaluation -- Background Many countries lack the resources for annual visits of peripheral laboratory by central laboratory staff
Laboratories in most countries are visited by a non-laboratory District supervisor
Optimum evaluation is performed by trained laboratory staff in supervisory role
18. EQA – AFB Microscopy Guidelines Key FeaturesOn-site Evaluation Develop a standard checklist of questions and indicators
Include minimal evaluation that can be performed by non-laboratory trained personnel (e.g., inventory supplies, reagents, equipment)
Include detailed evaluation that can be performed by supervisory laboratory staff
Train laboratory and non laboratory staff to assure consistent application
19. On-site Evaluation Results - Uganda
20. On-site Evaluation Performed by Non-laboratory Staff Examples: Are all staining reagents available and within expiration dates?
How are wire loops cleaned?
Is the laboratory register present and all columns completed properly?
How is maintenance on the microscope performed?
21. Quality Control in the TB Laboratory
Quality Control (QC) should be routinely
performed in the laboratory to ensure
reliability as well as reproducibility of results.
Perform QC on each new batch of stain. Observe these results before reading patient slides. This important step will verify correct staining procedure and microscope function.
When controls are not demonstrating proper staining,
determine the cause and correct it.
If the negative control appears red after the staining procedure, this is the result of incomplete decolorization. Quality Control in the TB Laboratory
Quality Control (QC) should be routinely
performed in the laboratory to ensure
reliability as well as reproducibility of results.
Perform QC on each new batch of stain. Observe these results before reading patient slides. This important step will verify correct staining procedure and microscope function.
When controls are not demonstrating proper staining,
determine the cause and correct it.
If the negative control appears red after the staining procedure, this is the result of incomplete decolorization.
22. On-site Evaluation Performed by Laboratory Staff Examples: Does the technician verify that the container is properly labeled?
How are slides labeled?
How often is the carbol fuchsin filtered?
How many fields are examined to report a negative smear?
23. EQA - AFB MicroscopyProficiency Testing - Background Uncommon in resource-limited countries
Prepared smears (South Africa) or patient slides (Senegal) sent from central laboratory
Consistent challenge of laboratory test performance
PT test performance may be different from testing routine patient specimens
24. EQA – AFB Microscopy Guidelines Key FeaturesProficiency Testing Laboratory may re-use patient slides but a procedure is provided to produce test slides for consistent slide sets
Recommended slide set is 10 slides: 5 stained and 5 unstained
Simple forms for slide production and collection of test results
25. PT Implementation in Mexico Inspected 587 of 637 laboratories
604 microscopists given a 2 hour, 10 slide test
54% had score >80
33% had score 60-79
13% had score <60
536/604 (88.7%) finished all 10 slides
216 persons with score <80 received training followed by second PT: average scores improved from 65 to 90 (P-value < 0.0001).
26. EQA - AFB MicroscopyRechecking - Background Recommended by IUATLD and WHO
Usually 100% of positive and 10% of negative smears
Usually un-blinded — adds bias*
Reviews patient testing; including smear preparation, staining, and interpretation
*Lan N.T.N. et al, 1999 Int J Tuberc Lung Dis 3(1): 55-61
27. EQA – AFB Microscopy Guidelines Key FeaturesRechecking Emphasizes “blinding” and random sample using the laboratory register
Sample size is based on Lot Quality Assurance Sampling (LQAS) with parameters selected for test volume and desired sensitivity
Positives and negatives sampled
Minor errors (FP or FN with 1-9 AFB/ 100 f) are included as a measure of performance
28. Comparison of random blinded re-checking to non-blinded rechecking of AFB Smears in Mexico
29. Comparison of random blinded re-checking to non-blinded rechecking of AFB Smears in Mexico
30. Sample Size Comparison of the old method* and new EQA guidelines
31. EQA - AFB Microscopy Guidelines Key FeaturesResource Analysis Inventory available resources (actual/projected)
Manpower, supplies, communication, administrative, financial
Examine effectiveness of current EQA activities
Gather laboratory service information
Planning—options for the evolution of EQA
Pilot test and document changes
Expansion based on availability of resources
32. Resources to implement on-site evaluation The NTP will need to consider: Capacity of the reference laboratory staff to provide on-site evaluation of all intermediate laboratories at least annually.
Capacity of intermediate laboratories to provide on-site inspection of the peripheral laboratories at least annually, and more frequently to correct problems identified.
Availability of properly trained non-laboratory personnel to make supervisory visits at least quarterly (as required for DOTS).
Capacity to implement necessary QI measures.
33. Resources to implement panel testingThe NTP will need to consider: Available financial support.
Proficiency of reference laboratory staff to perform ZN AFB smear microscopy.
Capacity of the reference laboratory staff to prepare panel testing slide sets for the laboratories to be evaluated
Available mechanisms to deliver slides to the peripheral sites, including mail and couriers.
Capacity of the reference laboratory staff to review and evaluate results from peripheral laboratories, and provide recommendations and follow-up for corrective action.
34. Resources to implement blinded rechecking The NTP will need to consider: Available financial support.
Capacity of peripheral laboratories to store smears for rechecking.
Availability of properly trained personnel to collect appropriate samples of slides from peripheral sites.
Capacity of the reference laboratory staff at central and intermediate level laboratories to reread smears from peripheral sites, including second rereading to resolve discrepancies as needed.
Capacity of reference laboratories to provide results of rechecking as well as feedback to implement effective corrective action.
35. EQA – AFB Microscopy GuidelinesResource AnalysisPhased Approach Assure the five elements of DOTS
Develop a central reference and intermediate laboratories to carry out EQA
Determine the existing capacity for EQA
Train district health officials to evaluate the minimal functions of microscopy laboratories
36. EQA – AFB Microscopy GuidelinesResource AnalysisPhased Approach (cont) Proficiency testing to evaluate performance
Pilot rechecking program
Determine resources: additional PT or phased implementation of rechecking
38. The Keys to Successful Quality Assurance Include: Adequately trained, interested, and committed staff
Common sense use of practical procedures
A willingness to admit and rectify mistakes
Effective communication
Ref: Laboratory Services in Tuberculosis Control, WHO