Gradient Echoes, Diffusion, & EPI

Gradient Echoes, Diffusion, & EPI. Two recent MRI clinical research tools - Echo Planar Imaging ( EPI ) Diffusion Weighted Imaging ( DWI ). Clinical EPI applications. Very rapid imaging (fastest clinical). 128x128 image < 100 ms.

aldona
Download Presentation

Gradient Echoes, Diffusion, & EPI

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Gradient Echoes, Diffusion, & EPI • Two recent MRI clinical research tools - • Echo Planar Imaging ( EPI ) • Diffusion Weighted Imaging ( DWI )

  2. Clinical EPI applications • Very rapid imaging (fastest clinical). • 128x128 image < 100 ms. • Multi-slice first pass contrast enhanced brain perfusion. • Functional MRI ( BOLD fMRI ). • Real-time cardiac imaging ?

  3. DWI • Measures diffusion of water. • Diffusion - • Random (Brownian) motion of water. • RMS distance travelled in a fixed time would be a measure of diffusion. Slow diffusion Rapid diffusion

  4. Clinical DWI applications • Ischemia • Ischemic tissue exhibits reduced diffusion. • Intracellular water • - low (restricted) diffusion. • Extracellular water • - higher diffusion. ischemia  cytotoxic edema  reduction in extracellular volume fraction  reduction in overall diffusion

  5. Clinical DWI applications • Ischemic tissue - reduced diffusion. • Reduction observed within about half an hour of ischemia. • T2 increase not seen for 6 - 12 hours (recruitment of excess tissue water). • Assessment of acute stroke. T2 weighted diffusion weighted

  6. Clinical DWI applications • Tumours. • Help distinguish tumour, cystic changes, vasogenic edema, normal white matter. • White matter disease. • Disruption of structure affects local diffusion of water.

  7. S N S N Nuclear Magnetic Resonance Imaging • Nucleus of atom is spinning. • Causes it to behave like a tiny magnet. • Nuclei align (almost) parallel to external magnetic field, like compass needle.

  8. Magnetic Resonance Imaging • Because nucleus is spinning, it can not align exactly parallel to the magnet field. • This causes the N-S axis of the nucleus to rotate around the N-S axis of the main magnetic field. • Precession. North of main magnetic field North of nucleus’s magnetic field

  9. Magnetic Resonance Imaging • The rate of precession is important in MRI. • The number of revolutions per second (frequency) of a precessing nucleus depends on the main magnetic field strength. • Although individual nuclei are not aligned exactly parallel to the main magnetic field, their average alignment is parallel. Millions of nuclei involved in MRI.

  10. Magnetic Resonance Imaging • In the Earth magnetic field • ( 0·00005 T ), hydrogen precesses at about 2100 revolutions per second (Hertz). • In the Vision MRI scanner ( 1·5 T ) hydrogen precesses at 63 million Hertz. • Larmor frequency.

  11. Magnetic Resonance Imaging • “wobble” a nucleus at same rate as it’s precessing  tips its alignment away from main magnetic field. • “flip angle”. • Use electromagnetic radiation to do the “wobbling”. • 1·5 T  63 MHz  radiowaves (FM) • Interaction due to resonance between precessing nucleus and radiowaves. • Radiowaves in  RF pulse. • Applied to whole sample.

  12. Magnetic Resonance Imaging • Switch off RF pulse  nuclei realign with main magnetic field. • As they realign, emit radiowaves at Larmor frequency  NMR signal out. • NMR signal detected by a RF coil (fancy FM aerial). • Vast majority of NMR signal from hydrogen in water or fat only.

  13. Magnetic Resonance Imaging • MR images are primarily images of water and fat. • To produce an image, apply smaller magnetic fields. • Add / subtract with main magnetic field. • Magnetic field gradients.

  14. 1·5 T Magnetic field, B Magnetic field, B 0 x x With gradient Without gradient Magnetic Resonance Imaging • Each point along x axis, different B. • Nuclei precess at different rates. • Emit r/w at different frequencies, depending on their x position. • Spatial encoding of NMR signal.

  15. Free Induction Decay (FID) time NMR signal out RF pulse in Readout gradient time Magnetic Resonance Imaging • To acquire one line of image • Apply gradient during acquisition to spatially encode NMR signal.

  16. Readout gradient time Magnetic Resonance Imaging • Consider nuclei in left & right eyes during FID and 2 ms readout gradient. 1·501 1·5 1·5 1·499 Rt. Lt. Magnetic field, B Magnetic field, B 0 0 x x Freq. 63 63 62·958 63·042 • After 2 ms, Rt. eye: 125916 rotations • Lt. Eye: 126084 rotations • Out of step  no gross NMR signal.

  17. Magnetic Resonance Imaging • Practically & theoretically better to separate application of RF pulse and reception on NMR signal. • Arrange peak (in phase) NMR signal in middle of readout gradient. Gradient Echo time RF pulse in NMR signal out Readout gradient time

  18. 1·501 1·501 1·499 1·499 Magnetic field, B Magnetic field, B 0 0 x Gradient Echo A B C D Readout gradient 2 2 2 time Rt. Lt. Rt. Lt. x A AB BC C D Rt. eye: 0 126084 125916 252000 377916 Lt. Eye: 0 125916 126084 252000 378084 in step initially de-phased by gradient in step again - echo

  19. 1·501 1·501 1·499 1·499 Magnetic field, B Magnetic field, B 0 0 x Gradient Echo - moving nucleus A B C D Readout gradient 2 2 2 time x A AB BC C D Rt. eye: 0 125916 126084 252000 378084 Lt. Eye: 0 126084 125916 252000 377916 moving: 0 126084 126084 252168 378252 Same argument for nuclei only 1 µm apart.

