Michigan’s Standard of Care Patient Exposure Values. Bruce Matkovich Radiation Safety Section Michigan Department of Community Health. Exposure at Skin Entrance. Measure common radiographic projections during regulatory inspection: Common patient thicknesses
E N D
Michigan’s Standard of Care Patient Exposure Values Bruce Matkovich Radiation Safety Section Michigan Department of Community Health CRCPD 38th Annual Conference - Detroit
Exposure at Skin Entrance • Measure common radiographic projections during regulatory inspection: • Common patient thicknesses • Measured with MDH X-Ray Monitor • Monthly workloads collected CRCPD 38th Annual Conference - Detroit
Exposure at Skin Entrance - 2 • Exposure data entered into Radiation Safety Data System • Recent update distinguishes between digital modes Computed Radiography (CR) and Direct Radiography (DR) CRCPD 38th Annual Conference - Detroit
Standard of Care Exposure Values • Established by advisory committee near the 85th percentile of ESE data • Last adjustment in 1995 • Mostly based on 200 speed screen-film • Non-regulatory CRCPD 38th Annual Conference - Detroit
Standard of Care Exposure Values CRCPD 38th Annual Conference - Detroit
Standard of Care Exposure Values and AAPM Reference Values CRCPD 38th Annual Conference - Detroit
Exposures Reported to Facility • Measured Entrance at Skin Exposure along with the Standard of Care for that projection • Advised to reduce exposure if over the Standard of Care • Good cooperation from facilities in reducing exposures when appropriate CRCPD 38th Annual Conference - Detroit
Exposure Distribution Graphs • Most recent 5 year period • Screen-film and digital imaging displayed separately • Number of machines tested, median exposure, and 85th percentile data included CRCPD 38th Annual Conference - Detroit
Exposure Trends • The difference between the 85th percentile and the median exposure is being reduced • “Outriders” are being becoming less of a problem CRCPD 38th Annual Conference - Detroit
Trend - AP Lumbar Spine CRCPD 38th Annual Conference - Detroit
Trend - AP Cervical Spine CRCPD 38th Annual Conference - Detroit
Trend - PA Chest (without a grid) CRCPD 38th Annual Conference - Detroit
Trend - PA Chest (with a grid) CRCPD 38th Annual Conference - Detroit
Trend - DP Foot CRCPD 38th Annual Conference - Detroit
Trend - Hand CRCPD 38th Annual Conference - Detroit
Trend – Dental Cephalometric CRCPD 38th Annual Conference - Detroit
Standard of Care Conclusions • Standard of Care values or reference values are effective in controlling exposures • They do not adversely drive exposures below what is necessary for diagnostic quality images CRCPD 38th Annual Conference - Detroit
For more information: Radiation Safety Section www.michigan.gov/rss CRCPD 38th Annual Conference - Detroit