XIS2 Calibration Status at Osaka U. 2004/03/16

XIS2 Calibration Status at Osaka U. 2004/03/16. Overview and current status of calibration in the soft X-ray region Energy-Pulse Height relation Energy-Resolution relation Quantum Efficiency Line Profile. K. Torii (Osaka U.), XIS2 team. Calibration method at Osaka U. Proportional counter.

afi
Download Presentation

XIS2 Calibration Status at Osaka U. 2004/03/16

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. XIS2 Calibration Status at Osaka U.2004/03/16 • Overview and current status of calibration in the soft X-ray region • Energy-Pulse Height relation • Energy-Resolution relation • Quantum Efficiency • Line Profile K. Torii (Osaka U.), XIS2 team

  2. Calibration method at Osaka U. Proportional counter XIS-EU (Seg. C) XIS-FM • We use a (calibrated) proportional counter as an absolute reference counter • Calibrate XIS-EU (Seg. C) against the reference proportional counter • Calibrate XIS-FM against XIS-EU (Seg. C)

  3. Calibration of EU Dispersed X-ray from spectrometer PD PC XIS-EU

  4. Calibration of FM Dispersed X-ray from spectrometer XIS-FM PD Fe55

  5. The datasets (2003/12-2004/3) • EU : 8s × 14000 fr. (2003/12/12-30, 2004/2/17-19) • FM S0: 8s × 5262 fr. (2004/2/26-27, 2004/3/8-10) • FM S1: 8s × 9225 fr. (2004/3/3-6) • FM S2: 8s × 11930 fr. (2004/1/6-14, 2/7-11) • FM S3: 8s × 12810 fr. (2004/1/26-2/6) • FM S1, S2, S3 are complete, some additional data have to be taken for S0 and EU • EU:125-126eV, FMS2:127-129eV, FMS3:128-130eV • EU:2.1-2.4e-, FMS2:1.9-2.5e-, FMS3:2.2-2.5e- (Normal, grade 0)

  6. Energy-PH relation (FM S2, Segment C) • PH=a・E + b • 1/a=3.720 eV/ch • b=1.38 ch Si-K edge

  7. Energy-PH relation (S0-A, S1-C) • Qualitatively similar results are obtained for all the sensors. S0, Seg.A, Grade0 S1, Seg.C, Grade0

  8. Energy-PH Relation (FM S3-A) • Fe55で決めたゲインと~0.1%で一致 • Qualitatively similar results are obtained for grade 0 and 02346.

  9. Energy-Resolution (FM S2, Segment C) • FWHM = sqrt (f02 + wf・ E) • f0=15.2eV, wf=0.45 Si-K edge

  10. Energy-Resolution (FM S3, Segment A) • Qualitatively similar results are obtained for all the sensors.

  11. Q.E.: Input flux is variable … Se-L (1.38 keV) W-M (1.77 keV) O-K C-K Auedge +15% Aledge Siedge

  12. W-M (1.77 keV) Ratio of 2004/2003 EU data Se L-α(1.38 keV)

  13. The dispersed spectra for Seg. C, Grade 02346 C-K O-K S0(×1024) S1(×256) EU2(×64) S2(×16) S3(×4) EU1 Se-L

  14. Uncorrected relative Q.E. (Seg. C, Grade 02346) S3 S2 EU2 S1 S0

  15. EXAFS and XANES at Oxygen K edge Oxygen K • FMS2, Segment 2, Grade 02346 Carbon K EXAFS, XANES

  16. Line profile (low energy tail) Oxygen-K (0.525 keV), H.V. = 1 and 5kV XIS2-S2 Counts 1 kV XIS1 (H.Katayama master thesis) • The effect of applied high voltage (5 kV) is almost negligible. • Tail component is significantly smaller than that for XIS 1. 5 kV Pulse Height

  17. Summary • Cal. Data were obtained with Fe55 and spectrometer (~0.2-3 keV). • E-PH, E-Resolution relations are as expected. • Low energy tail component is smaller than that for XIS1 sensors. • We find that X-ray flux from the generator is variable on long timescale. This has to be taken into account before calculating QE. • BPSG layer thickness does not seem to be significantly thinner for FM sensors than for EU (preliminary).

  18. Line profile (low energy tail) Counts • Tail component is significantly smaller than that for XIS 1. Pulse Height

  19. Dispersed spectra (Q.E.) S3 S2 S1 EU(2) S0 EU(1) Energy [keV]