An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. inoculating loop Maryam Monajemzadeh, pathologist Children Medical Center Hospital Tehran University Of Medical Sciences

  2. Aim?

  3. Significance of Colony Counts: AHistorical Background

  4. In retrospect, • all specimens were collected by catheterization • most asymptomatic women had counts below 10 3 CFU/mL

  5. In the 1960s and 1970s • women with symptomatic, acute lower UTls (urethra, bladder, or both), 29%to 45% had colony counts lower than 105 CFU/mL by suprapubic aspiration.

  6. 1980 102 CFU/mL.

  7. growth of ≥ 105CFU of bacteria per milliliter of urine highly indicative of infection • in young, sexually active women with the acute urethral syndrome (dysuria, frequency, and urgency), as few as 102CFU/mL is considered significant in the presence of concomitant pyuria

  8. True urinary infections associated with fewer than 105CFU/mL • infants and children • in males • in persons who are catheterized • recently treated with antimicrobial agents • drink large amounts of fluids (which dilutes urine) • symptoms and concomitant pyuria • urinary obstruction • pyelonephritis acquired from hematogenous spread (especially infections due to yeast and S. aureus) • Consequently, proper interpretation of urine culture results requires communication between clinicians and laboratory personnel.

  9. Quantitative bacterial culture of a urine specimen is done by inoculating appropriate media with a measured amount of urine • most commonly with a plastic or wire calibrated loop designed to deliver a known volume. • A 0.001 mL loop is used to inoculate all urine specimens • except those collected from women with suspected acute urethral syndrome and suprapubic aspirates. • Both the latter are inoculated with a 0.01 mL loop.

  10. Although the inoculating loops are calibrated to deliver the volume of urine prescribed • accuracy has an error rate of as much as 50%, particularly when using the 0.00 l-mL loop. • Vertical sampling from a small container may deliver only 50% of the prescribed volume • horizontal sampling at a 45-degree angle from a large container may deliver 150% of volume. • should be aware of these potential errors • derive a standard angle for sampling in their laboratory, based on the volume of containers being used.

  11. Accuracy and precision studies on the volume of inoculum can be done (I) photometrically by adding a loop of gentian violet sampled from a 60-mL reservoir of dye to 2 mL of water in a cuvette and reading it in a spectrophotometer set at 590 nm, or….. (2) manometrically by noting the change in weight when a loop of water is delivered to a tilter paper disk placed in the pan of a highly sensitive analytical balance.

  12. روش اسپکتروفتومتری • در این روش از یک ماده رنگی که دارای جذب نوری در طول موج خاص است استفاده می گردد . • اوانس بلو • متیلن بلو • کریستال ویوله

  13. 20mg evans blue • 10 mg distilled water

  14. با توجه به کاهش تصاعدی رقت لوله ها از کاغذ لگاریتمی برای تهیه منحنی استاندارد استفاده میشود . • پس از ترسیم منحنی برای کالیبره کردن لوپ۱۰ لوله برداشته و در هر یک از آنها ۱ میلی لیتر آب می ریزیم • به کمک لوپ از محلول رنگی در لوله ها وارد می کنیم. • خواندن جذب نوری آنها در طول موج ۶20 نانومتر • به دست آوردن میانگین جذب نوری • انتقال آن بر روی منحنی • مقدار حجم برداشته شده توسط لوپ

  15. 1/640 1/3200