Determination of trace impurities on Si wafers with x-ray fluorescence

Determination of trace impurities on Si wafers with x-ray fluorescence. Seminar Author: Bojan Hiti Mentor: doc. dr. Matjaž Kavčič. Si wafers – key resource in semiconductor production Contaminants: 3d transition metals (Fe, Ni, Zn, Cu) Light metals (Al, Na )

cameo
Download Presentation

Determination of trace impurities on Si wafers with x-ray fluorescence

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Determination of trace impurities on Si wafers with x-ray fluorescence Seminar Author: Bojan Hiti Mentor: doc. dr. Matjaž Kavčič

  2. Si wafers – key resource in semiconductor production • Contaminants: • 3d transition metals (Fe, Ni, Zn, Cu) • Light metals (Al, Na) Max. contamination: 109 atoms/cm2 ~ 1fg/g • The only nondestructive analytical tehnique with required sensitivity is X-ray fluorescence spectroscopy

  3. X-ray fluorescence spectroscopy • Excitation of inner electron shells with x-ray photons • X-ray tube • synchrotron radiation • Detection of the emitted • x-ray fluorescence. • Analytical tehnique used to determine elemental composition of the sample • Consists of:

  4. X-ray interaction with matter • Photoeffect on atomic inner shells • A core-hole is filled by a higher-energy-level electron • Exceeding energy is taken over by • Secondary X-ray • (X-ray fluorescence - XRF) • Auger electron

  5. Characteristic x-ray emission line energies – element identification

  6. Detector energy resolution should be high enough to distinguish between characteristic x-ray energies of neighboring elements. Semiconductor x-ray detector

  7. X-ray interaction with matter • Resonant x-ray Raman scattering (RRS) • Significant only at energies just below absorption edge • Intermediate virtual state • Secondary X-ray is emitted • Elastic scattering • Inelastic (Compton) scattering

  8. X-ray – Siinteraction cross-sections:

  9. Total reflection XRF • For x-rays: nmatter<1 • High yield (n°counts/s) due to detector proximity • Only first few nm of the sample are excited

  10. Highest sensitivity is obtained using excitation with synchrotron radiation • high photon flux ~ 1012 photons/sec • monochromatic (DE/E ~ 10-4) and polarized light Stanford Synchrotron Radiation Lightsource Total reflection XRF setup at SSRL synchrotron

  11. TXRF spectrum of “clean” Si wafer measured at SSRL • Detection limits: • 3d metals <109 atoms/cm2 Signal from Fe and Cu surface contamination is clearly observed

  12. Measurement of Al contamination • EAl < Eexcitation < Esi • Low energy Si tail despite fine energy tuning – RRS • Detection limit: • Alslightly below 1010 atoms/cm2

  13. Grazing emission XRF • Surface sensitivity achieved by grazing detection angle • Smaller yield than Total-reflection tehnique

  14. Grazing emission detection can be combined with high-resolution x-ray spectroscopy : • High energy resolution crystal spectrometer employing Bragg diffraction on the analyzer crystal • Position • sensitive • detector Energy resolution ~1eV • Analyzing crystal

  15. Experimental Grazing-emission setup at ESRF synchrotron in Grenoble European Synchrotron Radiation Facility Bragg crystalspectrometer installed at the ID21 beamline.

  16. High resolution enables discerning between Al and Si Raman scattering signal • Theoretically lower detection limits can be achieved • Problem: small detection efficiency • Achieved detection limit: 1012 atoms/cm2

  17. Conclusion • Can provide solution for the light metals measurement • Detection efficiency is substantialy smaller than in TXRF method • Main tehnique for trace impurity detection • Very good performance for transition metals measurement • Cannot lower detection limit for light metals

  18. Thank you for your attention.