An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Cardiac Case Study Inferior wall attenuation and correction with CT-based Attenuation Correction Linda Campbell, BS GE Applications Scientist

  2. Type of acquisition Time per stop Agent Dose Rest Ungated SPECT; CT 25 seconds 99m Tc Sestamibi 15 mCi Stress Gated SPECT; CT 25 seconds 99m Tc Sestamibi 30 mCi Patient History A 68 year old male was referred for myocardial perfusion imaging three months status post revascularization of the right coronary artery for evaluation of patency. Some atypical chest pain was noted. Imaging Procedure: The one day Sestamibi protocol was used, and rest imaging was performed first. SPECT imaging was done with 60 stops over 180 degrees. In addition, CT images were acquired. Acquisitions were done on a GE Hawkeye. Target heart rate (85% of maximum heartrate): 129 Maximal heart rate achieved on treadmill: 90%

  3. Imaging Analysis As part of the acquisition quality control for this data, fused SPECT-CT images were checked for proper registration of attenuation maps and emission data. SPECT-CTsagittal fusion SPECT-CT transaxial fusion

  4. Further quality control checks showed the acquisition to be of good quality: • The stress and rest selective linograms showed no significant motion in the vertical direction and showed no significant extra-cardiac activity that would interfere with the reconstruction. • The stress and rest sinograms showed no significant motion in the horizontal direction • Myocardial count density was adequate. The stress maximum pixel value was 201; rest maximal pixel value was 65. The guideline for an adequate count study for 60-64 projection images is to have a maximum pixel value of at least 50 counts within a tight circular/ elliptical ROI around the epicardial edge of the myocardium on an anterior view of the heart. • There were no gating errors

  5. Question 1 – Acquisition quality controlWhich one of the following acquisition factors has the least effect on slice image quality? A. Horizontal motion B. Myocardial count density C. Gating errors D. Vertical motion E. Intense extra-cardiac activity adjacent to the heart

  6. The correct answer is C - Gating errors Thickening is the parameter most prone to error in the presence of arrhythmias. Slice appearance is maintained in the presence of arrhythmias except in very severe cases. LV ejection fraction (EF) is a robust parameter that is little affected by arrhythmias. Quantitative perfusion measures were also little affected by arrhythmias except in the case of atrial fibrillation.1 As has been well documented in the literature, vertical and horizontal motion and myocardial count density are critical for slice image quality. Intense visceral activity close to the heart can also affect slice image quality as it results in changed slice count densities.

  7. The rotating projection images show soft tissue attenuation in the waist area which can cause increased attenuation of the inferior wall. Click Click Stress projection images Rest projection images Soft tissue attenuation may create artifacts that mimic true perfusion abnormalities and can lead to false positive interpretation.

  8. Question 2 – Attenuation correction What are the advantages of doing attenuation correction on cardiac studies? • Provide images that are more uniform and easier to interpret • Decrease acquisition time • Increase myocardial count density • Increase sensitivity and overall diagnostic accuracy • Both A and D

  9. The correct answer is E – Both A and D Attenuation correction can provide images that are more uniform and easier to interpret, and increase sensitivity and overall diagnostic accuracy. The diagnostic accuracy of SPECT imaging is compromised by soft tissue attenuation artifacts. Though education, experience, and the application of gated SPECT imaging have greatly helped, there is a need for improved diagnostic accuracy. Attenuation correction SPECT techniques represent a significant advance in myocardial perfusion imaging and hold great promise for improved assessment of cardiac patients.2

  10. Question 3 – CT-based attenuation correction What are the further advantages of doing CT-based attenuation correction on cardiac studies? A. CT-based AC produces high quality transmission maps B.CT-based AC is consistent with time C. CT-basedAC is consistent over a full range of body types D. All of the above

  11. The correct answer is D – all of the above There are significant differences in the ability of different commercial AC systems to reduce artifacts due to attenuation, hot liver / bowel activity and type of orbit. Systems that can generate high quality attenuation maps yielded the best results.A high quality attenuation map appears to be an important determinant of image quality.3 Stress attenuation map Rest attenuation map

  12. Stress and rest images were reconstructed by filtered back projection (FBP) and iterative reconstruction with attenuation correction (IR-AC). Count deficiencies in the inferior walls of both the stress and rest FBP slices have been noticeably improved in the IR-AC slices, correcting for inferior wall attenuation. FBP IR-AC Stress SA slices Rest SA slices FBP IR-AC Stress VLA slices Rest VLA slices

  13. EKG Results Non-specific ECG changes were noted with no ST depression at maximal exertion. Test terminated at maximal heart rate and SOB. No chest pain was noted. Findings FBP images were interpreted as inferior wall scar with possible ischemic component. This could also be a diaphragmatic attenuation artifact. With CT-based AC the inferior wall is shown to be viable with no ischemia. The validity of the revascularization of the RCA is confirmed. For more information on GE Nuclear Cardiology products and the unique Hawkeye CT based cardiac AC, visit:GE Hawkeye Cardiac Attenuation Correction Findings and Images Courtesy of VA Medical Center, Baltimore, Maryland

  14. References 1. Nichols K, Yao S, Kamran M, Faber TL, Cooke, CD, and DePuey EG. Clinical impact of arrhythmias on gated SPECT cardiac myocardial perfusion and function assessment. J Nucl Cardiol 2001; 8:19-30. 2. Hendel RC, Corbett JR, Cullom SJ, DePuey EG, Garcia EV, and Bateman TM. The Value and Practice of Attenuation Correction for Myocardial Perfusion SPECT Imaging: A Joint Position Statement form the American Society of Nuclear Cardiology and the Society of Nuclear Medicine. J Nucl Med 2002; 43:273-280 3. O’Connor MK, Kemp B, Anstett F, Christian P, Ficaro EP, Frey E, Jacobs M, Kritzman JN, Pooley RA, and Wilk M . A multicenter evaluation of commercial attenuation compensation techniques in cardiac SPECT using phantom models. J Nucl Cardiol 2002; 9:361-76.