1 / 23

Standard Operating Procedures

Standard Operating Procedures. At www.dep.state.fl.us/labs/qa/sops.htm Call for help at 407-893-3313 Orlando Office 850-488-2796 Tallahassee. Chapters of Interest. FS2400 Wastewater Sampling FT1000 Field Testing General FT1100 Field pH FT1400 Temperature FT1500 Field Dissolved Oxygen

chen
Download Presentation

Standard Operating Procedures

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Standard Operating Procedures • At www.dep.state.fl.us/labs/qa/sops.htm • Call for help at • 407-893-3313 Orlando Office • 850-488-2796 Tallahassee

  2. Chapters of Interest • FS2400 Wastewater Sampling • FT1000 Field Testing General • FT1100 Field pH • FT1400 Temperature • FT1500 Field Dissolved Oxygen • FT1600 Field Turbidity • FT1900 Continuous Monitoring • FT2000 Field Chlorine (TRC)

  3. Calibration Field pHFT1100 • Quarterly • And the good news, THERE ARE NO QUARTERLY CALIBRATION REQUIREMENTS FOR pH • Monthly • and for more good news, THERE ARE NO MONTHLY CALIBRATION REQUIREMENTS FOR pH

  4. Calibration of pH meter • Weekly - Cannot be done with all meters, calibrate and check meter’s % slope. • Should be 90% or greater • Record date, time , analyst, ph buffer readings after calibration, buffer lot # , buffer expiration date, temperature, and slope

  5. Daily Calibration • Calibrate with at least 2 buffers • Use 7 buffer first then 4 or 10 • Second buffer will bracken most samples • For meters using 2 point calibration, use third buffer as check

  6. Field Calibration • In the field, as minimum, check meter with 7 buffer. • Record readings • Return from the field as minimum, check meter with 7 buffer • No more than 0.2 pH units off from standard

  7. Continuous pH meter calibration • Perform an initial calibration • before installation • whenever taken off-line • after failed QC check • Daily checks • compare pH of grab sample taken at same location as in-line monitor • grab sample pH measured with calibrated meter

  8. Document, Document, Document • If its not written down you did not do it! • Date and time • Analyst • Reading from continuous meter • Reading from calibrated meter • No more than 0.2 pH units different than calibrated meter reading

  9. TemperatureFT1400 • Variety of thermometers for field • capable of reading 0.1 degree Celsius • Laboratory thermometer • NIST-traceable reading 0.1 • degree Celsius

  10. Temperature Calibration • Temperature should agree within 0.1 degree Celsius • Keep records of checks and sample measurements

  11. TurbidityFT1600 • Meter specifications are in EPA Method 180.1 or FT1600 • Daily Checks • Quarterly Checks

  12. Quarterly Calibration • Used to validate the secondary standards • Use properly calibrated nephelometer • Use primary standard for each range of secondary standard • Refer to FT1600 3.2 for acceptable limits for secondary standards • This calibrates the secondary standards

  13. Daily Calibration • Used for assurance meter is operating properly • Check each range with primary or secondary standard • Refer to FT1600 3.2 for acceptable limits for secondary standards

  14. Record Keeping • Type of standards and concentrations • Date, Time, Analyst • result of each check • Lot # and expiration date for all bought standards • Lot # and expiration date for • all bought reagents

  15. IN-line Turbidity Calibration • Initial calibration • Before installation • when taken off-line • when fails QC criteria • Daily Check • grab sample, measure with calibrated meter and compare

  16. In-line Turbidity • write it down!! • date, time, analyst • reading from in-line meter and other meter • should be no more than 20 % difference • use caution with very low readings

  17. Total Residual ChlorineFT2000 • DEP recommends using DPD colorimetric method for concentrations between 0.2 and 4.0 mg/L • DEP does not recommend equipment • Can’t use gel stds for <0.5 requirements • Color wheel comparison kits are unacceptable

  18. Quarterly Calibration • Used to calibrate/check the meter • Meter is checked with primary standards • At least one standard for each range • Refer to FT2011 for acceptable limits for primary standards • Can be performed by a laboratory certified in residual chlorine or Class A or B operator

  19. Quarterly Paperwork • If you took the time to do it, be able to prove it!!! • Date, time, analyst • Lot # and expiration date of chemicals used • True value of standard and meter reading

  20. Monthly Calibration • Used to validate the secondary standards • Use properly calibrated meter • Use primary standard for each range of secondary standard • Refer to FT2011 for acceptable limits for secondary standards • This validates the secondary standards • Can be performed by a laboratory certified in residual chlorine or Class A or B operator

  21. Monthly paperwork • when you write it down you make me so happy!!! • Date, time, analyst • Lot # and expiration date of standards used • True value of standard and meter reading

  22. Daily Calibration • Commercial secondary standards or primary standards • Zero with blank • Read the other three standards • Date, time, analyst • Lot # and expiration date of standards used • True value of standard and meter reading

  23. In-Line Calibration • Calibrate • initially • when taken off-line • when QC checks failed • Check against calibrated meter daily • should be no more than 20 % difference

More Related