An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER 12 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university

  2. CHAPTER 12 GAS CHROMATOGRAPHY (GC)

  3. GAS CHROMATOGRAPHY - Mobile phase is a gas - Column is open tubular - Sample is injected through a thin rubber disc (septum) - Sample is vaporized upon entering a heated glass port - A carrier gas (He, N2, H2) carries vaporized sample to a detector - Detector temperature is higher than column temperature to keep solutes in the gaseous state

  4. GAS CHROMATOGRAPHY Liquid Sample Size 0.1 to 2 µL for analytical chromatography 20 to 1000 µL for perspective chromatography Gas Sample Size 0.5 to 10 mL (gas tight syringe should be used)

  5. INSTRUMENTATION - Injectors (syringes, autosamplers) - The column - Detectors

  6. SAMPLE INJECTION - Sample is injected through a thin rubber disc (septum) to a heated glass port where it is vaporized - Injection is commonly done with syringes - Syringe needle must be very narrow - Wide needle could cut out a plug of the septum (coring) Autosamplers - Can analyze 100 or more liquid samples without manual injection - Samples are put in autosampler vials which are capped with septa

  7. SAMPLE INJECTION Split Injection - Good for open tubular columns - Complete injection may be too much for an open tubular column - 0.1 to 10% of the injected sample reaches the column - Not good for quantitative analysis (higher boiling point components may not be vaporized)

  8. SAMPLE INJECTION Splitless Injection - Suitable for quantitative analysis - Suitable for analysis of low concentrations of solutes (trace componenets) - Dilute solution with low-boiling solvent is used - About 80% of the injected sample reaches the column

  9. SAMPLE INJECTION Solvent Trapping - Sample is initially injected at about 40o below boiling point of sample - A thin band of solute is trapped - Column temperature is later raised

  10. SAMPLE INJECTION Cold Trapping - Used for high-boiling solutes - Sample is initially injected at about 150o below boiling point of solutes of interest - Solvent and low-boiling solutes are eluted - High-boiling solutes are trapped in a narrow band - Column temperature is later raised

  11. SAMPLE INJECTION On-column Injection - Sample is injected directly into the column - Used for compounds that decompose at temperatures above their boiling points - Solvent trapping or cold trapping is employed to trap a narrow band of analyte - Column temperature is increased afterwards to initiate chromatography

  12. OPEN TUBULAR COLUMN - Stationary phase is usually made of fused silica (SiO2) (silicone polymers) - Liquid or solid stationary phase is coated on the inner wall - Stationary phase may be porous carbon - Another type is porous layer open tubular (PLOT) for relatively small molecules

  13. OPEN TUBULAR COLUMN Molecular Sieves - Included in the stationary phase - Has cavities and made of inorganic materials - Used to dry gaseous solutes - Strongly retains H2O - Separates other small molecules (CH4, H2, O2, N2, CO2)

  14. OPEN TUBULAR COLUMN - Polar column is usually used for polar solutes - Nonpolar column is usually used for nonpolar solutes - Tailing is seen when stationary phase bakes and SiOH groups (silanol) forms on the silica surface

  15. OPEN TUBULAR COLUMN Nonpolar Stationary Phase - Solutes are eluted in order of increasing boiling point - Solutes with higher vapor pressure are eluted faster Polar Stationary Phase - Solutes are separated based on polarity - Less polar solutes are eluted faster than strongly polar solutes

  16. GUARD COLUMN - Collects nonvolatile solutes that are not eluted - Attached to the front of a chromatography column 5 to 10 meters long - Has no stationary phase and is silanized - Ends are cut off with time to discard nonvolatile solute buildup

  17. OPEN TUBULAR COLUMN VS PACKED COLUMN - Gives better separation - Narrower peaks - Handles smaller samples (analytical chromatography)

  18. COLUMN TEMPERATURE Increase in Column Temperature - Increases solute vapor pressure - Decreases retention time - Results in sharp peaks Temperature Programming - Used to separate compounds with a wide range of boiling points and polarities

  19. ELUTION AND RESOLUTION Elution and resolution behavior depends on - The composition of the stationary phase and mobile phase gas - Column dimensions (length) - Column phase ratio (internal diameter and film thickness) - Column gas flow (pressure) - Column temperature

  20. DETECTORS Mass Spectrometer (GC-MS) - Very sensitive and the most versatile - Major problem is interfacing Flame Ionization Detector (FID) - Destructive, mass-flow detector - For almost all organic compounds (not for CO, CO2, HCN) - Low detection limits Thermal Conductivity Detectors (TCD) - Nondestructive, concentration detector - Not sensitive to narrow columns (diameter < 0.53 mm)

  21. DETECTORS Electron Capture Detector (ECD) - Very sensitive to halogen-containing compounds - Insensitive to ketones, alcohols, and HCs - Nondestructive, mass-flow detector Electrolytic Conductivity Detector (ELCD) - For organic compounds with halogen, N, or S substituent - Destructive, Mass-flow detector Sulfur Chemiluminescence Detector (SCD) - For organic compounds with S atoms only - Destructive, mass-flow detector

  22. DETECTORS Sulfur – Phosphorus Flame Photometric Detector (SP-FPD) - For organic compounds with S or P atoms - Destructive, mass-flow detector Nitrogen – Phosphorus Detector (NPD) - Selectively sensitive to nitrogen and phosphorus - Used for analysis of drugs - Destructive, mass-flow detector Photoionization Detector (PID) - For organic compounds with more easily ionizable π-electrons - Nondestructive, mass-flow detector

  23. DETECTORS Helium Ionization Detector (HID) - Universal detector for everything except neon - Nondestructive, mass-flow detector Atomic Emission Detector (AED) - Selective for compounds containing many atoms - Tunable - Destructive, mass-flow detector

  24. HYPHENATED TECHNIQUES GC-MS GC-IR GC-GC or GC2 (2D-Gas Chromatography)

  25. APPLICATIONS - Environmental pollution monitoring - Analysis of contaminants - Biochemical, medical, and pharmaceutical research

  26. LIMITATIONS - Complex molecules (DNA, RNA) are too large to be volatilized and may decompose - HPLC is preferred for analyzing such molecules - Not suitable for solutions in aqueous media