  20. Gradient Echo - moving nucleus • On the scale of an image pixel within the object - • Perfusion: Coherent motion. Entire pixel has same phase shift. • Diffusion: Incoherent random motion. Signal in a pixel is sum of random phase shifts  cancel one another out  suppression of signal  diffusion weighting. • High diffusion  large suppression of signal  dark pixel in DWI. • Low diffusion  little suppression of signal  bright in DWI.

  21. Diffusion weighting • Diffusion  smaller effect than perfusion  small amount of dephasing. • Noticeable DW required very large magnetic gradients. • Separate DW gradients from imaging gradients.

  22. Diffusion weighting RF NMR signal time Readout gradient time Stationary: dephased contributes to nucleus rephased echo. Diffusing: dephased reduced contribution nucleus not fully to echo rephased Diffusion imaging gradients gradients

  23. Diffusion weighting • In practice, use spin echo with DW. 180º NMR signal 90º time Readout gradient time Stationary: dephased rephased contributes to nucleus echo. Diffusing: dephased not fully reduced contribution nucleus rephased to echo Diffusion imaging gradients gradients

  24. Diffusion weighting • Problem: DWI sensitive to any random movements. • Patient movements, e.g., cardiac pulsations, dominate over diffusion. • One solution - use very fast MRI and ‘freeze’ unwanted motions. • Echo Planar Imaging ( EPI )

  25. Gradient echo MRI • Single gradient echo - one line of image. RF time Readout gradient time • Build entire image with FLASH. RF …. time Readout gradient …. time • About 1 second. Poor signal to noise.

  26. EPI • Multiple gradient echoes RF time Readout gradient time Phase: in out in out (stationary nucleus) But, signal still potentially available. Re-phase it with another gradient. Readout gradient time Phase: in out in out in out

  27. EPI • Multiple gradient echoes RF time …. Sign: + - + - + - Readout gradient …. time …. Phase: in out in out in out in out in out in out in out

  28. EPI • NMR signal from one RF pulse lasts between 50 - 200 ms ( T2 decay ). Have to acquire all echoes within this time. • Limited to about 128 echoes, i.e., 128x128 image matrix. • The faster gradients can be switched + to - the more echoes in a fixed time.  Entire image in less than 100 ms.  Physiological motion frozen.  Relatively high S/N (c.f. FLASH).  More spatial distortion & artefacts.  Not a high resolution technique.  High spec. hardware required.

  29. DW-EPI 180º 90º time Readout gradient time Diffusion EP imaging gradients gradients

  30. DW-EPI T2 weighted EPI DW-EPI • High diffusion  large suppression of signal  dark pixel in DWI. • Low diffusion  little suppression of signal  bright in DWI.

  31. DW-EPI • DW another MRI parameter • (c.f., T1 and T2 weighting ) • DW-EPI also is heavily T2 weighted (need long TE to fit in extra diffusion gradients). EPI is inherently T2 weighted already. • Bright signal in DWI could also be due to long T2 and vice versa. • “T2 shine through” • To just measure diffusion, calculate the Apparent Diffusion Coefficient ( ADC ). • ‘Apparent’ because averaged over a pixel, and contains some perfusion.

  32. Apparent Diffusion Coefficient No diffusion (stationary) Low diffusion log( DWI ) High diffusion b 0 1000 • b is strength of DW gradients. • Larger b value  more DW. • slope of line is the ADC. • To calculate ADC, need minimum of 2 points on line. • We choose b=0 (i.e., T2 weighted EPI) and b=1000 (DW-EPI). • ADC is a quantifiable parameter.

  33. Apparent Diffusion Coefficient b=0 b=1000 ADC map ( T2 weighted ) • In ADC map - • Bright pixel  large ADC. • Dark pixel  small ADC. • infarct  dark • normal brain  grey • CSF  bright • No potential for “T2 shine through” in ADC map.

  34. Apparent Diffusion Coefficient • Amount of diffusion (ADC) also depends on direction. • In free water, diffusion should be the same in all directions (isotropic) • In structures (e.g., white matter tracts) get more diffusion along the tracts than perpendicular (anisotropic). • Shown DW gradients along x axis. Acquire separate DWI with diffusion along y or z axes. • Construct diffusion tensor. • A tensor gives directional information.

  35. Apparent Diffusion Coefficient Normal brain. ADC map. (amount of diffusion regardless of direction) Relative Anisotropy (RA) map (from tensor). (how uni-directional diffusion is) • White matter tracts bright - • all diffusion in one direction along tracts.

  36. Apparent Diffusion Coefficient Post radiotherapy (AVM). T2 weighted ADC map RA map Stroke. ADC map RA map

  37. Summary • Gradient echo always required to acquire a MR image. • Sensitive to motion. • Can use this to measure diffusion. • Fast MRI sequence needed to freeze all other motions  EPI. • Quantify amount of diffusion using ADC maps. • Quantify direction of diffusion using tensor maps. • http://www.nottingham.ac.uk/radiology/