The completely redesigned Endos DC system features a new generator, tubehead, arm, and timer for maximum energy efficiency and minimal patient dose. With high frequency constant potential technology and adaptive duty cycle, this system offers precise exposure control and optimized image quality. Connectivity options and various mounting configurations make it versatile for any dental practice.
E N D
Completely new • Endos DC has been designed from scratch. • Every component of the system is completely new: • new generator • new tubehead • new arm • new timer
New generator • High frequency, constant potential technology with ripple lower than 2%. • Inverter operating at 100kHz frequency. This value is at the highest levels in the market. • The two features above help in giving the maximum energy efficiency ant to keep the patient dose as small as possible.
New generator • Duty cycle 1:15, adaptive.After a 200ms exposure, you just have to wait 3.2 seconds for the next exposure. The adaptive duty cycle, automatically sets the waiting time according to loading conditions. • Exposure parameters: 65kV, 4/5 mA with DLC technology
New generator • The choice of a single high voltage (65kV) optimizes the range of visible gray levels by operating in the optimal dynamic range of digital detectors and films • This value gives the best representation of all the interesting anatomic structures.
DLC Technology • DLC = Dual Low Current • It is possible to select two anodic current values: 4 or 5mA. • The two mA values and the 65kV high voltage have been studied to give the best results with digital sensors and high speed films • Those values allow to operate with highly stable and repeatable times (not too short, thus unstable; not too long, thus subject to patient movement)
Energetic efficiency • Endos DC has been designed to make the best use of the electric energy taken from the mains • Endos DC is already compliant to the forthcoming IEC standards on the reduction of the mains harmonics which will be in force starting from 2004
New Tubehead • Compact design • 0.7mm focal spot • weight 4.5 kg • new design • accessories: • 30cm cone • 35x45mm collimator
New timer • Selection of 4 receptors: • Standard film (5mA) • High speed film (5mA) • Digital (4mA) • Programmable (4 or 5mA) • The selection of the receptor optimizes: • Current (4 or 5 mA) • Exposure times (depending also on the selected tooth)
New timer Anodic current 4/5mA indication 30cm cone indication X-ray on indication Manual time selection “Ready” indication Patient size selection Anatomic tooth selection X-ray pushbutton Receptor selection
New timer • A control on the keyboard, with a signal lamp, can be used to automatically increase the exposure times when the 30cm extension cone is in use • Display brightness adjustment • Automatic compensation of mains fluctuations • Exposure times from 0.01 to 2 s
Remote timer • Gives complete control over the system from outside the examination room. • The same keyboard installed on the local timer can be removed and installed outside the room by means of a simple adapter kit
Remote Timer • Connection is made with a simple telephone-type connector • It’s possible to convert a local system into a remote system at any time by using an adaptation kit
Connectivity • Endos DC incorporates a RS232 serial interface which will allow in the future to communicate with external devices and perform service functions
Configurations • Mounting options: • standard wall mount with 30, 60, 80cm extension arms • wall mount with remote timer keyboard • Mobile Wall mounted units can be installed using two (single stud) or three fixation holes. Two holes out of three are compatible with Explor-X 70 holes.
Service Average installation time: 4 hours, 1 technician Tubehead replacement: 1 hour, one technician Minimum adjustments required