Objective . To create an automotive glass database that will allow scientist to identify unknown glass samples from a crime scenes.. Why IR Spectroscopy?. There is no current forensic database for glass analysis using IR techniques.Refractive index is the only instrumental technique used for way glass analysis in forensic laboratories today...
E N D
1. IR Spectroscopy: A Promising Technique for Glass Analysis in Forensic Identification By :
Brian Vlaisavich, Biology, Junior, Lewis University
Pradip Patel, Biology, Junior, Lewis University
2. Objective To create an automotive glass database that will allow scientist to identify unknown glass samples from a crime scenes.
3. Why IR Spectroscopy? There is no current forensic database for glass analysis using IR techniques.
Refractive index is the only instrumental technique used for way glass analysis in forensic laboratories today..
4. Refractive Index Outdated instrumental method
Method is neither accurate nor precise.
Variation over time of these databases causes for improper statistical calculations.
5. Benefits of using FTIR Method is quick and reliable.
Provides:
reliable accuracy
desirable speed and efficiency of data analysis and collection.
6. History of IR Developed in the 1960's
Only intended for advanced research
High cost made it impossible to be used for commercial use
Advancements in technology - FTIR
reduced cost
enhanced capabilities making it a standard instrument in modern labs.
7. How does IR work? 1. consists of a source, beamsplitter, two mirrors, laser, and detector.
2. beamsplitter splits the beam into two parts.
3. One part is transmitted to a moving mirror; other part is reflected to a fixed mirror.
4. The moving mirror moves back and forth at a constant velocity. This velocity is timed according to the very precise laser wavelength in the system which also acts as an internal wavelength calibration.
5. The two beams are reflected from the mirrors and recombined at the beamsplitter.
8. Continued . . . 6. The beam from the moving mirror has traveled a different distance than the beam from the fixed mirror.
7. When the beams are combined an interference pattern is created, since some of the wavelengths recombine constructively and some destructively.
8. This interference pattern is called an interferogram.
9. This interferogram then goes from the beamsplitter to the sample, where some energy is absorbed and some is transmitted.
10. The transmitted portion reaches the detector.
11. The detector reads information about the wavelengths in the infrared range simultaneously.
9. Further explanation of IR… IR spectroscopy is the use of light in the mid-infrared range to excite the bonds in a compound in order to learn about it by seeing which frequencies are absorbed.
10. Two main uses of IR spectroscopy in Forensic Science: Confirmation of Functional Groups
Confirmation of the identity of a substance
11. Theory of IR Absorption IR absorption is restricted to compounds with small energy differences in the likely vibrational and rotational states.
For a molecule to absorb in IR, the vibrations or rotations must cause a net change in dipole moment of the molecule.
If the frequency of radiation matches the vibrational frequency of the molecule, then radiation will be absorb in the form of a peak, also known as a ‘fingerprint’
12. Interpretation of IR spectrum? IR peaks shows where the infrared radiation was absorbed in the molecule.
Collective analysis of peak shapes and frequencies, define the functionality/identity of the compound.
13. Why do IR peaks have different intensities and/or different frequencies? The intensity of an IR absorption is proportional to:
The strength of the dipole moment change associated with the vibration
The number of bonds in the molecule
14. Types of Stretching Vibrations Symmetrical stretching
the bonds stretch with the same strength at the same time.
Asymmetrical stretching
The bonds stretch at different times causing a change in the dipole moment.
15. Bench Status
Must have a Green Check
Collect
Experiment setup
Collect Tab
# of scans 32
Resolution 4
Final Format: % transmittance
Collect background after (Large #)
32 scans for background
17. Bench Tab
Gain: autogain
Min Range 600 nm
Select OK
18. Analyze
Library Setup
Select Glass Library
Select Add
Select OK
19. Collect background
Click Col Bkg
Select OK
Add to window 1: Select NO
20. Place glass into cardboard
Place into middle slit
Cover the top
21. Collect Sample
Click Col Smp
Add title of sample with date
Select ok
Add to window 1: Select Yes
Add to Library
22. Results Created a library for glass analysis:
Entered known samples previously collected from University of St. Francis students
Entered known samples collected at Canal St. Junkyard in Lockport, IL
Various other samples also entered
23. Confirmation of Samples Each sample was entered into the database multiple times
Spectra that had the largest % match was then saved to the library
24. Test: 1985 Ford Bronco II Rear Back Window
25. Experimental Problems Samples similar to each other produced indistinguishable spectra
Samples that are
Curved in shape
Bubbled in shape
Cracked in shape
Too small in size
All produced distorted spectra.
26. Discussion Car manufactures may use similar compositions of glass in production
A larger range of different makes and models of cars needs to be collected
Most of the samples collected:
American cars
Ford
Chevrolet
27. Curved/Bubbled Samples Produced different spectra depending on:
Angle and direction of light on sample
Example:
tail lights, head lights, etc.
28. Example: 1989 Pontiac Grand Prix Tail Light
29. Tail Light Spectra: Example 1
30. Tail Light Spectra Example 2
31. Identification of Unknown Samples Three unknown glass samples were analyzed
Goal:
To determine the source of glass based on the Automobile Library that we created
32. Unknown 1
33. Unknown 2
34. Unknown 3
35. Future Research More samples can be collect to build the library
Spectra collected can still use continued updating to be more accurate
36. Unknown Data Unknown 2 & 3
appear to not match well with anything in the library.
Unknown 1
70% match to 1995 Dodge Intrepid
Possible relationship?
37. Discussion/Conclusion From the initial research it appears that FTIR is a reliable method for glass analysis
Glass samples matched with corresponding cars
However:
Large % match was not always obtained
Future research necessary
38. References FT-IR Dispersive Infrared. Theory of Infrared Spectroscopy Instrumentation. www.thermo.com/nicolet
Infrared Spectroscopy. http://www.cem.msu.edu/~reusch/VirtTxtJml/Spectrpy/InfraRed/inf rared.htm
Introduction to IR Spectra. http://www.chem.ucla.edu/~webspectra/irintro.html
Koons, Robert D. Distribution of Refractive Index Values in Sheet Glasses. Forensic Science Research Unit. www.fbi.gov/hq/lab/fsc/backissu/jan2001/koons.htm
Lesney, Mark S. A Study in Infrared. Today’s Chemist at Work. October 2003. www.tcawonline.org
39. Acknowledgement Lewis University
University of St. Francis
Professor S. M. Diab
Br. Pierre St. Raymond
Canal St. Junkyard (Lockport, IL)