1 / 25

‘Point-of-Care’ Coagulation Monitors

‘Point-of-Care’ Coagulation Monitors . Evan Pivalizza, M.D. October, 2005. A. Anticoagulation (heparin) management. Function – whole blood clotting Quantitative- heparin level. ACT (Activated clotting time) Hemachron 2 ml blood + activator (celite → kaolin)

brooke
Download Presentation

‘Point-of-Care’ Coagulation Monitors

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ‘Point-of-Care’ Coagulation Monitors Evan Pivalizza, M.D. October, 2005

  2. A. Anticoagulation (heparin) management • Function – whole blood clotting • Quantitative- heparin level

  3. ACT (Activated clotting time) • Hemachron • 2 ml blood + activator (celite → kaolin) • Magnet in rotating test tube (dependent) • Blood clots → magnet moves/breaks contact → time • Normal 80-120 secs • Aim > 480 secs (animal, ltd human data)

  4. B. Hemotec • Cartridge kaolin • Blood automatically added → plunger • Clot impedes speed plunger → time

  5. Advantages • Easy, inexpensive • Sizable body data

  6. Disadvantages • 4% coefficient variability normal • Exaggerated hypothermia, hemodilutiuon • Affected thrombocytopenia, antiplatelet drugs • NOT reflect heparin level/concentration, e.g., ‘Heparin resistance’ • ↓ AT III

  7. 2. Heparin concentration A. Hepcon (Heparin management system – HMS) • Automated protamine titration (multiple channels) • < protamine = not antagonize heparin • > protamine = inhibits clot formation • √ protamine = fastest clot → heparin []

  8. Advantages • Not affected temperature, dilution • Disadvantages • NOT measure function (heparin resistance)

  9. B. High dose Thrombin Time (HiTT) • TT = time fibrinogen → fibrin by thrombin • Sensitive to low levels heparin • High dose thrombin added, bind heparin-ATIII complexes • Remaining unbound heparin ↑ time fibrin formation (measured)

  10. Advantages • Not affected temp, dilution, aprotinin • Disadvantages • Variable

  11. B. Platelet Function • Point-of care monitors • Rapid results • Help predict bleeding/cause • NOT always correlate laboratory analyses • NOT exclusive, ideally incorporated into transfusion algorithm • Clinical • Point of care • Laboratory

  12. Dynamic tests platelet function • Thrombelastograph (TEGTM), Haemoscope, IL

  13. Viscoelastic test of fibrin formation → fibrinolysis • Incorporation ↓ bleeding/ transfusion/ re-exploration • Cardiac surgery (esp complex) • Pedi cardiac surgery • Liver transplantation

  14. R: Time initial fibrin formation • K: Subsequent fibrin- plt interaction • Angle: Ditto • MA: 60% plt function (40% fibrinogen) • G: Manipulation MA - > plt function • Index: Summary first 4 measures • Lysis: At 30 minutes

  15. Modifications • Heparinase • Undercover defect during CPB • Is heparin reversed? • Tissue factor • ++ acceleration thrombin formation → MA (plt function) • Abciximab • GpIIb/IIIa antagonist • Differentiate platelet/ fibrinogen effect

  16. 2. Sonoclot (Sienco, CO)

  17. Changing impedance ultrasonic probe immersed sample • Start fibrin formation (ACT) • Fibrin cross-linkage (Rate = slope curve) • Platelet-mediated clot strength (Peak impedance and time to peak) • Data primarily cardiac

  18. B. Tests platelet response to activator • Platelet function analyzer (PFA – Dade Behring)

  19. Plt adhesive activity whole blood • Perfused across aperture in membrane coated agonist (epinephrine, ADP) • Plt adhesion seals aperture → closure time • High negative predictive value bleeding

  20. 2. Ultegra (Accumetrics, CA)

  21. Measure platelet response to TRAP (thrombin receptor agonist peptide) • Platelet-agglutination of fibrinogen-coated beads – optical detection • Sensitive GpIIb/IIIa – cath lab

  22. 3. Platelet aggregometry (lab gold std) • Photo-optical measure light transmittance through plt-rich plasma • Exposed agonist (ADP, collagen, epinephrine) → aggregation → ↓ turbidity → ↑ transmittance

  23. References • Shore-Lesserson, Anesthesiology Clinics N America 2003:21;511-26 • Shore-Lesserson, J Cardiothoracic and Vascular Anesthesia 2002:16;99-106

More Related