T1ρ MRI

T1ρ MRI . e xample images and review. T1rho MRI Hershey Medical Center 9/15/13. There is moderate B 0 field heterogeneity in this image. This field heterogeneity causes banding artifacts.

damisi
Download Presentation

T1ρ MRI

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. T1ρ MRI example images and review

  2. T1rho MRI Hershey Medical Center9/15/13 There is moderate B0 field heterogeneity in this image. This field heterogeneity causes banding artifacts. It’s difficult to see the artifact because it appears differently in each image. Compare closely the overall signal homogeneity in the ventricular cavity or myocardium and you will see it. Compare white arrows. The artifact is obvious on the movie in the upper-right corner.

  3. potential solutions to resolve the B0 artifact • Check the shim mode. • Is it ‘tune up’ or ‘standard’? • Set the center frequency using the the frequency scout (product sequence). • Shim locallyusing the local shim interface. • Localize this region around the heart. • Try a local shim in the short and long axis. • Copy the shim from a previous scan. • Right-click on the previous scan and select ‘copy parameter’ > ‘adjust volume’

More Related