An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Polarization Calibration of HMI • A. Norton, J. Schou, D. Elmore, • J. Borrero, S. Tomczyk, G. Card + • High Altitude Observatory • National Center for Atmospheric Research

  2. Polarization Modulation Schemes Mod A - 4 observations mix linear & circular (80 sec) I Q U V I Q U V Mod C - 6 obs separate linear & circular I±Q, I±U, I±V (120 sec) Can correct for velocity during Mod C.

  3. Polarization Calibration Unit (PCU)G. Card, D. Elmore, S. Tomczyk, K. Streander + Linear Polarizer & a 1/4 Waveplate: *Moved in/out of beam (1.25  resolution) *Rotate optics (0.001° resolution) Inputs fixed states of polarization into HMI for calibration purposes

  4. PCU HMI Tel HMI Polarizer Sout Sin Lin 1/4 Pol WP 1/2 1/4 1/2 Lin WP WP WP Pol HMI Tel Ret (,) 17 Unknowns in optical path: 3 - Input light - Slamp= ([1.0], [q], [u], [v]) 4 - PCU - transmissions, retardation error and mount error of WP 2 - HMI tel - retardation and angle 6 - HMI Polarizer - mount errors and retardation errors for WPs, 2 - CCD - gain, bias

  5. Calculating theT-Matrix & Polarimeter Response Matrix • Calibration sequences selected to constrain unknowns: With PCU elements in/out of beam and at various angles, step through a variety of HMI waveplate orientations. Short sequence - 69 exposures, includes Mod A Long sequence - 140 exposures, includes Mod C. • Forward model light path. • Use least squares fitting routine to solve for unknowns. • Combine for Telescope matrix & Polarimeter Response Matrix.

  6. Monte Carlo Simulations - Photon Noise Calibration requirement are that uncertainties in [XT] < 0.1%

  7. Calibration in Progress Polarization data taken Feb 9-10 With PCU in beam. Work in progress. Please check HAO website for updates http://web.hao.ucar.edu/hmi 1st magnetogram 2/7/06 courtesy of J. Schou