140 likes | 251 Views
TCDI Setup and validation. TCDI Setup. Changed by hand the beta functions in the TRIM application collimator positions moved to 5 s and position thresholds to ±0.25 s (dump) and ±0.125 s ( warning) automatically. TCDI Setup.
E N D
TCDI Setup • Changed by hand the beta functions in the TRIM application collimator positions moved to 5 s and position thresholds to ±0.25 s(dump) and ±0.125 s (warning) automatically.
TCDI Setup • Changed by hand the beta functions in the TRIM application collimator positions moved to 5 s and position thresholds to ±0.25 s (dump) and ±0.125 s (warning) automatically. • Energy thresholds set up by hand to 10 s + 1 s (equivalent in mm)
Calibration BLM • Close 1 jaw to 1 nominal snom shot the beam onto the collimator record losses at the near by BLM. 1 s Beam axis
Calibration • Close 1 jaw to 1 nominal snom shot the beam onto the collimator record losses at the near by BLM. • Normalize BLM reading with respect to the extracted beam intensity • Define real beam size sreal(measured emittance, dispersion, nominal Dp/p = 1e-4) • Define the fraction of beam intercepted by the collimator jaw (NORMDIST(snom,beam centre, sreal) )
Validation BLM • Knob producing n soscillation in H and V plane for different phases 0-30-60…-360° (the knobs had to be regenerated on the spot because of the change in the optics) 5 s Nominal beam axis 5 s
Validation • Knob producing n s oscillation in H and V plane for different phases 0-30-60…-360° (the knobs had to be regenerated on the spot because of the change in the optics) • Oscillations with 5 snom amplitude look at BLM readings normalize with respect to the extracted intensity • Convert BLM readings in fraction of beam intercepted • Convert fraction of beam intercepted in amplitude cut in sreal by the collimator jaws (to be added to the amplitude of the applied oscillation) TCDI aperture in sreal
Validation Results Knobs and validation fine for TI2 horizontal and TI8 vertical Should measure ~3 points per phase (4,4.5,5 σ) – here we aimed only at the 5 σ oscillation due to time constraints and previous validation
Validation Results • Some problems with: • TI2 vertical knob: 10 s amplitude needed to have ~ 5 mm oscillation (1 s = 0.9 mm) in the TL Recalculated amplitudes
Validation Results • Some problems with: • TI2 vertical knob: 10 s amplitude needed to have ~ 5 mm oscillation (1 s = 0.9 mm) in the TL • TI8 horizontal knob: 1.5-2 s amplitude needed to have ~ 5 mm oscillation in the TL Recalculated amplitudes
Validation Results • Some problems with: • TI2 vertical knob: 10 s amplitude needed to have ~ 5 mm oscillation (1 s = 0.9 mm) in the TL • TI8 horizontal knob: 1.5-2 s amplitude needed to have ~ 5 mm oscillation in the TL • For 5 mm oscillation (seen in YASP) we see a loss level at the TCDIs as expected from the ‘good’ knobs validation OK • In principle the validation was redundant since we already measured that the TCDIs at 5 ssettings provide the required protection
General Critical Issues • TCDIs are the only protection in case of large trajectory drifts • TCDI setup and validation takes minimum 2-3 shifts • Knob generation is critical and should never be done on the spot • Two people are needed for both procedures • At least two teams covering 8h shifts are needed or at least 8h machine downtime has to be accepted between setup and validation • Automatize as much as possible both procedures • Automatics setup tool will be tested in MD on Wednesday