Stony Brook Update Bill and Ted’s Cosmic Adventure. T.K. Hemmick for the Wyld Stallions. Cosmic Studies are Finished. Ted has now finished all his Cosmic data taking. This includes first measurements of gain on all stacks at dV=475 (running simultaneously).
E N D
Stony Brook UpdateBill and Ted’s Cosmic Adventure T.K. Hemmick for the Wyld Stallions
Cosmic Studies are Finished. • Ted has now finished all his Cosmic data taking. • This includes first measurements of gain on all stacks at dV=475 (running simultaneously). • Also includes 2nd measurements for outliers Excellent!!!
Scintillation used to get Gain • Fit to Gaussian + Exponential. • Gain from: • Use ADC gain from manual (pC/channel) • 2000 e’s/mV from pre-Amp • Checks with Bob to ~10%
Bill S Preston, esq. • The left column is the gain at dV=475. • The blue box shows the required dV to get gains of 2000, 4000, 6000, 8000, 10000 • The green box is the chain V. • Blue & Green are calculated (not measured). • See below for technique…
Ted Theodore Logan • The left column is the gain at dV=475. • The blue box shows the required dV to get gains of 2000, 4000, 6000, 8000, 10000 • The green box is the chain V. • Blue & Green are calculated (not measured). • See below for technique…
How to get Gain Curves • 3 “Gain-Slope” Determinations: • Vlad’s Cosmics = 0.0695 • Tom’s 55Fe spectrum = 0.0624 • Bob’s Scintillation Cube Measurement = 0.0672 • All yield same answers to w/in a few volts dV
Remaining Topics • When to move (Thursday Dec 4 preferred) • Scintillation Cube Spectra. • Trip Strategy w/ 2 power supplies. West East