250 likes | 623 Views
Hg Monitoring Options and Requirements. Region 7 Kansas City August 2007 Charles Frushour USEPA, CAMD. Hg Monitoring Options . CEMS Sorbent Trap – Appendix K Hg LME . Hg CEMS. Initial Certification 7-day calibration error test Linearity Check (elemental Hg)
E N D
Hg Monitoring Options and Requirements Region 7 Kansas City August 2007 Charles Frushour USEPA, CAMD
Hg Monitoring Options • CEMS • Sorbent Trap – Appendix K • Hg LME
Hg CEMS • Initial Certification • 7-day calibration error test • Linearity Check (elemental Hg) • 3-level System Integrity Check (oxidized Hg)monitors with Hg converter only • Cycle Time Test • RATA • Bias Test
Hg CEMS • On-going Quality Assurance • Daily calibration error test (elemental or oxidized Hg) • Weekly 1-level System Integrity Check • Required when daily calibrations use elemental Hg for monitors with Hg converters • Weekly = 168 operating hours • Quarterly Linearity Check (elemental Hg) or3-level System Integrity Check (oxidized Hg) • Required for each QA operating quarter • Annual RATA (no semi-annual threshold) • Bias Test
Calibration Tests • The initial 7-day calibration error test and subsequent Daily calibrations may be conducted using either elemental Hg or a NIST-traceable source of oxidized Hg. • Specification: ≤ 5.0 % of span value with alt spec of ≤ 1.0 µg/scm absolute difference • The form of Hg used for daily calibrations will need to be reported in the Monitoring Plan so that EPA can track whether Weekly System Integrity Checks are required. • The monitoring plan may be updated as needed
Weekly System Integrity Check • One point version of the 3-level System Integrity Check required for initial certification of Hg monitors with an Hg converter. • Required only if daily calibrations are performed using elemental Hg and the monitor uses an Hg converter to monitor total mercury. • Required once every 168 unit/stack operating hours. • Specification: ≤ 10.0 % of reference value with alt spec of ≤ 0.8 µg/scm absolute difference
Linearity Check • Linearity Checks are performed with an elemental Hg standard • Required for initial certification of all Hg monitors • Required for quarterly QA unless a 3-level System Integrity Check is substituted • Specification: ≤ 10.0 % of reference value with alt spec of ≤ 0.8 µg/scm absolute difference
3-level System Integrity Check • System Integrity Checks are performed with an oxidized Hg standard generated from a NIST-traceable source • Required for initial certification of Hg monitors with Hg converters • 3-level System Integrity Check may be used instead of a linearity for Quarterly QA. • Specification: ≤ 10.0 % of reference value with alt spec of ≤ 0.8 µg/scm absolute difference
Relative Accuracy Test Audit (RATA) • Required for initial certification of all Hg monitors • Required annually (once every 4 QA operating quarters) thereafter • For Reference Methods that require paired sampling, the RM value for each run should be the average of the values observed from each of the paired samples for the run. • Specification: ≤ 20.0 % RA with alt spec of ≤ 1.0 µg/scm if average RM value is ≤ 5.0 µg/scm • No semi-annual threshold
Bias Test • Use standard bias test logic • Low Emitter BAF cap is 1.250 for when the average RM concentration is < 5.0µg/scm
Appendix K – Sorbent Trap System Basics • Appendix K – Sorbent Trap systems consist of a pair of sampling trains that collect an integrated sample over a specified collection period • Sorbent traps collect the Hg and have 3 sections. • Section 1; Main Sample Collection Section • Section 2; Breakthrough Sample Collection Section • Section 2; Spiked Section for QA • Each sampling train uses a dry gas meter* to record the volume of stack gas sampled over the course of each sample collection period *Note: Agency plans to revise Appendix K in the upcoming OAQPS direct-final alternative mercury reference method rule package to allow the use of any type of gas flow meter that accurately measures gas volumes within 2 percent. This is consistent with our performance-based measurement policy.
Appendix K – Sorbent Trap Systems • Initial Certification • RATA • Bias Test • Initial 3-level Dry Gas Meter Calibrations* • Initial DGM Temperature Sensor and Calibration Check* • Initial DGM Barometer Calibration Check* * Gas meter calibration procedures are being revised as part of the alternative reference method rule package.
Appendix K – Sorbent Trap Systems • On-going Quality Assurance for Appendix K systems and components: • Annual RATA (no semi-annual threshold) • Annual Bias Test • Quarterly Gas Meter Calibrations* • Quarterly Gas Meter Temperature Sensor and Calibration Check* • Quarterly Gas Meter Barometer Calibration Check* • Quality Assurance for each Appendix K sample collection period: • Pre- and Post- sample collection Leak Check • Monitor Ratio of stack gas flow to sample collection flow • Sorbent Trap Breakthrough check (section 2 of trap) • Percent Trap Agreement • Spike Recovery (section 3 of trap) * Gas meter calibration procedures are being revised as part of the alternative reference method rule package.
Appendix K – Sample QA • The following QA is applicable to each sample collection period for evaluating the acceptability of each sorbent trap sample collected • If any of the following QA checks are failed, the sample for that trap is invalidated. • If data for the other trap is validated, a single trap adjustment factor may be used to adjust the data from that trap for uncertainty in lieu of applying substitute data for the collection period. • If the QA checks are failed for both samples then substitute data must be used for the entire collection period.
