1 / 29

RHIC Polarimetery in Run9, sqrt (s)=200 GeV

RHIC Polarimetery in Run9, sqrt (s)=200 GeV. A.Bazilevsky for RHIC Polarimetry group May 12, 2009. HJet. HJet. ToF vs Energy. Forward scattered proton. proton beam. proton target. recoil proton. Energy vs Channel #. YELLOW side. BLUE side.

aldan
Download Presentation

RHIC Polarimetery in Run9, sqrt (s)=200 GeV

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. RHIC Polarimetery in Run9, sqrt(s)=200 GeV A.Bazilevsky for RHIC Polarimetry group May 12, 2009

  2. HJet

  3. HJet ToF vs Energy Forward scattered proton proton beam proton target recoil proton Energy vs Channel # YELLOW side BLUE side Array of Si detectors measures TR&ToFof recoil proton. Channel # corresponds to recoil angle R. Correlations (TR & ToF ) and (TR & R )  the elastic process

  4. Hjet Fill 10663

  5. Hjet: StabilityTarget asymmetry • Should be constant vs fill (for fixed Ptarget) • Should be consistent between blue and yellow • 0.04170.0004 vs 0.04180.0004 • Should be consistent between Years (for fixed Ptarget) • Run5: 0.04090.0004 • Run6: 0.04130.0002 • Run8: 0.04220.0003 Multi-Year consistency within ~1%

  6. Hjet: StabilityBackground Background (with two beam mode) is on the same level as with one beam mode ~Constant vs fill On the same level as in Run5&6 (removed collimators in Run8  larger background)

  7. Hjet: results <P>=55% <P>=55% Fill-by-fill variations as well as variations within a fill should be estimated from pC polarimter

  8. pC

  9. 6 1 2 5 3 4 Ultra thin Carbon ribbon Target (5 mg/cm2) pC 18cm Two independent polarimeters in each ring (so 4 polarimeters!) 12 carbon targets in each polarimter (6 vertical and 6 horizontal) 6 detectors in each polarimeter 3 types of detectors: “BNL strip”: 12 strip in each detector “Hamamatsu strip”: 12 strip in each of 2 detector (blue2) “Hamamatsu pad”: 2 pads in each of 2 detector (yellow2) Si strip detectors (TOF, EC)

  10. pC Rate history s=200 GeV • Severe rate problems observed during Run9 sqrt(s)=500 GeV • Higher rate  lower measured polariation • Target changed in Yellow1 to thicker one (from Fill 10686)  Rate problems!

  11. pC measurements Online Polarization , not normalized (!) vs fill Fills 10616 (Apr 18) – 10713 (May 8) “Online” polarizations: 0.50-0.60 Pol-1 measure slightly lower than Pol-2: by ~6% Blue1/Blue2: consistent within stat. uncertainties Yell1/Yell2: shows variations above stat. uncertainties May show a gradual decrease (5-7%) from fill 10616 to 10713 may come from detector degradation

  12. pC-BluevsHJet Hjet/pC is stable over fills within (large) stat. errors (of HJet) HJet: <P>=55% (fills 10616-10713) HJet/Blue1  1.05 HJet/Blue2  0.99

  13. pC-Yellow vs HJet Hjet/pC is stable over fills within (large) stat. errors (of HJet) HJet: <P>=55% (fills 10616-10713) HJet/Yell1  1.07 HJet/Yell2  1.00

  14. More precise Hjet-pC comparison Hjet: fills combined in 8 periods Clear correlation between Hjet and pC Consistency vs fill within 5% pC-blue HJet pC-yellow HJet

  15. pC: Pol. Profile Polarimeters 2 Polarimeters 1 Horizontal profile Vertical profile Vertical profile Horizontal profile Usual… R0.15 in previous years (100 GeV beams) R0.1  Experiments see 5% more polarization than Hjet

  16. pC: Consistency within a Fill

  17. pC: Consistency within a Fill Clear polarization decay Consistent between Pol1 and Pol2 10685-Blue 10704-Blue

  18. pC: Consistency within a Fill Prob(2, NDF) – from the fit to a constant in a fill Should be uniform if variations within a fill are only due to stat. errors Should show higher density near 0 if there are “sizable” syst. effects No systematic effects (comparable to stat. errors) are seen within a fill

  19. Pol. Decay In a fill: fit to exp(-t/Tdecay) <Tdecay> = 80-180 hours <Tdecay> = 250-800 hours May need rate correction! • Run6: • <Tdecay> ~150 hours • Run8: • <Tdecay> ~400 hours • <Tdecay> ~100 hours

  20. pC: Injection

  21. pC: pol. at injection <P>~55% (similar to flattop) Pol1 measures smaller values than Pol2 by ~5% (similar to flattop, or slightly smaller due to smaller rate effect at injection)

  22. pC: injection vs flattop On the average no difference within 4%: Assuming that we don’t lose polarization on the ramp  AN(inj)/ AN(ftp) is known within 4% Correction due to rate effect and polarization decay at store may be needed Assuming that AN(inj)/ AN(ftp) is correct  Polarization loss on the ramp <4%

  23. Summary • HJet: • Running in stable conditions: <P>~55% • pC: • Blue1 vs Blue2 consistent behavior • Yell1 vs Yell2 show systematic effects ~5-7% (may be due to larger rate effects in Yell1) • pC vs Hjet: consistent within stat. errors • Hjet/Pol1 ~ 1.06; Hjet/Pol2 ~ 1 • Might be ~5% drop in the pC measurements from the beginning to the end of the run due to detector degradation (“dead layer” increase) • Measurements are statistically consistent within a fill • Polarization decay Tdecay ~ 100-200 hours • Polarization profile no sharper than in previous years • Experiments see ~5% more polarization than Hjet • Measurements at injection and flattop are consistent within ~4% • Correction for polarization decay and rate effects at flattop may lead to even better consistentcy… • “Fast offline” results can be used only with special care/caution • Needs monitoring: parameter retuning sometimes fails

  24. Backups

  25. Rate history s=500 GeV

  26. C Mass

  27. P 2. Obtain R directly from the P(I) fit:    I R=0.290.07 pC: Polarization Profile Scan C target over the beam cross: pC 1. Directly measure I and P : P Polarization I Intensity Target Position Precise target positioning is NOT necessary

  28. Statistically Ok !

More Related