350 likes | 732 Views
Auger Surface Detector . . . . . . . . . . . . . . . . Plastic Tank. . 1.2 m. Tyvek Liner. Purified Water. . 3.2 m. Muons. PMT. . . Cherenkov Light. Spectrum Response. Cherenkov photons are peaked at 350~450 nm after reflection and absorption in water tank.It is matched to QE of Bi-Alkali photo-cathode..
E N D
1. PMT testing at UCLA-STRB
3. Spectrum Response
4. Signal from 6x1019eV Iron Shower
5. Correlation between Energy, Arrival Time and Density (1019eV, 900m from the core)
6. Specification of 8-9 PMTs for Auger
7. Three Candidate PMTs for Auger
8. PMT Test Facility at UCLA
9. Single Photoelectron Peak
11. Quantum Efficiency
12. Gain & Dark Current of Hamamatsu
13. Gain & Dark Current of Photonis
14. Gain & Dark Current of ETL
15. UCLA Setup for Linearity Measurement
16. Linearity - Hamamatsu
17. Linearity - Photonis
18. Linearity - ETL
19. Dark Pulse Rate - Hamamatsu
20. Dark Pulse Rate - Photonis
21. Dark Pulse Rate - ETL
22. After Pulse
23. After Pulse
24. Earth Magnetic Field
25. Earth Magnetic Field
26. Summary of New Measurements
27. Specs and Comparison of Three Companies
28. Ideal PMT for Auger
29. Summary of New Measurements
30. DC measurements and Specs Cathode sensitivity:
Sk > 50 uA/lm
Skb > 7 uA/lm-B (equivalent to QE > 16% at 420nm)
QE at 350 nm > 16% (337nm, 384nm for preproduction)
Gain and supply voltage:
Standard operation at Gain = 2x105
High Voltage should be less than 2000 V for Gain=1x106
Dark current and stability:
HV and Dark current will be measured at the following gains: 5x104, 2x105, 5x105, 1x106, 2x106
PMT should be stable at 1x106. This is defined by the following
I_dark(G=1x106) < 20 nA.
I_dark(G=1x106)/I_dark(G=2x105) < 10.0
I_dark(G=1x106) must be stable in time (no significant fluctuation).
The test listed above will be performed for all the PMTs,
prior to base attachment.
31. Pulse Measurements and Specs Dark pulse and after pulse
Measured at G=1x106 within 2 hours.
Dark pulse rate (>0.25pe) <10KHz
After pulse rate < 5% within the gate of 200nsec ~ 5,000nsec from the main signal. This is measured by taking the charge ratio of Q_after/Q_main.
Pulse linearity (no measurement for ~3 months)
Measured at Gain=2x105
Anode current should be linear within +-5% up to 50mA.
Single PE resolution
Measured at G=1x106
Pulsed light source must be tuned to have 90% pedestal, 10% signal for Poisson distribution of average 0.1 PE.
Peak to Valley ratio > 1.2
Cathode and Anode Uniformity
The test listed above will be performed for 1 out of 5 for the first 100 PMTs, then 1 out of 50 for the rest.
UCLA will do the all (except 4.)
32. PMT Potting and Assembly
33. 12 PMTs Package from Photonis
34. Fez and Collar from Colorado State.
35. PMT is aligned vertically
36. Collar is glued
37. Collar is glued
38. Collar is glued
39. Collar is aligned
40. Silicone Mixer (20:1 Ratio)
41. Silicon poured into Collar
42. Silicon poured into Collar
43. 12 PMTs after Potting, Waiting for Cure
44. PMT and Base potted by Silicon
45. Finished Assembly
46. Cables are glued
47. Environmental Testing