1 / 60

Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis

Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis. Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis. Physicists (University of Arizona) RJ Hamilton CJ Watchman J Gordon NJ Stanley Radiation Oncologist (University of Arizona) B Stea

creda
Download Presentation

Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis

  2. Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis Physicists (University of Arizona) RJ Hamilton CJ Watchman J Gordon NJ Stanley Radiation Oncologist (University of Arizona) B Stea Neurosurgeon (Center for Neurosciences, Tucson, AZ) A Sanan

  3. Accuracy of Frame-Based Novalis SRS for TGN • Rahimian et al (J. Neurosurg S3 101:351-55, 2004) • Image Fusion 0.4 mm • MR 0.2 mm • Gantry 0.3 mm • Couch 0.6 mm • Overall 0.9 mm (0.6 mm w/o couch) • Target position 0.2 mm • Laser 0.2 mm • Phantom Radiographic Test 0.7 ± 0.1 mm

  4. Frame-Based SRS for TGN Human Issues Pins fix frame to skull Uncomfortable during treatment delivery too Patient remains in department for several hours waiting for treatment Difficult to abort procedure (machine problems) Scheduling: patient, neurosurgeon, radiation oncologist, physicist, treatment time, etc…

  5. Frame-Based SRS for TGN • Technical Issues • Frame-based systems require that the position of a patient in the frame is unchanged from application of the pins through treatment completion • The stereotactic frame (not the patient) is aligned with the machine isocenter for treatment • Differential torque on the head ring deforms it • Head ring is bolted to the table assembly, patient may be displaced relative to it.

  6. Image-Based SRS for TGN • Human Issues • Plastic mask (BrainLab U-PLAST mask fixation system)helps patient remain still • Scheduling: specialists (neurosurgeon, radiation oncologist) may work at their convenience, treatment time set for patient

  7. Image-Based SRS for TGN • Technical Issues • Positioning accuracy • Depends on the accuracy of the registration of images between the planning system (DRRs) and imaging system (ExacTrac) • Depends on the calibration of the imaging system with the mechanical isocenter • Variations of the position of the patient in the mask (stereotactic coordinate space) are not important provided that the imaging system can correct for it (ExacTrac X-Ray 6D and Robotics) • Once positioned for treatment, the patient must remain fixed.

  8. Feasibility of Image-Based SRS for TGN • Demonstrate System Accuracy • Anthropomorphic Phantom Experiment • Demonstrate that variations of the patient position in the mask are not important • Anthropomorphic Phantom Experiment • Confirm for each patient that deviations of their head within the mask may be corrected by the ExacTrac X-ray 6D and robotics • Perform planning CT scan plus two addition CT scans. The final scan is performed 1 hour prior to the scheduled treatment time. • Confirm that the patient does not move during treatment • Repeat the ExacTrac X-ray positioning 3-4 times (every 3rd couch position) during treatment delivery

  9. Demonstrate System Accuracy • Anthropomorphic Phantom • 0.25” BB embedded in head • BrainLab U-PLAST mask constructed • Axial CT scan (35 cm FOV 1.5mm slice) with BrainLab Head & Neck Localizer • BrainScan Plan • ExacTrac 6D X-Ray and Robotics

  10. Anthropomorphic Phantom CT

  11. Anthropomorphic Phantom CT

  12. Anthropomorphic Phantom CT

  13. Anthropomorphic Phantom Plan

  14. Accuracy of Isocenter Determination • Two planners contoured the BB in the planning system and then placed the isocenter at the center. Repeated the procedure 10 times each.

  15. Pixel Size Pixel Size = 0.5 mm

  16. Pixel Size Pixel Size = 0.5 mm

  17. Pixel Size Pixel Size = 0.5 mm

  18. Pixel Size Pixel Size = 0.5 mm 0.5 mm

  19. ExacTrac Calibration

  20. ExacTrac Calibration

  21. ExacTrac Calibration

  22. ExacTrac Calibration

  23. ExacTrac Calibration

  24. ExacTrac Calibration

  25. ExacTrac Calibration

  26. ExacTrac Calibration

  27. ExacTrac Calibration

  28. ExacTrac Calibration

  29. ExacTrac Calibration

  30. Anthropomorphic Phantom Treatment Head Holder

  31. Anthropomorphic Phantom Treatment

  32. Anthropomorphic Phantom Treatment Select Isocenter

  33. Anthropomorphic Phantom Treatment Automatic Preposition

  34. Anthropomorphic Phantom Treatment

  35. Anthropomorphic Phantom Treatment Automatic Preposition Film Verification Difference = ( 0.3, -0.2 ) = 0.4 mm 10 mm cone BB shadow

  36. Anthropomorphic Phantom Treatment

  37. Anthropomorphic Phantom Treatment

  38. Anthropomorphic Phantom Treatment

  39. Anthropomorphic Phantom Treatment

  40. 10 mm cone BB shadow Anthropomorphic Phantom Treatment Film Verification Difference = ( 0.1, 0.1 ) = 0.1 mm

  41. Phantom Displacement Experiment • Phantom moved by exactly 2.0 cm in all three directions as determined by a ruler • Perform X-ray Alignment with ExacTrac • Shifts Found by ExacTrac • AP 19.85 mm • Lateral 20.05 mm • Vertical 20.40 mm

  42. Phantom Moved: Not in Mask !!! Head Rotated

  43. Head Holder Removed Phantom Moved: Not in Mask !!! No Head Holder  Head Lower and Rotated

  44. Phantom Moved: Not in Mask !!! Position using standard procedures with array

  45. Phantom Moved: Not in Mask !!!

  46. Phantom Moved: Not in Mask !!!

  47. Phantom Moved: Not in Mask !!!

  48. 10 mm cone BB shadow Phantom Moved: Not in Mask !!! Film Verification Difference = ( 0.1, 0.1 ) = 0.1 mm

  49. Phantom Moved: Not in Mask !!! Rotation too large for robotics

  50. Phantom Moved: Not in Mask !!!