ISO 10 360-5 (2000) 1 / 2 Performance of CMM‘s using multiple stylus probing systems

465. l PE. l. ISO 10 360-5 (2000) 1 / 2 Performance of CMM‘s using multiple stylus probing systems. Fixed Probe Head Articulating Probe Head. ML, MS, MF. ISO Test procedure:. AL, AS, AF. 5 orthogonal probe pins of length l are calibrated.

alissa
Download Presentation

ISO 10 360-5 (2000) 1 / 2 Performance of CMM‘s using multiple stylus probing systems

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 465 lPE l ISO 10 360-5 (2000) 1 / 2 Performance of CMM‘s using multiple stylus probing systems Fixed Probe Head Articulating Probe Head ML, MS, MF ISO Test procedure: AL, AS, AF 5 orthogonal probe pins of length l are calibrated. 1 probe pin (length 20mm) with extension lPE is calibrated in 5 orthogonal positions. A reference sphere is measured with each probe resp. with each calibrated position. Every measurement takes 25 probings, total number of probings is 5 x 25 = 125. Evaluations: Multiple Stylus Location errorML resp. AL = Max. Range of the 5 centre coordinates in X, Y or Z. Multiple Stylus Size error MS resp. AS = Deviation from the calibrated diameter (all 125 points). Multiple Stylus Form error MF resp. AF = Form error of the calculated sphere (all 125 points). JR - 2/2002

  2. 465 lPE AL / ML l Location error => Biggest axial distance between the 5 measured centerpoints ISO 10 360-5 (2000) 2 / 2 Performance of CMM‘s using multiple stylus probing systems Fixed Probe Head / Articulating Probe Head AS / MS Size error over 125 points (from 5 different probes (fixed head) or 5 different orientations (Renishaw head) AF / MF Form error over 125 points (from 5 different probes (fixed head) or 5 different orientations (Renishaw head) JR - 2/2002