130 likes | 302 Views
Simulator Theory and Operation. Use and Theory. EACH BREATH TEST MUST BE PRECEEDED BY A REFERENCE ANALYSIS REQUIRED BY THE CDPHE RULES AND REGULATIONS, 5 CRR 1005-2 THE DEVICE USED TO GENERATE THE REFERENCE VAPOR IS CALLED BREATH ALCOHOL SIMULATOR OR “SIMULATOR”
E N D
Simulator Theory and Operation Rev 3.09
Use and Theory • EACH BREATH TEST MUST BE PRECEEDED BY A REFERENCE ANALYSIS • REQUIRED BY THE CDPHE RULES AND REGULATIONS, 5 CRR 1005-2 • THE DEVICE USED TO GENERATE THE REFERENCE VAPOR IS CALLED BREATH ALCOHOL SIMULATOR OR “SIMULATOR” • H2O AND ETOH IS USED IN LUE OF BLOOD AND ETOH Rev 3.09
Use and Theory • THE AMOUNT OF ETOH IN 2100 ml OF DEEP LUNG AIR AT 34.0C ETOHIN1 ml OF ATERIAL BLOOD • SIMULATOR SOLUTION 0.100 grams OF ETOH PER 2100 Liters OF AIR AT 34.0C • PREPARED BY CERTIFICATION UNIT • VALIDATES THE PERFORMANCE OF THE 5000EN Rev 3.09
Functional Structure AIR INLET TUBE HIGH INTENSITY L.E.D. DISPLAY ON/OFF-RESET SWITCH HEATER INDICATOR GLASS CONTAINER VAPOR PORT AGITATOR BAFFLE PLATES TEMPERATURE SENSOR HEATING ELEMENT DISPERSION TUBE Rev 3.09
Exclusive Features of the Guth Model 2100 Simulator • MICROPROCESSOR ENHANCED: -PERFORMSINTERNAL DIAGNOSTICS -MONITERS TEMPERATURE SENSOR, RFI CIRCUITRY -CONTROLS AGITATOR MOTOR, DISPLAY DRIVER, COMMUNICATIONS AND A/V INFORMATION Rev 3.09
Exclusive Features of the Guth Model 2100 Simulator • TEMPERATURE SENSOR: -RUGGED DESIGN: STAINLESS STEEL, BRASS -STABLE TEMPERATURE RANGE:MAINTAINS A PRECISE RANGE OF 34C .02C -ACCURACY OF INSTRUMENT TEMPERATURE READING IS .05C Rev 3.09
Exclusive Features of the Guth Model 2100 Simulator • ELECTRONICALLY PROTECTED: -INTERNAL POWER TRANSFORMER AND OPTICALLY ISOLATED CONTROL COMPONENTS PREVENTS A.C. LEAKAGE TO ELECTRONICS -R.F.I. PROTECTED -HEATING ELEMENT PROTECTED FROM OVER HEATING WHEN TEMPERATURE SENS0R IS DEFECTIVE OR REMOVED FROM SOLUTION -EXTERNAL FUSE PROVIDES PROTECTION IN CASE OF ELECTRICAL MALFUNCTION Rev 3.09
Guth Simulator Error Codes MESSAGE PLATE MOUNTED ON SIMULATOR QUICK REFERENCE GUIDE Rev 3.09
Guth Simulator Error Codes * NOTIFY CDPHE Rev 3.09
Guth Simulator Error Codes * NOTIFY CDPHE Rev 3.09
Guth Simulator Error Codes * NOTIFY CDPHE Rev 3.09
Correct Positioning of Simulator Tubing and Clamp #1 • #1 Shows the correct orientation of the Simulator Tubing Elbow • #2 Shows the correct positioning of the Simulator Clamp • Make sure the clamp is tightened all the way closed • When sending the simulator in for repairs, remove the tubing at the clamp and • disconnect the heated tube at the disconnect joint • Keep all tubing and the jar at your location #2 Rev 3.09
QUESTIONS? Rev 3.09