An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. An Introduction to Chromatographic Separations Lecture 34

  2. An Introduction to Chromatographic Separations • It was Mikhail Tswett, a Russian botanist, in 1903 who first invented and named liquid chromatography. • Tswett used a glass column filled with finely divided chalk (calcium carbonate) to separate plant pigments. • He observed the separation of colored zones or bands along the column, hence name chromatography, where Greak chroma means color and graphein means write.

  3. The development of chromatography was slow for reasons to be discussed later and scientists waited to early fifties for the first chromatographic instrument to appear in the market (a gas chromatograph). • However, liquid chromatographic equipment with acceptable performance was only introduced about two decades after gas chromatography.

  4. According to the nature of the mobile phase, chromatographic techniques can be classified into three classes: • Liquid chromatography (LC) • Gas chromatography (GC) • Supercritical fluid chromatography (SFC) Other classifications are also available where the terms: Column chromatography: Chromatographic separations take place inside a colum Planar chromatography: The stationary phase is supported on a planar flat plate, are also used.

  5. General Description of Chromatography • In a chromatographic separation of any type, different components of a sample are transported in a mobile phase (a gas, a liquid, or a supercritical fluid). • The mobile phase (also called eluent) penetrates or passes through a solid or immiscible stationary phase. • Solutes (eluates) in the sample usually have differential partitioning or interactions with the mobile and stationary phases. • Since the stationary phase is the fixed one then: • Those solutes which have stronger interactions with the stationary phase will tend to move slower (have higher retention times) than others. • But solutes which have lower or no interactions with the stationary phase will tend to move faster.

  6. Therefore, chromatographic separations are a consequence of differential migration of solutes. • It should be remembered that maximum interactions between a solute and a stationary phase take place when both have similar characteristics, for example in terms of polarity. • However, when their properties are so different, a solute will not tend to stay and interact with the stationary phase and will thus prefer to stay in the mobile phase and move faster. • A good example: A polar solvent and a non polar stationary phase.

  7. Elution Chromatography • The term elution refers to the actual process of separation. • A small volume of the sample is first introduced at the top of the chromatographic column. • Elution involves passing a mobile phase inside the column whereby solutes are carried down the stream but on a differential scale due to interactions with the stationary phase. • As the mobile phase continues to flow, solutes continue to move downward the column. • Distances between solute bands become greater with time and as solutes start to leave the column they are sequentially detected.

  8. Rretention time: The time a solute spends in a column. It depends on the fraction of time that solute spends in the mobile phase. • Band broadening: As solutes move inside the column, their concentration zone continues to spread. The extent of spreading depends on the time a solute spends in the columns. • Factors affecting band broadening are very important and will be discussed later. • The dark colors at the center of the solute zones in the above figure represent higher concentrations than are concentrations at the sides. This can be represented schematically as:

  9. Chromatograms • The plot of detector signal (absorbance, fluorescence, refractive index, etc..) versus retention time of solutes in a chromatographic column is referred to as a chromatogram. • The areas under the peaks in a chromatogram are usually related to solute concentration and are thus very helpful for quantitative analysis. • The retention time of a solute is a characteristic property of the solute which reflects its degree of interaction with both stationary and mobile phases. • Retention times serve qualitative analysis parameters to identify solutes by comparison with standards.

  10. Migration Rates of Solutes • The concepts which will be developed in this section will be based on: separation of solutes using a liquid mobile phase and an immiscible liquid stationary phase. • This case is particularly important as it is a description of the most popular processes.

  11. Distribution Constants • Solutes traveling inside a column will interact with both the stationary and mobile phases. • If, as is our case, the two phases are immiscible, partitioning of solutes takes place and a distribution constant, K, can be written: Where; CS and CM are the concentrations of solute in the stationary and mobile phases, respectively. • If a chromatographic separation obeys the previous equation , the separation is called linear chromatography. • In such separations peaks are Gaussian and independent of the amount of injected sample.

  12. The retention time tR: The time required for an analyte to travel through the column after injection till the analyte peak reaches the detector is termed tR. The dead or void time, tM: The time spent by species to exit the column if the sample contains an un-retained species, such species travels with the mobile phase. • Solutes will move towards the detector in different speeds, according to each solute’s nature.

  13. Retention Times tM = retention time of mobile phase (dead time) tR = retention time of analyte (solute) tR’ = time spent in stationary phase (adjusted retention time) L = length of the column

  14. Velocity = distance/time  length of column/ retention times Velocity of solute: Velocity of mobile phase:

  15. average linear velocity of a solute, v, can be written as: Where, L is the column length and tR is the retention time of the solute. The mobile phase linear velocity, , can be written as: The linear velocity of solutes is a fraction of the linear velocity of the mobile phase. This can be written as:

  16. Chromatography Velocity/Retention time and Kc

  17. Dividing both nominator and denominator by CMVM we get Now, let us define a new distribution constant, called the capacity or retention factor, k’, as:

  18. By Substitution we get: Rearrangement after subst. for V and  gives: as and This equation can also be written as: or as

  19. The Selectivity Factor • For two solutes to be separated, they should have different migration rates. • This is referred to as having different selectivity factors with regard to a specific solute. • The selectivity factor, a, can be defined as: Therefore, a can be defined also as: For the separation of A and B from their mixture, the selectivity factor must be more than unity.

