An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Lecture 10

  2. The time-dependent transport equation Spatial photon gradient Absorbed photons Photons scattered to direction ŝ' Light source q Photons scattered into direction ŝ from ŝ'

  3. Time-Dependent Transport Equation • Typically the transport equation is expressed in terms of the radiance ( I(r,ŝ,t) =N(r,ŝ,t)hnc) , and after dropping the integrals

  4. Time-Independent Transport Equation • For the steady-state situation, we assume that radiance is independent of time, and the transport equation becomes

  5. Approximations • The transport equation is difficult to solve analytically.In order to find an analytical solution we need to simplify the problem. • Discretization methods • Discrete ordinates method • Kubelka-Monk theory • Adding-doubling method • Expansion methods • Diffusion theory • Probabilistic methods • Monte Carlo simulations

  6. Diffusion approximation • Expand the photon distribution in an isotropic and a gradient part • Where r(r,t) is the photon density • And J(r,t) is the photon current density (photon flux)

  7. Fick’s 1st law of diffusion Movement or flux in response to a concentration gradient in a medium with diffusivity c Photon flux (J cm-2 s-1) in response to a photon density gradient, characterized by the diffusion coefficient D, defined as

  8. Diffusion approximation Transport equation: Photon distribution expansion: Diffuse intensity is greater in the direction of net flux flow Photon source expansion:

  9. Diffusion approximation • Plug in, integrate over w, and assume only isotropic sources (refer to supplementary material for full derivation) • Assume a constant D and use the relation for the fluence rate • To get

  10. Types of diffuse reflectance measurements frequency domain (TD) Continuous wave (CW) Time domain (TD) I0 I0 It It intensity intensity t (ns) tissue t=0 ~ns phase shift intensity ac dc t (ns)

  11. Point source solution: time-domain • The solution to the diffusion equation for an infinite homogeneous slab with a short pulse isotropic point source S(r,t)=d(0,0) is • This is known as the Green’s function solution and can be used to solve more complicated problems

  12. Point source solution: frequency domain • Harmonic time dependence is given by factor exp(-iwt), so that ∂/∂t -iw • Diffusion equation takes the form

  13. Point source solution: frequency domain frequency domain (TD) • Green’s function for homogeneous, infinite medium containing a harmonically modulated point source of power P(w) at r=0 is intensity t (ns) tissue phase shift intensity ac intercept (ms’) t (ns) dc phase ln(r2*Idc) slope (ma, ms’) slope (ma, ms’) r r intercept = 0

  14. Frequency domain measurements • The slope of r*IDC as a function of r and the slope of the phase as a function of r depend on ma and ms'. • Find the slopes and extract the optical properties

  15. Medical applications of reflectance spectroscopy Pulse Oximetry Frequency domain NIR spectroscopy and imaging Steady-state diffuse reflectance spectroscopy

  16. The Pulse oximeter • Function: Measure arterial blood saturation • Advantages: • Non-invasive • Highly portable • Continuous monitoring • Cheap • Reliable

  17. The pulse oximeter • How: • Illuminate tissue at 2 wavelengths straddling isosbestic point (eg. 650 and 805 nm) • Isosbestic point: wavelength where Hb and HbO2 spectra cross. • Detect signal transmitted through finger • Isolate varying signal due to pulsatile flow (arterial blood) • Assume detected signal is proportional to absorption coefficient (Two measurements, two unknowns) • Calibrate instrument by correlating detected signal to arterial saturation measurements from blood samples

  18. The pulse oximeter • Limitations: • Reliable when O2 saturation above 70% • Not very reliable when flow slows down • Can be affected by motion artifacts and room light variations • Doesn’t provide tissue oxygenation levels

  19. Near-infrared spectroscopyand imaging of tissue Sergio Fantini Department of Biomedical Engineering Tufts University, Medford, MA

  20. volume probed by near-infrared photons outline source Near-infrared spectroscopy and imaging of tissues applications to skeletal muscles • hemoglobin oxygenation (absolute) • hemoglobin concentration (absolute) • blood flow and oxygen consumption applications to the human breast • detection of breast cancer • spectral characterization of tumors applications to the human brain • optical monitoring of cortical activation • intrinsic optical signals from the brain detector source detector source detector

  21. Why near-infrared spectroscopyand imaging of tissues? Non-invasive Non-ionizing Real-time monitoring Portable systems Cost effective

  22. Dominant tissue chromophoresin the near infrared ultraviolet near infrared visible light 770 nm 410 nm 600 1300 absorption coefficient (cm-1) wavelength (nm) Hb, HbO2 from: Cheong et al., IEEE J. Quantum Electron. 26, 2166 (1990) H2O from: Hale and Querry, Appl. Opt. 12, 555 (1973)

  23. Diffusion of near-infrared light inside tissues low power laser optical fiber optical detector biological tissue

  24. high scattering problem is there a car in front of me? is there a cookie in the milk?

  25. Frequency-domain spectroscopy (FD) intensity (a.u.) t (ns) tissue phase shift intensity (a.u.) ac dc t (ns)

  26. Diffusion equation: frequency domain • Harmonic time dependence is given by factor exp(-iwt), so that ∂/∂t -iw • Diffusion equation takes the form

  27. Point source solution: frequency domain • Green’s function for homogeneous, infinite medium containing a harmonically modulated point source of power P(w) at r=0 is frequency domain (TD) intensity t (ns) tissue phase shift intensity ac intercept (ms’) t (ns) dc phase slope (ma, ms’) ln(r*Idc) slope (ma, ms’) r r intercept = 0

  28. TISSUE OXIMETRY

  29. Time-domain oximetryMiwa et al., Proc. SPIE 2389, 142 (1995)

  30. 0 5 1 0 2 0 2 5 3 0 Configuration for tissue oximetry ) % ( n o i t a r u t a S n i b o l g o m e H e ( m i n ) T i m RF electronics detector optical fiber detector source optical fibers laser driver 2.0 cm laser diodes multiplexing circuit measuring probe main box

  31. Frequency-domain oximetry HbO2and Hb (mM) saturation (%) ischemia ischemia time (min) time (min) 750nm 830nm ms’ (1/cm) ma (1/cm) 750nm 830nm ischemia ischemia time (min) time (min)