An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Degradation Monitoring - IFE/DLR Manfred W. Wuttke Envisat Calibration Review 9-13 September 2002 ESTEC, Noordwijk, The Netherlands

  2. The Monitoring Concept Monitoring the behaviour of SCIAMACHY in-orbit is an essential function during the Commissioning and Routine Operations Phase. It ensures that • the actual instrument status is known • countermeasures preventing instrument malfunctions can be initiated • data processing can incorporate the most up-to-date instrument characterisation

  3. Degradation Model (1/2) • Lightpath through instrument components is described as product of Müller matrices. • Assumption: Degradation of a component can be described as a linear factor on the corresponding Müller matrix.

  4. Degradation Model (2/2): Aging coefficients •  ASM •  ESM •  Diffuser •  Optical Bench Module to Science Detectors •  NDF •  Optical Bench Module to PMD 1-6 •  Optical Bench Module to PMD 7

  5. The m-factor concept(1/6) M-factors are used in the 0-1b processor to compensate for the radiometric degradation of SCIAMACHY. In general, a m-factor • is defined as a product of inverse aging coefficients and • can be determined as the ratio between a measured solar spectrum at the time t to a reference spectrum obtained for the same optical path at the time 0, i.e. BOL. Ideally, these reference spectra should be taken on-ground, but in practice this is not possible because measurements of the unobscured sun cannot be performed on-ground. Therefore, „earliest“ in-flight measurements shall be taken as reference.

  6. The m-factor concept (2/6) • mcal =1/ Calibration light path (via diffuser, NDF) to science detectors • mdl =1/  Limb light path to science detectors • mpl =1/  Limb light path to PMD detectors • mql =1/  Limb light path to 45° PMD detector • mdn =1/  Nadir light path to science detectors • mpn =1/  Nadir light path to PMD detectors • mqn =1/  Nadir light path to 45° PMD detector

  7. The m-factor concept(3/6) Strategy: • M-factors shall be computed on the basis of calibrated sun spectra • Calibration shall be consistent with operational L0-1b processing • Therefore tools are derived from Level 0-1b processor prototype software package

  8. The m-factor concept(4/6) Thus needed: • Continuous flow of sound Level 1b products: • To have consistent calibration data. • To monitor and extrapolate trends in instrument history • Up to date version of L0-1b processor prototype software package: • To apply consistently all corrections but the m-factors.

  9. The m-factor concept(5/6) Reference spectra have to be determined ASAP! Instrument seems to show signs of degradation already. The assumption m-factor = 1 @ “BOL“ is therefore questionable!

  10. The m-factor concept(6/6) For the ASM the aging coeffient  can be determined without reference to BOL. This will give a hint about at least the one assumption But: Calibrated spectra are needed!

  11. Conclusions(1/4) • We already see that the instrument is altering. • The basic idea of the m-factor concept, that the instrument didnt change between on ground characterization (keydata) and early inflight phase, is thus waning. • The Operational Long Term Monitoring concept depends on the level 1b product, which is not yet in a useful state (completeness, quality and supply of data).

  12. Conclusions(2/4) • Current monitoring activities thus depend heavily on • level 0 data, the supply of which is not guaranteed for long • tools for level 0 which were not foreseen to be used for • long term, • routine and • mass application.

  13. Conclusions(3/4) Under these circumstances • to get a comprehensive picture of the instrument status • as required to procure high quality data products • on adequate timescales is hard to achieve.

  14. Conclusions(4/4) • What is needed /missing: • Reliable data distribution • Sound Level 1b product • Up to date version pf L0-1 processor prototype package • What would (have) be(en) useful: • A more open policy concerning release of official software sources (extraction, calibration).