ADC by SunTech Adview features • Overview of the ADC by SunTech Adview • Taking an automated BP measurement • Taking a manual BP measurement • Taking a temperature measurement • Taking an SpO2 measurement • Recalling the last reading • Clearing the last reading • Need help? Contents
Overview of the ADC by SunTech Adview BP Module Pressure column 1 bar = 10mmHg Systolic BP/ Cuff pressure Diastolic BP Automatic button Manual button Pulse rate icon Warning icon MAP icon Pulse rate/ Mean Arterial Pressure (MAP) Cuff/Hose Error icon Contents
Overview of the fully configured ADC by SunTech Adview Temperature Mode button Temperature Module BP Module Signal quality SpO2 Module Retention clip Contents
Taking a BP Measurement Press the auto button to take an automatic measurement OR Press and hold the manual button to inflate the cuff, and take a traditional manual reading with a stethoscope Contents
Taking an automated BP measurement • Wrap cuff snugly around upper arm. Click here to visit the ADC website for further details on proper placement. • Ask the patient to stay still and quiet, press and release the automatic button to start an automatic measurement. • A measurement is displayed in about 35-40 seconds Contents
Taking a manual BP measurement • Wrap cuff snugly around upper arm and position your stethoscope. • Press and hold the manual button to inflate the cuff to the desired pressure. • Watch the pressure displayed and quickly release the button to deflate the cuff at 3mmHg/sec. • When the measurement is determined, press the auto button to completely deflate the cuff. Contents
Taking a temperature measurement • Place a disposable cover on the blue probe; the display shows the type of measurement, (oral). • Wait for the device to beep. • Then, place the probe under the patient’s tongue; the posterior medial sublingual pocket is preferred for accuracy. • A “traveling dash” displays for 10-15 seconds until a reading is displayed. Contents
Taking an SpO2 measurement For the reusable finger sensor, insert the patient’s digit into the sensor; a “traveling dash” displays for 10-20 seconds until a reading is displayed. The reading is updated every second. Contents
Recalling the last reading Note: Measurements are displayed for 2 minutes after which the display shows- - - Press and hold the Automatic BP button for 2 - 3 seconds to display the last measurement values. Contents
Clearing the last reading Note: If there are privacy concerns with regards to measurements on the display, you can clear the display. Once cleared, they cannot be recalled. Press and hold the Automatic BP button for 5 - 6 seconds to clear the last measurement values. Contents
Need help? Contact us by telephone: • 1.800.232.2670 or 1.631.273.9600 Or by email: • CustomerService@adctoday.com Or find help on the Internet: • www.adctoday.com Contents