1 / 20

Fat/Water Separation

Fat/Water Separation. Hochong Wu 2008/01/25. In vivo Fat signal. Fat (lipids, CH 2 ) freq shift of -3.5 ppm -3.5x10 -6 x 42.576x10 6 Hz/T x 1.5 T ≈ -223 Hz Fat signal is only second to water Chemical-shift artifacts Obscure tissue contrast. Reeder SB, et al., MRM 2004. Fat Suppression.

christmas
Download Presentation

Fat/Water Separation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Fat/Water Separation Hochong Wu 2008/01/25

  2. In vivo Fat signal • Fat (lipids, CH2) freq shift of -3.5 ppm -3.5x10-6 x 42.576x106 Hz/T x 1.5 T ≈ -223 Hz • Fat signal is only second to water • Chemical-shift artifacts • Obscure tissue contrast

  3. Reeder SB, et al., MRM 2004

  4. Fat Suppression • Inversion-Recovery (Fat T1 ≈ 250 ms at 1.5 T) • Fat Saturation: Excite fat and dephase it • Spectral-Spatial: Excite water only, then image • Limitations to these methods • Instead, decompose an MR image into a Fat image and a Water image

  5. Chemical Shift Imaging MR Signal Equation: m(x,y) actually contains many freq components: Want to make an image of each f of interest. Fully resolve the frequency dimension --> MR Spectroscopic Imaging

  6. Fat/Water Separation Two Component Model: Suppose we acquire M images of m(x,y) at different TEk: k = 0 … M-1 Then we can solve for W and F. TE can be varied for Gradient-Echo and Spin-Echo sequences.

  7. Two-Pt. Dixon • Number of pts. means the number of images acquired at different TE (M) • F complex, W complex --> M = 2 to solve (0, π) acquisition Dixon WT, Radiology 1984

  8. Dixon WT, Radiology 1984

  9. Three-Pt. Dixon • Field inhomogeneity • Need more than 2 measurements (0, π, 2π) acquisition (0, π, -π) acquisition Glover GH, JMRI 1991

  10. SNR Efficiency • Effective NSA = var(est) / var(source) • Relative SNR = sqrt( NSA / M ) Glover GH, JMRI 1991

  11. (0, π, -π) acquisition (0, π, 2π) acquisition Glover GH, JMRI 1991

  12. Extended 3-pt. (0, π, 2π) Dixon Glover GH, JMRI 1991

  13. Considerations • Field Map • Phase fitting, unwrapping, processing • Scan Time • Proportional to M • Misregistration among sources

  14. Four-Pt. Dixon • And even more points … • General estimation problem • Model • Spectral Peaks • Spectral Widths (T2) • Amplitude loss (T2’) • Field inhomogeneity

  15. Other Methods • Single-point Dixon • Direct Phase Encoding (generalized 3 pt.) • CSISM • IDEAL • Many more!

  16. IDEAL • Iterative Decomposition of Water and Fat With Echo Asymmetry and Least-Squares Estimation • Arbitrary TE shifts • Can be asymmetric • Least-squares estimation problem • Solve iteratively • Support multiple coils Reeder SB, et al., MRM 2004 Reeder SB, et al., MRM 2005

  17. IDEAL Removing effects of the field ψ: Least-squares estimate: Iteratively estimate ψ

  18. Source Calculated Water Calculated Fat Reeder SB, et al., MRM 2004

  19. Complementary Techniques • Multi-echo, RARE • Fast readout trajectories • Partial Fourier • Parallel Imaging

  20. Discussion • Fat needs your attention ! • Fat/Water Separation – important case of Chemical-Shift (Spectroscopic) Imaging • With more sophisticated model  MR Spectroscopy