1 / 18

Case Study : High Voltage Leak Detection for Low Conductive Product

Case Study : High Voltage Leak Detection for Low Conductive Product. Pharmapack Limited Nikka Densok Limited. Contents . Test Product Information (Low Conductivity) Inspection Principle Introduction of Production Line – Enema Formulation ( Weighing ~ Dispensing ~ Filling ~ Packaging)

Download Presentation

Case Study : High Voltage Leak Detection for Low Conductive Product

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Case Study : High Voltage Leak Detection for Low Conductive Product Pharmapack Limited Nikka Densok Limited

  2. Contents • Test Product Information (Low Conductivity) • Inspection Principle • Introduction of Production Line – Enema Formulation (Weighing ~ Dispensing ~ Filling ~ Packaging) • Scope of Pinhole Inspector (Specifications, Transfer System, Inspection Screen Indication) • Report of Experiment to Select Pinhole Inspector (Good Product Signal, Bad Product Signal, False Rejection of Good Product) • Report of Result in Production

  3. Test Product Information • Mainly Glycerin Enema • Product is Low Conductive between 2.0 – 3.5μS/cm.

  4. Introduction of Production Line – Enema Formulation(Weighing ~ Dispensing ~ Filling ~ Packaging)

  5. InspectionPrinciple

  6. Examination to Select Pinhole Inspector • Step 1 Examination of Electrode and Transfer System • Step 2 Analysis of Good /Bad Signal Values • Step 3 Definition of Upper Reference for Detection

  7. Step 1Examination of Electrode Configuration ①Definition of Inspection Area : where Sealing Mold and Main Mold contact each other. ②Examination of Electrode Configuration - should be able to inspect the defined inspection area. Inspection Area

  8. Transfer and Inspection System

  9. Analysis of Good/Bad Product Signal Values

  10. If Upper Limit is set higher, chance of failure to discharge bad products to good product line will be higher. If Upper Limit is set lower, chance of false rejection to reject good product will be higher. Upper Limit

  11. Pinhole Inspector Model HDY-02+DL

  12. Application Product Size Maximum Capacity : BFS Bottles : 30ml – 150ml : 85pcs/m. General Specifications

  13. Transfer and Inspection System

  14. Production Bach Display

  15. An Example of Inspection Result

  16. Pinhole Test Result of Glycerin Product

  17. An Example of Pinhole(Digital MicroScope VHX-99 KEYENCE) 20μm Signal Value 1Ch 4.3V        2Ch 3.0V

  18. An Example of Pinhole(Digital MicroScope VHX-99 KEYENCE)