1 / 18

Lab tests of Thick GEMs (THGEM)

Lab tests of Thick GEMs (THGEM). S. Dalla Torre, Elena Rocco , L. Ropelewski, F. Tessarotto May – August 2007. Outline:. Geometry of the THGEM tested; Sources & Setup; First discouraging results; Different geometry and encouraging results; The rim effect; Conclusions. Ions.

delila
Download Presentation

Lab tests of Thick GEMs (THGEM)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Lab tests of Thick GEMs (THGEM) S. Dalla Torre, Elena Rocco, L. Ropelewski, F. Tessarotto May – August 2007

  2. Outline: • Geometry of the THGEM tested; • Sources & Setup; • First discouraging results; • Different geometry and encouraging results; • The rim effect; • Conclusions.

  3. Ions 40 % 60 % Electrons GEM Principle 5 µm 50 µm 55 µm 70 µm GEM hole cross section Avalanche simulation

  4. Some THGEM pictures W2: D=0.3 mm Pitch=0.7 mm Rim=0.1 mm Thick=0.4mm R3: D=0.2 mm Pitch=0.5 mm Rim=0.01 mm Thick=0.2mm R3 W2 P1: D=0.8 mm Pitch=2 mm Rim=0.04 mm Thick=1mm R3 section P1

  5. Sources & Parameters of the THGEMs used *Except for the Pi geometry we always used 30/70 CO2/Ar gas mixture !!

  6. DRIFT d_drift THGEM d_ind ANODE Structure of the chamber used for testing GAS INLET • Non segmented anode (copper foil); • Inlet and outlet (on the cover) for the gas; • Flux gas of 5 l/h. Section view of the structure inside the chamber DRIFT IMPORTANT: Before installing bath, backing in the oven of the THGEM to avoid leakage current. THGEM

  7. Cu X-Ray setup

  8. Electronics Setup and acquisition Power Supply CAEN 471A FAN I/O Le Croy 428F DISCRIMINATOR Le Croy 821 SCALER CAEN N 145 G 472 ORTEC Amplifier 142 A ORTEC Preamplifier Delay ADC (LRS 2259 12 ch) + DAQ (CAEN controller C111) THGEM in the chamber Digital Oscilloscope Gas system with mass flow meter mixing (30%CO2 70% Ar)

  9. d=0.3mm Rim=0.1mm Pitch=0.7mm W2 One of the first trial: scan in time @ 40 Hz with the THGEM characterized by the “Weizmann geometry” @ out of the ADC range Gain variations larger than a factor of 2 !

  10. R3 Our best result so far with R3(d=0.2mm pitch=0.5mm rim=0.01 thick=0.2mm) X-Ray Source Collimated Rate =6.6 KHz 55Fe Source Uncollimated Rate =260 Hz <10% 15% Long time scan (~ 4 days) Short time scan (< 1 days)

  11. Comparison between different sources R3 primary e- in 55Fe R= primary e- in X-Ray 210 0.66 = R= 320 ADC ch. peak position with 55Fe R= ADC ch. peak position with X-Ray 417.8 R= = 0.74 563.5 RATE=460 Hz

  12. R3 Rate capability Rate effect on signal amplitude: ~ 20%, varying the rate by 3 orders of magnitude! Also, from current measurement  gain ~ 700

  13. R3 Gain Estimation for different signal amplitudes ‘ Cut spectrum due to the threshold on the discriminator giving the trigger signal Gain  const e –V/t t ~ 50V

  14. Rim effect Is this dramatic gain increase with time a rim effect? (Recall that the increase is much smaller with 10 micron rim). • Try thicker THGEM, larger holes w/o rim … We come back to the Weizmann geometry (d=0.3 mm, pith=0.7 mm, thick=0.4), but w/o rim

  15. NO gain increase with time ! but again stability only at moderate gains (~ 700)  next step: chemical polishing to remove sharp edges and asperities due to copper drilling.

  16. Time stability- THGEM polished chemically Residuals after mechanical drilling Before chemical polishing After chemical polishing THEN WE CREATE A SMALL RIM!!!

  17. Rate capability – THGEM polished chemically

  18. Conclusions: • In the very near future we are : • Characterizing a DOUBLE THGEM configuration; • Measuring the first THGEM coated with the CsI in the test beam; • The work is in progress (and promising!), but there’s still a long way to go: • Geometry role; • Technological production; • Different gasses….