1 / 18

ABX Pentra 400 Specifications Reagents Main features analysis Strengths

ABX Pentra 400 Specifications Reagents Main features analysis Strengths.  Sample Tray.  Reagent Tray.  ISE Module. Physical Specifications.  Cuvette changer. . . . .  Windows XP. . Physical Specifications. Bench-top analyzer Dimensions : 100 x 65 x 57 cm

chi
Download Presentation

ABX Pentra 400 Specifications Reagents Main features analysis Strengths

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ABX Pentra 400 • Specifications • Reagents • Main features analysis • Strengths

  2.  Sample Tray  Reagent Tray  ISE Module Physical Specifications  Cuvette changer      Windows XP 

  3. Physical Specifications Bench-top analyzer Dimensions : 100 x 65 x 57 cm Weight : 120 kg Screen: 12 inches 57 cm 65 cm 100 cm

  4. Technical Specifications Analysis Method Types - Spectrophotometry : Colorimetry and Turbidimetry, - Potentiometry : Direct (Serum or Plasma) and Indirect (Urine) On-board Test Capacity : 55 parameters on board - 52 cassette positions or 39 tests positions with Reagent racks - 3 ISE's (Na+, K+, Cl-)

  5. Technical Specifications • External Modules • Cooling unit • Options • ISE Module • P400 Stand Alone ISE Module

  6. Technical Specifications Throughput : - Without ISE : up to 300 Tests per hour - With ISE : up to 420 Tests per hour

  7. Technical Specifications • Sample Handling • Sample Capacity • - 60 samples on board + Continuous loading • - 6 racks with 10 sample positions for primary tubes or cups Positive Sample Identification Integrated Barcode Readerand tube presence detector Sample Needle - Level detection - Shock detection - Clot detection

  8. Technical Specifications • Sample Handling Sample Specimen Serum, Plasma, Urine, CSF, Whole blood, Homogeneous fluids Sample Volume - Photometry mode:2 to 95 µL (380 in 4 steps) - ISE mode : 65µL Automatic Sample Dilution from 1/2 to 1/22 500 Automatic Post Concentration from x2.0 to x10.0

  9. Technical Specifications • Calibrator and Control Handling Calibrator/Controls Positioning Sample Tray or Reagent Tray Automatic Standard Dilution Series - Dilution ratio:main direct : 1/2 - 1/150 - Dilution ratio: main indirect : 1/2 - 1/1508

  10. Technical Specifications • Reagent Handling Number of Reagents on-board : 52 positions - 44 in a closed refrigerated area (2-10°C) - 8 at room temperature Positive Reagent Identification Integrated barcode reader Reagent Needle - Capacitive Level detection - Preheating

  11. Technical Specifications • Reagent Handling Reagent Management with Cassettes - Automated Reagent On board stability check plus test countdown - Unique cassette barcode - back-up cassette Reagent Volume - Minimum : 2µL with Sample needle 15µL with Reagent needle - Maximum : 600µL

  12. Technical Specifications • Reaction Handling Reaction Cuvettes - Disposable acrylic cuvettes - Cuvette volume : Min 150µl - Max 600µl Automatic Cuvette Changer - Continuous loading and unloadind of cuvette segments - Total Capacity : 36 segments (432 cuvettes)

  13. Technical Specifications • Reaction Handling Reaction Temperature Temperature controlled Air bath at 37°C Mixing By Rotation Incubation Time 1 to 100 cycles (12 seconds to 20 minutes)

  14. Specifications Reagents Cassette Features • - 4 types of cassette (1 single and 3 twin) • - Side barcode indentifies for : • - type of reagent • - expiration date • - filling volume • - index of the cassette • - scrambling key • - Cassette Packaging • - On board Reagent Stability

  15. Specifications Reagents Advantages of the cassette • - Easy to handle • - Automatic Reagent management • - Lot number • - Open stability handled • - Test countdown • - Back-up cassettes

  16. ABX Pentra Range Characteristics • 52 reagents (30 in cassette format) • 9 Enzymes • 21 Substrates • - 22 Specific Proteins

  17. Compact Benchtop Integrated System 52 refrigerated positions Comprehensive menu + Open channels Bar coded RTU Cassette 3 ISE’s : Na, K, Cl High Autonomy Optical system Holographic grating technology Choice of 15 wavelengths from 340-700 nm Safe and secure environment Sampling constant check GLP/Logs Reagent management Appealing Easy to use software Integrated Validation Station Help on line Windows XP touch screen Main Features analysis

  18. Strengths • High autonomy Throughput Cuvette capacity Sample and reagent on-board capacity • Comprehensive menu Routine and Proteins • Reagent management RTU cassettes, easy handling Higher reagent stability Test countdown, lot and shelf life Cassette / vial flexibility • Easy to use and Intuitive software Touch screen interface Integrated validation station Remote diagnostic capabilities Contextual help • Flexibility • Closed system with number of • open channels # configurable • Cal / Ctrl placement on system • Most liquid chemistries adaptable

More Related