1 / 12

Chapter 19 Screen-Film Image Artifacts

Chapter 19 Screen-Film Image Artifacts. Usually TECHNOLOGIST dependant Patient Preparation Clothes left on (zippers, jewelry, dentures Positioning errors (patient, tube, bucky , receptor) Wrong screen-film combination (detail screens with fast film and vice-versa) “ Bling ” left on.

ciqala
Download Presentation

Chapter 19 Screen-Film Image Artifacts

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Chapter 19 Screen-FilmImage Artifacts • Usually TECHNOLOGIST dependant • Patient Preparation • Clothes left on (zippers, jewelry, dentures • Positioning errors (patient, tube, bucky, receptor) • Wrong screen-film combination (detail screens with fast film and vice-versa) • “Bling” left on

  2. Processing Artifacts • Roller marks; guide shoe lines, Pi lines • Dirty rollers (curtain effect/run down, etc.) • Chemical fog • Wet-pressure sensitization (varied amount of pressure on the rollers = uneven film “splotches”)

  3. Handling Artifacts • Light/radiation fog • Kink marks • Static (tree, smudge are the most common) • Hypo retention (fixer retention – required for mammo)

  4. Chapter 20 Film-Screen Quality Control • Quality Assurance is a JCAHO required 10 step PEOPLE PROCESS! • Quality Control = imaging instrumentation and equipment process

  5. QC = 3 step process • Acceptance testing • Routine performance monitoring • Maintenance

  6. Radiography QC • Filtration – annually • Collimation – semiannually (+/- 2% SID) • F.S. size – Annually (+/-50%) FREQUENTLY MISSED PARAMETER! • kVp calibration – Annually (+/- 10%) • Exp. Timer accuracy – annually (based on mS used – more leeway with longer exposures) • Linearity – annually (+/- 10%) of same mAs with DIFFERING mA/times • Exp. REPRODUCEIBILITY – SAME contrast/density, etc with SAME exp. factors repeated later on

  7. Focal Spot Size Check Tools • Pin hole camera (somewhat hard to use?) • Star pattern (down to .3mm sizes) • Slit camera – preferred method

  8. Other RAD QC • Filtration • Collimation (PBL) • kVp/Timer • Exposure Linearity (same mAs with VARIED mA and times) • Intensifying screens • Protective apparel (annually) • Film illuminators

  9. Fluoro QC • Exposure rate (max table top dose/min) • Spot film exposures • Automatic Exposure System (ABS/ABC – AEC)

  10. Tomo QC • Tube travel/ ARC • Fulcrum level • Patient exposure

  11. Processor QC • See Table 20-6 for QC RAD Processor items/intervals • Processor Maintenance • Scheduled maintenance • Preventative maintenance • Nonscheduled maintenance

  12. Processor Monitoring • Temperatures • Replenishment • Sensitometry/Densitometry

More Related