Intoxilyzer 5000EN Factory Technical Specifications. Rev 3.09. Intoxilyzer 5000EN Factory Technical Specifications. RANGE: 0.005 to 0.450 BrAC grams/210 liters Majority of BrAC results are between 0.005 BrAC and 0.450 BrAC Results less than 0.005 BrAC = 0.000 BrAC on display
E N D
Intoxilyzer 5000ENFactory Technical Specifications • RANGE: 0.005 to 0.450 BrAC grams/210 liters • Majority of BrAC results are between 0.005 BrAC and 0.450 BrAC • Results less than 0.005 BrAC = 0.000 BrAC on display • Results greater than 0.450 Brac = Range Exceeded Error Message 0.005 BrAC 0.450 BrAC <0.005 BrAC = 0.000 BrAC >0.450BrAC = Range Exceeded Error Message Majority of BrAC Results Rev 3.09
Intoxilyzer 5000ENFactory Technical Specifications • Accuracy: +3% or +0.003 BrAC which ever is greater • Same accuracy as similar laboratory equipment Rev 3.09
Intoxilyzer 5000ENFactory Technical Specifications • Precision: Standard Deviation of 0.003 BrAC or better • If 10 tests of a known ethanol solution were run on the instrument, all results would be within 0.003 BrAC of each other Rev 3.09
Intoxilyzer 5000ENFactory Technical Specifications • Test Time: Approximately 7 minutes • For Colorado test sequence data entry time + (ACABAIA2ABACA) Rev 3.09
Intoxilyzer 5000ENFactory Technical Specifications • Operating Temperature: • 68 F to 86 F (20 C to 30 C) room temperature • Keep room temperature at approximately 75 F Rev 3.09
QUESTIONS? Rev 3.09