An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Computed TomographyRAD309 Data Acquisition

  2. Data Acquisition • Data acquisition represents the first step in process of image production • X-ray tube & detectors collect information systematically • Collect large number of x ray transmissions around the patient

  3. Data Collection Basics Patient • X-ray source & detector must be in & stay in alignment • Beam moves (scans) around patient • many transmission measurements taken X-Ray beams

  4. Data Collection Basics • Pre-patient beam • collimated to pass only through slice of interest • shaped by special filter for uniformity

  5. Data Collection Basics (cont) • Beam attenuated by patient • Transmitted photons detected by scanner • Detected photon intensity converted to electrical signal (analog) • Electrical signal converted to digital value • A to D converter • Digital value sent to reconstruction computer

  6. CT “Ray” • That part of beam falling onto a single detector Ray

  7. Each CT Ray • attenuated by patient • projected onto one detector • detector produces electrical signal • produces single data sample

  8. CT Projection -or- View • # of simultaneously collected rays

  9. Acquisition Geometries • Pencil Beam • Fan Beam • Spiral

  10. DA Geometries • Parallel beam, translate rotate motion • Fan beam, translate rotate motion • Fan beam, complete rotation tube/detector • Fan beam, complete rotation of tube around stationary ring of detectors • Special: high speed CT, stationary/stationary, multiple targets tube • Spiral, rotate/translate • Multiple detector rows

  11. Interconnect Wiring Tube Slip Rings Detector Spiral Geometry • X-ray tube rotates continuously around patient • Patient continuously transported through gantry • No physical wiring between gantry & x-ray tube • Requires “Slip Ring” technology

  12. X Ray System • Initially used low energy gamma rays • Problem: low radiation intensity rate, large source size, low source strength, high cost • Use of X ray tubes • Benefit: high radiation intensity, high contrast ct scanning • Problem: heterogeneous beam , does not obay Lamber-Beer Exponential Law

  13. Radioactive Source instead of an X-Ray Tube? • High intensity required • X-ray tubes produce higher intensities than sources • Single energy spectrum desired • Produced by radioactive source • X-ray tubes produce spectrum of energies

  14. CT Beam Filtration Shapes beam to appear monochromatic and satisfy reconstruction process • Hardens beam • Removes greater fraction of low-energy photons than high energy photons • reduces patient exposure • Shapes energy distribution to produce uniform intensity & beam cross section Patient Filter

  15. Patient Protection • Pre-collimators • between tube & patient Tube • Post-collimators • between patient & detector Detector

  16. Tube Detector Pre-Collimation • Constrains size of beam • Reduces amount of scatter produced • Designed to minimize beam divergence • Often consists of several stages or sets of jaws Pre-collimator

  17. Tube Detector Post-Collimation • Helps define slice (beam) thickness • Reduces scatter radiation reaching detector • Improves image quality Post-collimator

  18. Detectors • Capture radiation from patient • Converts to electrical signal • Then they are converted to binary coded information

  19. CT Detector Characteristics • Efficiency • Response time • Dynamic range • Reproducibility and Stability

  20. 1. Efficiency • Ability to capture, absorb & convert x-ray photons to electrical signals

  21. Efficiency Components a. Capture efficiency • Efficiency of detector to obtain transmitted photons from patient • Size of detector area, distance between 2 detectors b. Absorption efficiency • no. of photons absorbed • Z , density, size, thickness of detector c. Conversion efficiency • fraction of absorbed energy which produce signal

  22. Overall Detector Efficiency capture efficiencyXabsorption efficiencyXconversion efficiency

  23. Absorption Efficiency • Depends upon detector’s • atomic # • density • size • thickness • Depends on beam spectrum

  24. 2. Response Time • “Speed with which detector can detect an x ray event and recover to detect the next one” • Minimum time after detection of 1st event when detector can detect 2nd event • If time between events shorter than response time, second event may not be detected • Shorter response time better

  25. 3. Dynamic Range • Ability to faithfully detect large range of intensities • “Ratio of largest signal to be measured to the precision of the smallest signal to be discriminated” • Typical dynamic range: 1,000,000:1 • much better than film

  26. 4. Stability • “Steadiness” of detector system • Consistency of detector signal over time • The less stable, the more frequently calibration required

  27. Detector Types 2 principles: • Convert x-ray into light ---electrical signal • Scintillation detector • Convert x-ray directly into electrical signal • Gas ionization detector

  28. Scintillation Detectors • Crystal couple to photomultiplier tube • X ray falls on crystal ---light flashes (glow) • Light directed to PM • Light hits Photocathode in PM and releases electrons

  29. Light X-Rays Scintillation • X-ray energy converted to light • Light converted to electrical signal Photomultiplier Tube Electrical Signal Scintillation Crystal

  30. Photomultiplier Tubes • Light incident on Photocathode of PM tube • Photocathode releases electrons + - Light X-Rays Scintillation Crystal Photocathode PM Tube Dynodes

  31. Gas Ionization Detector • Series of individual chambers separated by tungsten plates • X ray falls on each chamber– (+/- ions) • + ions move to – plate, - ions to + plate • The migration produces electrical signal

  32. Electrical Signal X-Rays Gas Ionization • X-rays converted directly to electrical signal Filled with Air Ionization Chamber + - - +

  33. CT Ionization Detectors • Many detectors (chambers) used • adjacent walls shared between chambers • Techniques to increase efficiency • Increase chamber thickness • x-rays encounter longer path length • Pressurize air (xenon) • more gas molecules encountered per unit path length thickness X-Rays

  34. Detector Array • Slice by Slice – one arc of detector array • Volume – one arc of detector array, acquires volume of tissue then separated by computed to slice by slice

  35. DAS • Detector electronics • Location: between detector and computer • Role of translator • Measure transmitted radiation beam • Encodes measurement to binary data • Transmits binary data to computer

  36. Components of DAS • Amplifier • Log Amplifier • Analog to Digital Converter (digital data) • Digital Transmission to computer

  37. Log Amplification • Transmission measurement data must be changed into attenuation and thickness data • Attenuation = log of transmission x thickness

  38. Detector Electronics From Detector Increases signal strength for later processing Amplifier Compresses dynamic range; Converts transmission intensity into attenuation data Logarithmic Amplifier Analog to Digital Converter To Computer

  39. DA and Sampling • Radiation falling on detector • Each samples the beam intensity on it • Not enough samples = artifacts appear • To increase number of measurement/samples available for reconstruction and improve image quality

  40. Improving Quality & Detection • Geometry • Smaller detectors • Closer packed detectors • Smaller patient-detector distance • Thinner slices • less patient variation over slice thickness distance