An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Non-Invasive Oxygen Monitoring Pulse Oximetry & Conjunctival Oxygen Monitoring

  2. OBJECTIVES • At the end of this module, the student will be able to… • define terms associated with pulse oximetry. • explain the theory of operation of a pulse oximeter. • differentiate between functional and fractional saturation. • list the indications, limitations, advantages, disadvantages and contraindications for use of a pulse oximeter.

  3. OBJECTIVES • At the end of this module, the student should be able to… • list some of the things that affect the accuracy of a pulse oximeter. • explain when co-oximetry would be more appropriate than pulse oximetry. • describe patient application for each of the noninvasive monitors.

  4. ASSIGNMENTS • Read: • Egan Chapter 16, pgs. 377 – 383 • AARCClinical Practice Guidelines - Pulse Oximetry • Articles on Pulse Oximetry (look on website)

  5. Circle of Oxygenation Lungs Uptake PaO2 Dissolved Oxygen Tissues Utilization O2 Oxygen Consumption Blood Carrying SaO2 Carried Oxygen Heart Pumping DO2 Oxygen Delivery

  6. 120 100 80 60 40 20 0 Hypoxia Develops Rapidly 98 97 96 95 SaO2 % PaO2 mmHg  4 —> 2 L/min A 90 80 Central Apnea 70 Obstructive Apnea 50 30 Hypoxia + Obstructive Apnea 10 0 1 2 3 4 from Hanning,"Oximetry and Anaesthetic Practice", 1985

  7. Pulse Oximetry The Fifth Vital Sign

  8. Oxygen Content • The total amount of oxygen present in arterial blood. • Dissolved oxygen (PaO2) • Attached to hemoglobin (SaO2) • Measured by co-oximeter • Estimated by pulse oximeter • Calculated by blood gas analyzer • Actual amount of hemoglobin (Hb)

  9. Oxygen Carrying Capacity

  10. The Science behind Oximetry • Spectrophotometry • Lambert-Beer Law • Light Transmission Spectra and Oximetry • Light Emitting Diode (LED) technology • Optical Plethysmography

  11. Spectrophotometry • Spectrophotometry is the measurement of the amount of light that is absorbed as it passes through a substance. • Reflected • Absorbed • Transmitted • Spectrophotometry is an application of Beer’s (Beer-Lambert) law.

  12. Absorption Spectrum

  13. Schematic block diagram of a pulse oximeter

  14. Wavelength Selection • Two wavelengths are selected for analysis. • A red light at about 660 nanometers (nm). • An infrared light at about 940 nanometers. • So how do we get such specific measurements?

  15. Light Emitting Diode Technology • A pulse oximeter sensor is made up of two different parts: • An light emitting diode • A photodetector

  16. Misalignment of the sensor by inserting finger too far.

  17. Optical Plethysmography • Plethysmography is the measurement of changes in volume in a particular part of the body. • Originally designed to measure vascular perfusion. • Changes in volume at a site is due to the pulsation of arterial blood.

  18. Output signal generated by pulse oximeter. Saturation is based on the ratio of light absorption during pulsatile and baseline phases.

  19. Functional vs. Fractional Saturation • SpO2¹ SaO2 • Functional Saturation (SpO2) • Fractional Saturation (SaO2)

  20. Measurement vs. Monitoring • Measurement: A one time snapshot. Example: ABG sample or a “spot-check” SpO2. • Monitoring: Ongoing measurements.Example: Continuous pulse oximetry

  21. SAFE ACCURATE RELIABLE CONVENIENT Golden Rule For Safety Monitors • A Safety Monitor must be: “Always detect danger” and “No false goodness”

  22. Technical Limitations • Calibration Assumptions • Optical Interference • Absorption Interference • Signal Artifact

  23. Calibration Assumptions • Calibration curves • Accuracy range

  24. Optical Interference • Ambient Light • Optical Shunt • Optical Cross-Talk • Edema

  25. Absorption Interference • Anemia • Skin Pigmentation • Nail Polish • Carbon Monoxide • Methemoglobin • Intravascular Dyes

  26. Signal Artifact • Artifact can cause errors in measurement. • Artifact may obscure the signal • Causes: • Low perfusion • Hypotension • Cold • Motion Artifact

  27. The Solution? • Improvements in technology that: • Improve the signal to noise ratio present in low perfusion. • Improve the ability of the oximeter to “read” through motion.

  28. Sensor Application • Properly place sensor. • Single use sensor are meant for a single patient. • Use sensor as designed.

  29. Patient Application - Sites • Adults & children • Finger • Toe • Ear lobe • Cheek • Tongue • Nose • Small children & babies • Wrist • Upper arm • Across foot

  30. Pulse oximetry probes

  31. Reporting Results • Four things to verify prior to reporting results: • Waveform • Signal strength • Heart rate • Clinical Correlation

  32. Plethysmograph – Pulse oximeter tracing

  33. Oxyhemoglobin Dissociation Curve

  34. Contraindications (relative) • When arterial blood gases are indicated for measurement of pH or PaCO2. • When CO-oximetry is indicated for evaluation of total hemoglobin and abnormal hemoglobins • HbCO% • HbMet%

  35. Maintenance • Pulse oximeters usually do a self-check when they are first turned on. • They do not require regular routine manual calibration. • Wipe probe and monitor with an alcohol soaked cloth between patients.

  36. Mini Clini - Egan: Page 363

  37. Conjunctival • Principles of Operation • A sensor is inserted under the eyelid. • It contains a Clark PO2 electrode similar to a blood gas analyzer. • It is used mainly in to OR for short-term application. • Conjunctival PO2 is reported and correlates to the arterial PaO2.

  38. Conjunctival PO2(PcjO2)