  20. The Shapes of Chromatographic Peaks • Chromatographic peaks will be considered as symmetrical normal error peaks (Gaussian peaks). • This assumption is necessary in order to continue developing equations governing chromatographic performance. • However, in many cases tailing or fronting peaks are observed.

  21. Gaussian peaks (normal error curves) are easier to deal with since statistical equations for such curves are well established and will be used for derivation of some basic chromatographic relations. • It should also be indicated that as solutes move inside a column, their concentration zones are spread more and more where the zone breadth is related to the residence time of a solute in a chromatographic column.

  22. Plate Theory • Solutes in a chromatographic separation are partitioned between the stationary and mobile phases. • Multiple partitions take place while a solute is moving towards the end of the column. • The number of partitions a solute experiences inside a column very much resembles performing multiple extractions. • It may be possible to denote each partitioning step as an individual extraction. • The column can thus be regarded as a system having a number of segments or plates, where each plate represents a single extraction or partition process.

  23. Therefore, a chromatographic column can be divided to a number of theoretical plates where eventually the efficiency of a separation increases as the number of theoretical plates (N) increases. • In other words, efficiency of a chromatographic separation will be increased as the height of the theoretical plate (H) is decreased.

  24. Column Efficiency and the Plate Theory If the column length is referred to as L, the efficiency of that column can be defined as the number of theoretical plates that can fit in that column length. This can be described by the relation: N = L/H

  25. The plate theory successfully accounts for the Gaussian shape of chromatographic peaks but unfortunately fails to account for zone broadening. • In addition, the idea that a column is composed of plates is unrealistic as this implies full equilibrium in each plate which is never true. • The equilibrium in chromatographic separations is just a dynamic equilibrium as the mobile phase is continuously moving.

  26. Definition of Plate Height • From statistics, the breadth of a Gaussian curve is related to the variance التباينs2. • Therefore, the plate height can be defined as the variance per unit length of the column: H = s2/L • In other words, the plate height can be defined as column length in cm which contains 34% of the solute at the end of the column (as the solute elutes). • This can be graphically shown as:

  27. The peak width can also be represented in terms of time, t, where: t = s/v(s ) is the standard deviation t = s/(L/tR) as • The width of the peak at the baseline, W, is related to t by the relation: W = 4t where 96% of the solute is contained under the peak. s = L t/tR s = LW/4tR

  28. s2 = L2W2/16tR2 s2 = HL H = LW2/16tR2 N = L/H N = 16(tR/W)2 Also, from statistics we have: W1/2 = 2.354 t s2 = LH

  29. s = L t/tR • = L (W1/2/2.354)/tR s2 = L2 (W1/2/2.354)2 /tR2 Substitution in the prev. eq. (H = s2/L )gives: LH = L2 (W1/2/2.354)2 /tR2 H = L (W1/2/2.354 tR)2 N = L/H N = 5.54 (tR/W1/2)2 Another way to derive the efficiency equation:

  30. From statistics we know that the standard deviation (or W of peak) is supposed to be inversely related to the square root of the number separation stages (theoretical plates): (N is the efficiency) However, we also know that the standard deviation is proportional to retention time (as it  to W): We can write: The proportionality constant is practically = 1 Therefore It is also true that 95.5% of molecules in a peak are contained within + 2swhich means that the base of the peak is equal to 4s ; or Substitution for s gives using: so

  31. Asymmetric Peaks The efficiency, N, can be estimated for an asymmetric chromatographic peak using the relation: At 10% Peak Height N = 41.7 (tR/W0.1)2 / (A/B + 1.25) Draw a horizontal line across the peak at a height equal to 1/10 of the maximum height. Where W0.1 = peak width at 1/10 height = A + B

  32. B distance > A distance A distance > B distance

  33. Suggested As values : for improvement

  34. An Introduction to Chromatographic Separations Lecture 35

  35. W1/2

  36. Rate Theory or Kinetic Theory Band Broadening • Apart from specific characteristics of solutes that cause differential migration, average migration rates for molecules of the same solute are not identical. • Three main factors contribute to this behavior:

  37. 1. Longitudinal Diffusion Molecules tend to diffuse in all directions because these are always present in a concentration zone as compared to the other parts of the column.  This contributes to H as follows: HL = K1DM/V  Where, HL is the height of plate resulted from the long. Diff., DM is the diffusion coeff. of solute in the mobile phase, K 1is const. This factor is not very important in liquid chromatography except at low flow rates. But is of strong effect in GC.

  38. Longitude diffusion

  39. 2. Resistance to Mass Transfer Mass transfer through mobile and stationary phases contributes to this type of band broadening. a. Stationary Phase Mass Transfer • This contribution can be simply attributed to the fact that not all molecules penetrate to the same extent into the stationary phase. • Therefore, some molecules of the same solute tend to stay longer in the stationary phase than other molecules.

  40. Quantitatively, this behavior can be represented by the equation: Hs = K2 ds2V /Ds Where, Hs: Height of plate results from molecules resistance in SP, ds is the thickness of stationary phase and Ds is the diffusion coefficient of solute in the stationary phase. V: average velocity of MP

  41. b. Mobile Phase Mass Transfer • Solute molecules which happen to pass through some stagnantراكد mobile phase regions spend longer times before they can leave. • Molecules which do not encounter such stagnant mobile phase regions move faster. • Other solute molecules which are located close to column tubing surface will also move slower than others located at the center. • Some solutes which encounter a channel through the packing material will move much faster than others.

  42. HM = K3dp2V /DM Where dp is the particle size of the packing.