1 / 13

Simulator Theory and Operation

Simulator Theory and Operation. Use and Theory. EACH BREATH TEST MUST BE PRECEEDED BY A REFERENCE ANALYSIS REQUIRED BY THE CDPHE RULES AND REGULATIONS, 5 CRR 1005-2 THE DEVICE USED TO GENERATE THE REFERENCE VAPOR IS CALLED BREATH ALCOHOL SIMULATOR OR “SIMULATOR”

cianna
Download Presentation

Simulator Theory and Operation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Simulator Theory and Operation Rev 3.09

  2. Use and Theory • EACH BREATH TEST MUST BE PRECEEDED BY A REFERENCE ANALYSIS • REQUIRED BY THE CDPHE RULES AND REGULATIONS, 5 CRR 1005-2 • THE DEVICE USED TO GENERATE THE REFERENCE VAPOR IS CALLED BREATH ALCOHOL SIMULATOR OR “SIMULATOR” • H2O AND ETOH IS USED IN LUE OF BLOOD AND ETOH Rev 3.09

  3. Use and Theory • THE AMOUNT OF ETOH IN 2100 ml OF DEEP LUNG AIR AT 34.0C  ETOHIN1 ml OF ATERIAL BLOOD • SIMULATOR SOLUTION 0.100 grams OF ETOH PER 2100 Liters OF AIR AT 34.0C • PREPARED BY CERTIFICATION UNIT • VALIDATES THE PERFORMANCE OF THE 5000EN Rev 3.09

  4. Functional Structure AIR INLET TUBE HIGH INTENSITY L.E.D. DISPLAY ON/OFF-RESET SWITCH HEATER INDICATOR GLASS CONTAINER VAPOR PORT AGITATOR BAFFLE PLATES TEMPERATURE SENSOR HEATING ELEMENT DISPERSION TUBE Rev 3.09

  5. Exclusive Features of the Guth Model 2100 Simulator • MICROPROCESSOR ENHANCED: -PERFORMSINTERNAL DIAGNOSTICS -MONITERS TEMPERATURE SENSOR, RFI CIRCUITRY -CONTROLS AGITATOR MOTOR, DISPLAY DRIVER, COMMUNICATIONS AND A/V INFORMATION Rev 3.09

  6. Exclusive Features of the Guth Model 2100 Simulator • TEMPERATURE SENSOR: -RUGGED DESIGN: STAINLESS STEEL, BRASS -STABLE TEMPERATURE RANGE:MAINTAINS A PRECISE RANGE OF 34C  .02C -ACCURACY OF INSTRUMENT TEMPERATURE READING IS .05C Rev 3.09

  7. Exclusive Features of the Guth Model 2100 Simulator • ELECTRONICALLY PROTECTED: -INTERNAL POWER TRANSFORMER AND OPTICALLY ISOLATED CONTROL COMPONENTS PREVENTS A.C. LEAKAGE TO ELECTRONICS -R.F.I. PROTECTED -HEATING ELEMENT PROTECTED FROM OVER HEATING WHEN TEMPERATURE SENS0R IS DEFECTIVE OR REMOVED FROM SOLUTION -EXTERNAL FUSE PROVIDES PROTECTION IN CASE OF ELECTRICAL MALFUNCTION Rev 3.09

  8. Guth Simulator Error Codes MESSAGE PLATE MOUNTED ON SIMULATOR QUICK REFERENCE GUIDE Rev 3.09

  9. Guth Simulator Error Codes * NOTIFY CDPHE Rev 3.09

  10. Guth Simulator Error Codes * NOTIFY CDPHE Rev 3.09

  11. Guth Simulator Error Codes * NOTIFY CDPHE Rev 3.09

  12. Correct Positioning of Simulator Tubing and Clamp #1 • #1 Shows the correct orientation of the Simulator Tubing Elbow • #2 Shows the correct positioning of the Simulator Clamp • Make sure the clamp is tightened all the way closed • When sending the simulator in for repairs, remove the tubing at the clamp and • disconnect the heated tube at the disconnect joint • Keep all tubing and the jar at your location #2 Rev 3.09

  13. QUESTIONS? Rev 3.09

More Related