RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology. RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. Part 15.3: Optimization of protection in radiography Practical exercise - HVL measurement. Overview / Objectives.

curry
Download Presentation

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION INDIAGNOSTIC ANDINTERVENTIONAL RADIOLOGY Part 15.3: Optimization of protection in radiography Practical exercise - HVL measurement

  2. Overview / Objectives • Subject matter : HVL measurement • Step by step procedure to be followed to implement the considered QC test • Interpretation of results 15.3: Optimization of protection in radiography

  3. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 15.3: Optimization of protection in radiography HVL (Filtration) Measurement

  4. Half Value Layer (HVL) (I) • Possibly the most important test • Checks whether there is sufficient filtration in the x-ray beam to remove damaging low energy radiation 15.3: Optimization of protection in radiography

  5. What Minimum Equipment is Needed? • Electronic device (multi-function meter) to measure dose • High purity aluminium layers. (Most aluminium has high levels impurities, e.g., Cu) • Stand to hold the dosimeter, e.g., laboratory stand • Tape measure, spirit level • Roll of removable tape 15.3: Optimization of protection in radiography

  6. Tape Measure, Spirit Level 15.3: Optimization of protection in radiography

  7. Half Value Layer (II) • Method (simplest - using a multifunction test meter) : • Place the detector half way between the front of the collimator and the table top. • Set 80 kVp, fixed mAs (e.g., 50 mAs) • Collimate x-ray beam to size of detector • Measure exposure three times 15.3: Optimization of protection in radiography

  8. HVL Measurement Geometry X-Ray Tube Collimator Aluminium Sheets X-Ray Beam Detector Table top 15.3: Optimization of protection in radiography

  9. Half Value Layer (III) • Tape a 1 mm sheet of aluminium to the front of the collimator and measure the dose • Add another 1 mm sheet of aluminium, measure again, and repeat until the dose has fallen to below 50% of the initial, unattenuated, value • Remove all aluminium, and make 3 more exposures 15.3: Optimization of protection in radiography

  10. Half Value Layer (IV) Analysis : • Average all the “no Al” measurements • Plot all results on semi-log graph paper • From the graph, find the thickness of Al required to reduce the unattenuated dose by 50% - this is the HVL • The HVL must be at least that specified in local regulations, e.g., 2.3 mm Al at 80 kVp • Ideally, it should be at least 3.0 mm Al or higher to assist in minimizing patient dose 15.3: Optimization of protection in radiography

  11. HVL Measurement Dose • Be careful of beam hardening (semi-log plot is not a straight line) • Use points either side of half initial value • Calculate HVL : (initial value = 9 50% of this = 4.5, thus HVL = 2.6 mm Al) mm Al 15.3: Optimization of protection in radiography

  12. Minimum HVL Values (IEC) Acceptable Not acceptable 15.3: Optimization of protection in radiography

  13. Where to Get More Information Quality Control in Diagnostic Imaging, Gray JE, Winkler NT, Stears J, Frank ED. Available at no cost. http://www.diquad.com/QC%20Book.html 15.3: Optimization of protection in radiography