80 likes | 469 Views
ABG Instrumentation Analytical Errors. Levy-Jennings Chart In Control. +2SD. X. -2SD. Random Error. +2SD. X. -2SD. Contributing Factors Statistical probability Contamination of the sample Improper handling of the sample. Corrective Actions
E N D
Levy-Jennings ChartIn Control +2SD X -2SD
Random Error +2SD X -2SD
Contributing Factors Statistical probability Contamination of the sample Improper handling of the sample Corrective Actions Reanalyze control sample after performing a 2-point calibration If this works no further action is needed If this does not work, follow manufacturer recommendations Random Errors
Systematic or Bias ErrorTrend +2SD X -2SD
Systematic or Bias ErrorShift +2SD X -2SD
Contributing Factors Contaminated buffer solutions Incorrect gas concentrations Component degradation/ failure Incorrect procedure Corrective Actions Troubleshoot suspected problems Repair or replace problematic components Re-run 2 point calibration Bias errors