An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Establishing Rules for: Medical Device Reports (803) & Correction and Removal Reports (806) Presented by: Cap Uldriks, FDA Daniel P. Olivier, CCS

  2. Objectives • Define regulatory reporting rules based on established standards • Simplify regulatory reporting decision making • Establish defensible basis to support compliance with reporting regulations • Reduce risk of non-compliance observations

  3. AGENDA • Reporting Regulations • Evaluating Safety Risks • Defining the Strategy • Summary

  4. + Reporting Regulations

  5. MDR reportable event means: (1) User facilities become aware of information that reasonably suggests a device has or may have caused or contributed to a death or serious injury; or (2) Manufacturers or importers become aware of information that reasonably suggests one of their marketed devices: (i) May have caused or contributed to a death or serious injury, or (ii) Has malfunctioned and that the device or a similar device marketed by the manufacturer or importer would be likely to cause or contribute to a death or serious injury if the malfunction were to recur. 21 CFR Part 803: Medical Device Reporting

  6. If the event is “likely to cause or contribute to death or serious injury" then the event is reportable The preamble offers the following guidance, i.e., a malfunction report is required when: 1.The chance of a death or serious injury occurring as a result of the recurrence of the malfunction is not remote 2.The consequences of the malfunction affect the device in a catastrophic manner that may lead to a death or serious injury (Class I Recall) Compliance Program Guidance for 21 CFR Part 803:Medical Device Reporting

  7. Sec. 806.10 Reports of corrections and removals. (a) Each device manufacturer or importer shall submit a written report to FDA of any correction or removal of a device initiated by them: (1) To reduce a risk to health posed by the device; or (2) To remedy a violation of the act caused by the device which may present a risk to health unless the information has already been provided under Part 803 or Part 1004 or the action is exempt from the reporting under 806.1(b). 21 CFR Part 806: Corrections and Removal Reporting

  8. Risk to Health - The key concept to determine whether an event is reportable under Part 806: Use of, or exposure to, the product may cause temporary or medically reversible adverse health consequences, or an outcome where the probability of serious adverse health consequences is remote (Class II Recall). Compliance Program Guidance for 21 CFR Part 806: FDA Corrections and Removals Reporting

  9. Class I is a situation in which there is reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death (criteria for 803 reporting) Class II is a situation in which there is reasonable probability that the use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote (criteria for 806 reporting) Class III is a situation in which there is reasonable probability that the use of, or exposure to, a violative product is not likely to cause serious adverse health consequences 21 CFR Part 7.3: Recall Class Definitions

  10. + Evaluating Safety Risk

  11. Severity* - Assess the worst possible consequence, defined by degree of injury or illness that could occur. Severity classification identified as follows: Major, may cause death or serious injury Moderate, may cause non-serious injury Minor, not expected to cause injury Risk Evaluation, Severity *Based on “Levels of Concern” defined in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, Food and Drug Administration, May 11,2005.

  12. Severity: Suggested mishap categories: Catastrophic, could result in death, permanent total disability, loss exceeding $1M Critical, could result in permanent partial disability, or that may result in hospitalization of at least three people, $200k-$1M loss Marginal, could result in injury resulting in more than one lost work day, $10K-$200K loss Negligible, could result in injury resulting in less than one lost day, less than $10K MIL-STD 882D* Severity Levels * Standard Practice for System Safety, MIL-STD-882D, Department of Defense, February 10, 2000, Appendix A, Table A-1, page 18

  13. Severity: levels may be descriptive, e.g. Catastrophic Serious Critical Marginal Negligible ISO 14971:2000* Severity Levels * Medical device - Application of safety risk management to medical devices, ANSI/AAMI/ISO 14971:2000section E.2.2, page 32

  14. Consolidated Severity Levels MIL-STD 882D ISO 14971:2000 FDA Guidance Catastrophic Catastrophic Major Serious Major Critical Critical Major Marginal Marginal Moderate Negligible Negligible Minor

  15. Probability* - Probability that a mishap will occur during the planned life expectancy of the system: Frequent - Likely to occur often (X > 10-1) Probable - Will occur several times in the life of the system. (10-1 >X >10-2) Occasional - Likely to occur sometime in the life of the system. (10-2 >X >10-3) Remote - Unlikely but possible to occur in the life of the system. (10-3 >X >10-6 ) Improbable - So unlikely, it can be assumed occurrence may not be experienced (X < 10-6) MIL-STD-882D* Risk Evaluation, Probability * Standard Practice for System Safety, MIL-STD-882D, Department of Defense, February 10, 2000, Appendix A, Table A-2, page 19

  16. Probability - Probability estimation examines the initiating events or circumstances and the sequences of events that are of concern: Frequent Probable Occasional Remote Improbable Incredible ISO 14971:2000* Risk Evaluation, Probability * Medical device - Application of safety risk management to medical devices, ANSI/AAMI/ISO 14971:2000section E.2.1, page 31

  17. Proposed Risk Assessment Matrix for FDA Adverse Event Reporting Minor Moderate Major Frequent Probable Occasional Remote Improbable Note: Probability level of “Incredible” from ISO 14971:2003 not included

  18. Defining the Strategy

  19. If the event is “likely to cause or contribute to death or serious injury" then the event is reportable The preamble offers the following guidance, i.e., a malfunction report is required when: 1.The chance of a death or serious injury occurring as a result of the recurrence of the malfunction is not remote; 2.The consequences of the malfunction affect the device in a catastrophic manner that may lead to a death or serious injury Compliance Program Guidance for 21 CFR Part 803:Medical Device Reporting

  20. Proposed Risk Assessment Matrix for FDA Adverse Event Reporting Minor Moderate Major Frequent Probable Occasional Remote Improbable MDR MDR MDR Note: Probability level of “Incredible” from ISO 14971:2003 not included

  21. Reports are NOT required IF: The information has already been reported to FDA under the MDR regulation, 21 CFR Part 803 or under 21 CFR 1004 (Replacement of Electronic Products). The key concept for determining when an event is reportable is the CAR regulation's definition of risk to health: Use of, or exposure to, the product may cause temporary or medically reversible adverse health consequences, or an outcome where the probability of serious adverse health consequences is remote (Class II Recall). Compliance Program Guidance for 21 CFR Part 806: FDA Corrections and Removals Reporting

  22. Proposed Risk Assessment Matrix for FDA Adverse Event Reporting Minor Moderate Major Frequent Probable Occasional Remote Improbable CAR CAR CAR NR NR MDR MDR MDR CAR NR NR NR NR NR NR NR - Not reportable CAR - Reportable under 21 CFR Part 806 MDR - reportable under 21 CFR Part 803

  23. Summary

  24. Reference to external standards/guidelines provides defensibility for decision making Established rules provide: consistency use of common terminology facilitate training Established rules reduce compliance risk Defining Reporting Strategy

  25. Medical device - Application of safety risk management to medical devices, ANSI/AAMI/ISO 14971:2000. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, FDA, Office of Device Evaluation, May 11, 2005. Standard Practice for System Safety, MIL-STD-882D, Department of Defense, February 10, 2000. 21 CFR Part 7 Enforcement Policy, US Federal Register, Food and Drug Administration. References