Appendix K – Sample QA • Each sampling train must pass a pre- and post sample collection leak check. The post collection check will be reported to validate each sample • The Ratio of the hourly Stack Gas Flow Rate to the hourly Sample Flow Rate must be maintained to within ± 25% of the initial ratio from the first hour of each collection period. Samples are invalidated if more than either 5 hours or 5% of the hourly ratios do not meet this criteria. • Sorbent Trap Breakthrough check (section 2 of trap); The mass of Hg captured on section 2 of the trap must be ≤ 5% of the total Hg collected on section 1. • Spike Recovery (section 3 of trap); The analysis of Hg on section 3 must result in 75% - 125% of the spike referenced spike level for the trap.
Appendix K – Calculating the Collection Period Concentration • The Hg concentration for each trap sample is calculated by adding the Hg mass from section 1 and section 2 of the trap. • There are plans to remove the procedure to normalize the Hg mass in the upcoming reference method rule package. • Percent Trap Agreement; For each pair of traps, the Relative Deviation (RD) must be ≤ 10%. (Alternate Specification: ≤ 20% if average Hg concentration is ≤1.0µg/scm) • If the Percent Trap Agreement is within spec., the average Hg concentration is reported for each hour of the collection period • If the Percent Trap Agreement is not within spec., the higher of the two Hg concentrations is reported for each hour of the collection period
Substitute Data for CEMS and Appendix K • In August 2006, EPA proposed to consolidate the Substitute Data requirements for CEMS and Appendix K. • This decision allows for the use of an Appendix K system as a back-up to CEMS in a way that does not complicate the Substitute Data requirements for when neither system is available. • For hours where there is valid data from the CEMS, that data will be used to populate the lookback. This is also the value that must be used to calculate Hg mass for the hour. • The Appendix K data will populate the lookback period for any hour where CEMS data are unavailable. • For hours where neither system is available, substitute data is determined from the consolidated lookback of actual concentration data used for the previous hourly emissions calculations.
Substitute Data for CEMS and Appendix K • The substitute data for Hg has four tiers: • Tier 1; If the PMA is 90 or more: • If the duration of the outage is less than 24 hrs substitute the average of the hour before and after • If the duration is greater than 24 hrs, substitute the higher of the hour before/hour after average or the 90th percentile value from a 720 hour lookback • Tier 2; If the PMA is less than 90 and equal to 80 or more: • If the duration of the outage is less than 8 hrs substitute the average of the hour before and after • If the duration is greater than 8 hrs, substitute the higher of the hour before/hour after average or the 95th percentile value from a 720 hour lookback
Substitute Data for CEMS and Appendix K • Tier 3; If the PMA is less than 80 and equal to 70 or more, substitute the maximum value from the 720 hour lookback • Tier 4; If the PMA is less than 70 substitute the maximum potential concentration. • For units with add-on controls, if the controls are documented to be operated properly per §75.34, the unit may substitute the greater of the following instead of the maximum potential concentration: • The maximum expected concentration as listed in the monitoring plan, or • 1.25 times the maximum controlled value measured from the previous 720 hour lookback
Hg-LME Monitoring Methodology • Rather than install Hg analyzers or sorbent traps, affected units with lower potential levels of Hg may qualify for an excepted monitoring methodology • This method is referred to as the Hg-LME methodology • The Hg-LME excepted methodology is found in §75.81
Hg LME • Low Hg emitting units (≤ 29 lbs/yr potential emissions) may qualify for the Hg-LME methodology • This methodology requires an initial and on-going emission testing • Initial test stack test results are used in Eq. 1, with the maximum potential stack flow rate to demonstrate that the unit’s potential to emit is not greater than the 29 lb threshold. • The Hg concentration from the test is then used with actual stack flow data to determine hourly emissions • Ongoing retesting is required annually if the calculated potential Hg emissions are ≤ 9 lb/yr, and semiannually if the potential emissions are between 9 and 29 lb/yr (NO GRACE PERIODS!) • Actual reported Hg mass is used for ongoing qualification.
Hg-LME Testing • A minimum of 3 runs at normal load are required • Minimum run time is 1 hour per run is required for the instrumental method • Run time for other methods to be determined by the amount of sampling needed to acquire a measurable amount of Hg • Use the highest Hg concentration from the 3 runs or 0.50 µg/scm (whichever is greater) in when calculating the potential to emit • The same default is used for hourly reporting until the completion of the next test.
Hg LME • Equation 1 in §75.81 provides the conservative estimate of the annual Hg mass emissions from the unit (i.e., potential to emit) • Equation 1 is also used to determine whether ongoing test should be performed semi-annually or annually • Equation 1 uses the Hg concentration determined based on reference method testing consisting of three runs
Hg LME • Equation 1: E = 8760 K CHg Qmax • E = Estimated annual Hg mass emissions from the affected unit, (ounces/year) • K = Units conversion constant, 9.978 x 10-10 oz-scm/µg-scf • 8760 = Number of hours in a year * • CHg = The highest Hg concentration (µg/scm) from any of the test runs or 0.50 µg/scm, whichever is greater (paired sampling trains are required for each test run) • Qmax = Maximum potential flow rate, determined according to section 2.1.4.1 of appendix A to this part, (scfh) * Considering allowing less than 8760 hours if a permit condition restricts hours of operation to less than 8760
Hg LME • On August 22, 2006 EPA proposed revisions to 40 CFR 75 and 72.2 • Proposed revisions includethe following requirements • Coal combusted during testing must be from the same source of supply as the coal combusted at the start of the Hg mass emissions reduction program • Test a subset of identical units in accordance with §75.19(c)(1)(iv)(B)in lieu of testing each unit individually • The calculated value of E shall be divided by the number of units sharing the stack. If the result, when rounded to the nearest ounce, does not exceed 464 ounces, the units qualify to use the low mass emission methodology • Common Stack testing to be allowed with conditions