180 likes | 330 Views
Orbital verification of the performance of Suzaku XIS. K. Hayashida, K. Torii, M. Namiki, N. Anabuki, S. Katsuda, N. Tawa, T. Miyauchi, H. Tsunemi, Osaka Univ. (Japan); H. Matsumoto, T. G. Tsuru, H. Nakajima, H. Yamaguchi, K. Koyama, Kyoto Univ. (Japan);
E N D
Orbital verification of the performance of Suzaku XIS K. Hayashida, K. Torii, M. Namiki, N. Anabuki, S. Katsuda, N. Tawa, T. Miyauchi, H. Tsunemi, Osaka Univ. (Japan); H. Matsumoto, T. G. Tsuru, H. Nakajima, H. Yamaguchi, K. Koyama, Kyoto Univ. (Japan); T. Dotani, M. Ozaki, H. Murakami, H. Katayama, Japan Aerospace Exploration Agency (Japan); S. Kitamoto, Rikkyo Univ. (Japan); H. Awaki, Ehime Univ. (Japan); T. Kohmura, Kogakuin Univ. (Japan); B. LaMarr, E. Miller, S. E. Kissel, M. W. Bautz, R. F. Foster, Massachusetts Institute of Technology
X-ray Imaging Spectrometer (XIS) • X-ray CCD Cameras onboard Suzaku • 3 Cameras (XIS0,XIS2,XIS3) contains FI-CCD、1 Camera(XIS1) has BI-CCD • CCD Operation Temp = -90℃ • 1024x1024pixels • FOV18’x18’ • Energy Range 0.2-12keV
Initial Operation and Current Status • 2005/7/10 Suzaku Launch • 2005/7/25 XIS System Power On • 2005/8/12-8/13 XIS Door Open =First Light • 2006/05/31More than 100 objects were observed. 4 CCD Cameras are functioning properly. E0102-72 JAXA press release
~0.5keV incident BI Small tail component!
Chemisorption Process Burke et al., 2004, IEEE transactions on Nuclear Science, 51, p.2322
すざくXIS • 7/10 0パス非可視で電磁バルブの開閉。 • 7/11 電磁バルブの開(4パス)閉(5パス)。 • 7/21 温度を上げてバルブ開。 • 7/24 MPU, PPU立ち上げ。 • 7/25 AE立ち上げ。データ取得開始(frame mode) • 7/26 CCD設定温度:−60℃ • 7/27 CCD設定温度:−90℃(ノミナル動作温度) • 8/11 CCD設定温度: −80℃。HP設定温度: −35℃。 • 8/12 XIS-3ドア開。 • 8/13 XIS-2,1,0 ドア開。 Better Energy Resolution at Low Energies
Suzaku Team Compiled by Fujimoto et al.
Orbital Calibration Items • Energy Scale • Tracking evolution of Charge Transfer Inefficiency (CTI) • Verification of Energy Scale in orbit • Quantum Efficiency • Verification of QE measured on ground • Monitoring possible change of QE in time • Background • Energy Resolution and Response Profile Updating Calibration Data Base continuously. Software tools are also required to be updated.
Ex-PHA relation (Calibration on ground) Residual to straight line fit Residual to broken line fit +10eV -10eV Si K edge (E=1839 eV)
Mn-Ka 5.9keV Mn-Kb 6.5keV 55Fe Cal-Source→Gain monitor counts • Gain Decrease ~2%/year • Energy Resolution 140eV ->170~180eV CTI increase induced by orbital radiation damage of the CCD PH [ch] Peak ch of Mn-Ka (Normalized to the ch at 1st ligt) Energy Resolution FWHM@5.9keV
Determination CTI parameters included in rev0.6 data Q’ = Q(1-CTI)N → CTI = (Q- Q’) / Q / N Q = PHA(ACTY=0) Q : Initial charge Q’ : Readout charge N : Number of P transfer PHA(Y=0) – PHA (Y=896, T) PHA(Y=0)×896 Whole Area cal src data (8/11) CTI = CTI(Seg1) = CTI(Seg 2) = [CTI(Seg0) + CTI(Seg3)] / 2 Corner cal src data (8/15~11/20) CTI = CTI_CONST + CTI_NORM×(PHAS)CTI_POW In rev0.6… CTI_POW = -0.5 CTI_CONST = 0 T Q’(T) = PHA(ACTY=896, T)
1% 1% <2% at lower energy, <1% at higher energy side. Revised correction (rev0.7 processing) is almost ready now. • Rev0.6 Correction as a function of time, location, assuming • CTI energy dependence Ex^-0.5 • CTI is constant at ACTY=0 Energy Scale Correction (Charge Trail&CTI) Cyg Loop Galactic Center Sgr C XIS0 , XIS1, XIS2 , XIS3 XIS0 , XIS1, XIS2 , XIS3 line center energy [keV] line center energy [keV] broken line: expected energy broken line: expected energy ACTY ACTY
By S.Okada et al., XRT team 20060213 QE + XRT effective area verification with Crab
QE (low energy part)CalibrationRXJ1856.5-37542005/10/24-26 Isolated NS C-K edge ~0.3keV Rev0.3 data -10eV offset 63.5eV blackbody a: Based Cal on the Ground b: a x excess0.15mmC c: Dead Layer =Design Value d: c x excess0.15mmC >0.3keV 1/3-1/2 of expected value
NeIX OVIII NeX MgXI OVII 2005-08-13 2005-08-31 2005-12-16 2006-01-17 2006-02-02 E0102-72 Repeated Observation→Degradation of QE • E0102: SNR in SMC, bright in soft X-ray lines • excellent calibrator for low-E gain, QE changes model thermal bremss + 24 Gaussian emission lines Galactic + SMC absorption pure C absorption from contaminant (varabs) gain shift -5 eV ~ -15 ev r2 ~ 1.6 (FIs) to 2.5 (BI)
Low background level is confirmed →Efficient for low surface brightness
Summary • 4 XIS CCD cameras are functioning properly • More than 100 targets were observed already. • Gain and energy resolution are monitored with built-in calibration source. CTI correction to compensate radiation damage induced gain decrease (~2%/year) is introduced. • Energy scale error in rev0.6 processing is 1%-2%. Rev0.7 processing with higher accuracy is now almost ready to use. • QE at high energy is almost as expected (error ~10%) from Crab data. • Low energy QE suffers significant degradation. • Possibly C-dominant material put on Optical Blocking Filter (OBF) • (Details will be presented in a separate paper) • Activity to update calibration and their tool is under going.