An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. MODIS-Terra cross-calibrationfor ocean color bands Ewa Kwiatkowska Bryan Franz, Gerhard Meister Ocean Biology Processing Group 13 May 2008 MODIS Science Team Meeting

  2. MODIS calibration changes since launch band 8 412nm SD – Solar Diffuser frame 979, mirror AOI 50.30 SV – Space View frame 23, mirror AOI 11.40 MS1 – Mirror Side 1 MS2 – Mirror Side 2 B side electronics pre-launch damage to mirror coating, MS2 Terra SD door permanently opened A side electronics

  3. MODIS trends in ocean color products Terra/Aqua ratios normalized water-leaving radiances Lwn Lwn(412nm) band 8 Lwn(443nm) band 9 Lwn(488nm) band 10 Lwn(531nm) band 11 Lwn(551nm) band 12 SD door permanently opened

  4. MODIS-Terra ocean color RVS band 8 412nm Terra response versus scan angle (RVS) in terms of normalized water-leaving radiances Lwn Lwn(412nm) band 8 Mirror side 1 MS1 Mirror side 2 MS2 +10% -10% per detector

  5. Issues and limitations with MODIS-Terra • Temporal and scan-dependent trend in Lwn of MODIS-Terra suggests • instrument RVS may be in error • polarization sensitivity may be changing • on-board calibration capabilities may be degrading (SD, SDSM) • On-board calibration (lunar or solar) CANNOT assess • changes in RVS “shape” • changes in polarization sensitivities • OBPG developed a vicarious approach for on-orbit Terra characterization • RVS • polarization sensitivity

  6. Lt Qt Ut 0 Modeling of TOA Stokes vector over oceans air whitecap glint gas aerosol water Lt(λ) = [ Lr(λ) + La(λ) + tLf(λ) + TLg(λ) + td(λ)Lw(λ) ] · tg(λ) from MODIS NIR assumes MCST NIR band characterization λ’ → λ fit based on bio-optical models SeaWiFS 9-day mean Lwn(λ’)

  7. Lm Qm Um 0 Lt Qt Ut 0 M11 M12 M13 M14 M21 … M24 M31 … M34 M41 M42 M43 M44 = · Vicarious TOA MODIS total signal Lm(λ) = M11Lt(λ) + M12Qt(λ) + M13Ut(λ)

  8. TOA sensor cross-calibration MODIS measured TOA radiance, polarized m Lm(λ) – M11Lt(λ) + M12Qt(λ) + M13Ut(λ) • minimize over global distribution of path geometries • find best M11, M12, M13 per band, detector, and mirror-side • M11, M12, M13 = f (mirror AOI) • do this for one day per month over the mission lifespan

  9. Polarization of the atmosphere √Qt2 + Ut2 dp = Lt degree of atmospheric polarization dp air molecule (Rayleigh) and glint scattering polarization correction fp pre-launch MODIS characterization MODIS Terra swath 412nm band 8 Lm fp = Lt

  10. Blue band temporal trends Mirror side 1 Mirror side 2 RVS = 1/M11 412 Detector 4 443 Space View (lunar) frame Nadir frame Solar Diffuser frame 488

  11. Blue band RVS & polarization sensitivity MS1 M13 fixed to pre-launch 412 443 488 Detector 4

  12. Blue band RVS & polarization sensitivity MS2 M13 fixed to pre-launch 412 443 488 Detector 4

  13. MODIS-Terra vicarious characterization MS1, Detector 4 MS2, Detector 4 RVS M12 M13 RVS M12 M13 412 443 488 531 551 667 678

  14. MODIS Terra and Aqua comparison Terra MS2, Detector 4 Aqua MS2, Detector 4 RVS M12 M13 RVS M12 M13 412 443 488 531 551 667 678

  15. Comparison of MODIS-Terra and MODIS-Aqua Lwn Before Vicarious Characterization Lwn(412nm) band 8 Lwn(443nm) band 9 Lwn(488nm) band 10 Lwn(531nm) band 11 Lwn(551nm) band 12 After Vicarious Characterization

  16. MODIS-Terra residual RVS in Lwn Before Vicarious Characterization 412 443 488 MS1 MS2 After Vicarious Characterization 412 443 488

  17. MODIS-Terra temporal anomalies Before Vicarious Characterization Lwn(488nm) band 10 chlor_a After Vicarious Characterization Lwn(488nm) band 10 chlor_a

  18. Summary • MODIS-Terra shows evidence of residual error in instrument response versus scan-angle and/or changes in polarization sensitivity. • The MODIS on-board calibration system can not assess changes in RVS “shape” or changes in polarization; these characteristics are restricted to our pre-launch knowledge, which itself may be in error. • The OBPG developed a vicarious on-orbit characterization for visible band polarization and RVS changes over time, using SeaWiFS Lwn. • Results show significant improvement in agreement between MODIS-Terra and MODIS-Aqua Lwn over the combined mission lifetime. • Future work will focus on vicarious characterization of the NIR. Franz, B.A., E.J. Kwiatkowska, G. Meister, and C.R. McClain (2008). Moderate Resolution Imaging Spectroradiometer on Terra: limitations for ocean color applications. J. Appl. Rem. Sens., in press.