110 likes | 255 Views
GIFTS System Calibration Requirements Flowdown. Hank Revercomb 9 March 2004. GIFTS Overall Absolute Calibration Budget. Subject of BB CDR. GIFTS Top-level Absolute Calibration Budget. GIFTS Spectral Calibration Budget. Standard Atmosphere used for conversion to brightness temperature.
E N D
GIFTS System Calibration Requirements Flowdown Hank Revercomb 9 March 2004
GIFTS Overall Absolute Calibration Budget Subject of BB CDR
GIFTS Spectral Calibration Budget Standard Atmosphereused for conversion tobrightness temperature ** **3-sigma for single sample is 3 ppm-100 sample average assumed
Radiance N derived from raw spectra of Earth (CE), Space (CS), and the internal Hot (CH) and Cold (CC) Blackbodies t is the signal transmission of the telescope mirrors &m is the transmission of the Blackbody pick-off mirror Bi is the Radiance from the Hot, Cold, and Space References [eg. BH=H B(TH) + (1- H) B(Tstr), where B(T) is the Planck function and Tstr is the effective structure temperature for reflection] GIFTS Radiometric Calibration Equation:Basis for Uncertainty Modeling
GIFTS Absolute Calibration-Longwave Longwave Band (800 cm-1) Tc=255, Th=290, Ts=240, Tt=280 Requirement External BB GIFTS (excluding uncertainty in correction forNon-linearity & Polarization)
GIFTS Absolute Calibration-Shortwave Shortwave Band (1800 cm-1) Tc=255, Th=290, Ts=240, Tt=280 Requirement External BB GIFTS (excluding uncertainty in correction forNon-linearity & Polarization)
GIFTS Calibration Error Sources-LW Longwave Band (800 cm-1) Tc=255, Th=290, Ts=240, Tt=280 0.3% 0.3 K NEN/4=0.05
GIFTS Calibration Error Sources-SW Shortwave Band (1800 cm-1) Tc=255, Th=290, Ts=240, Tt=280 0.5% 0.3 K NEN/4=0.0025
GIFTS Blackbody Budget Subject of this BB Subsystem CDR:Budget is 0.5 K 3-sigma Absolute & 0.1 K 3-sigma Reproducibility, by analogy