210 likes | 610 Views
ICD Implant Overview. Today's ICDs are implanted in much the same way a pacemaker is implantedTransvenous leadsPectoral implantSometimes done on outpatient basisMay require only conscious sedationDevice-based testing may be done to test defibrillation thresholds (DFTs). Before the Implant. Inte
E N D
1. The Electrical Management of Cardiac Rhythm DisordersTachycardiaImplant Procedures The Electrical Management of Cardiac Rhythm Disorders, Tachycardia, Slide Presentation 20 Implant ProceduresThe Electrical Management of Cardiac Rhythm Disorders, Tachycardia, Slide Presentation 20 Implant Procedures
2. ICD Implant Overview Today’s ICDs are implanted in much the same way a pacemaker is implanted
Transvenous leads
Pectoral implant
Sometimes done on outpatient basis
May require only conscious sedation
Device-based testing may be done to test defibrillation thresholds (DFTs)
3. Before the Implant Interrogate the device
Be sure to verify that battery voltage > 3.2 V
Check charge times
Perform a capacitor maintenance charge
If desired, enter patient data into device
ICD will store patient’s name, diagnosis, device and lead information
Very valuable if patient is seen by other physicians, travels a lot, may be treated out-of-clinic
Defibrillation is OFF in the device as shipped; verify this for the safety of the implant team!
4. Basic Preparations Heart monitor
Pulse oximeter
Oxygen delivery system
Blood pressure monitor
External defibrillation equipment
Anesthesia equipment The stylized ECG says help. It is a reminder that there should always be external defibrillation equipment and trained personnel on hand during an ICD implant.The stylized ECG says help. It is a reminder that there should always be external defibrillation equipment and trained personnel on hand during an ICD implant.
5. Markers of High DFTs Although it is impossible to accurately predict DFTs prior to testing, there are indicators or markers that suggest high DFTs
Cardiomyopathy
Low ejection fraction (if known)
Antiarrhythmic agents
High NYHA Class
Prior myocardial infarction (MI)
High DFTs can (and do) occur in patients without these markers, too!
6. Patient’s Diagnosis Is this ICD for primary prevention?
Patient is at high risk for SCD but has no documented history of arrhythmia
This does not affect the implant procedure but will affect programming
Primary-prevention therapy is typically one-zone (VF only)
Is this ICD for secondary prevention?
Patient has a history of arrhythmias
This does not affect the implant, but does affect programming
Need zones to accommodate known rhythm disorders
Does the patient have a standard pacing indication?
Again, affects how the device will be programmed
7. Single-Chamber versus Dual-Chamber A single-chamber system requires one lead (RV)
RV lead both senses/paces and shocks
A dual-chamber system requires two leads (RA and RV)
RA lead senses/paces only
RV lead both senses/paces and shocks
Other tools
Introducers (based on size of leads and techniques)
Guide wires, stylets
Pacing System Analyzer (PSA)
Measures mode, rate, pacing parameters perioperatively
8. During the Implant Leads are implanted first
Single-stick or dual-stick technique
Cephalic cutdown or subclavian stick are most common approaches
Venous access may be more lateral in ICD implantation than for pacemaker implantation
Typically, the ventricular lead is placed first
RV apex
RV outflow tract
9. Thresholds and Impedance Once the leads are in place, thresholds should be tested
Lead impedance values should be evaluated (and fall within the specified range, ideally in the middle)
10. Securing the Leads First, verify appropriate sensing and pacing
Suture the lead(s) in place using the suture sleeve
Plug the lead pins into the epoxy header (the device indicates which lead goes into which port)
Assure electrical connection by making sure the lead is properly in place (set-screw)
11. Implant Testing Implant testing should be done with the actual device
ICD should be in the pocket
Requires a patient care system
Use wireless telemetry (if available)
Or put wand in sterile bag for introduction into the field
Obtain a real-time EGM
This shows what the ICD “sees” from signals picked up by the leads
Look for these landmarks
Strong, clear intrinsic signals
Possible far-field sensing
12. Perioperative Real-Time EGM
13. Real-Time Measurements Obtain real-time measurements from the device
Shows what the device is actually doing (not what it is programmed to do)
Battery voltage (should have a BOL value)
Signal amplitudes
Should be large enough for proper sensing
Verify sensitivity settings correspond to these signal sizes
Pacing lead impedance
Should fall within the broad range specified by the manufacturer
Values should be near middle of range (not at either extreme)
14. High-Voltage Lead Integrity Check Verifies proper lead connection and set up (without delivering a high-energy shock)
12 V through RV lead
Most patients will feel this and perceive it as painful
May require anesthesia
Warn the patient before conducting this test
ICD can be programmed to do daily HV lead integrity checks automatically using a multivector system
15. HV Lead Integrity Check
16. DFT Testing DFT is the lowest amount of energy required to reliably defibrillate the heart
DFTs vary by patient and will vary in one patient over time
Conventional safety margin is 10 J
Patient is tested to have DFT of 13 J
Program device to 23 J (13 + 10 J safety margin)
DFT testing requires VF induction and therapy delivery from the device
Device-based testing can induce the arrhythmia
External equipment can also be used
17. Device-Based Arrhythmia Induction Device-based testing (DBT) allows
Burst pacing (programmed stimulation protocol)
Shock-on-T
DC Fibber
DC Fibber (also known as Ventricular Fibber) uses 12 V of direct current
Works 96.4% of the time on the first application
Will not awaken a patient out of conscious sedation
Easy to set up and activate
Once arrhythmia is induced, use device to detect, diagnose, and deliver therapy
18. DFT Testing Always have external defibrillation equipment available and ready to use during DFT testing
Make sure trained personnel are there to defibrillate externally, if required
Always alert the team before arrhythmia induction and DFT testing
Once test is over, check diagnostic data to confirm proper function (can take a few minutes for data to be available)
19. DFT Testing Considerations DFT testing is painful; make sure the patient is properly sedated or anesthetized
Confirm that the ICD is appropriately programmed
Ideally, set up two shocks
One lower-energy, the second maximum energy
During DFT testing, watch detection and therapy delivery on the programmer
If you need to do DFT testing again, allow at least three minutes
Give “stunned” tissue time to recover
20. Implant Troubleshooting: Impedance High intraoperative lead impedance
Suggests lead is not properly connected
Verify leads are plugged in and set-screws are tightened
Make sure leads are in the right ports
Low intraoperative lead impedance
Suggests that there is an “open circuit”
May be a lead insulation break
Check insulation for nicks or cuts
Look for possible lead damage (particularly if passing two leads)
21. Implant Troubleshooting: High DFTs High DFTs
Use programmable fixed pulse width
Use programmable tilt
Reverse the polarity
Change the shocking vector
Reposition the lead
Playing the odds—DFT problems are more likely to be solved with
Programmable fixed pulse width (rather than tilt)
RV anode (rather than RV cathode)
22. Conclusion ICD implants today
Pectoral implants with transvenous leads
Often done under conscious sedation
May be performed on out-patient basis
Steps
Venous access
Lead placement/testing
Capture and sensing thresholds
Device testing
High DFTs can be a challenge
Close pocket/final programming