Improvement of the 11 C gas target Erol Bars, Munawwar Sajjad , Mike Haka, Steve Toorongian

Department of Nuclear Medicine.

doctor
Download Presentation

Improvement of the 11 C gas target Erol Bars, Munawwar Sajjad , Mike Haka, Steve Toorongian

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Department of Nuclear Medicine Design goal :-Increased Yield By 50%. -Low and quick maintenance cycle. -Suitable to Multi-geometrical internal volume Chamber. -Stand alone uPLC- PC GUI. -Low cost High reliability.Future Projects:-New valves layout for less radiation damage.-One isolation foil used in grid flange, Water cooled, no helium. Improvement of the 11C gas target Erol Bars, Munawwar Sajjad , Mike Haka, Steve Toorongian

  2. 11C Target Properties: Internal gas Volume:188cc Length internal Chamber: 8.5” Orifice in: 0.472” Orifice OUT: 2” Water Flow: 6-8Lm 2X2way GE valves used (COLD CUBE control - Cole Parmer) 1 kulite Pressure sensor All Aluminum parts 6061 Isolation Foils: 25um Al (Tron Side) 0.5mm Al 2024(0.02”) (Trg) Target face O-ring :Viton Helium flow: 50L/min. Energy: 22.5Mev Beam Current: 35-40uA 11C in Maintenance condition Previous Target

  3. 11C Design concept

  4. Target 11C (Nemo) Software uPLC Pic microcontroler: PIC16F876:22I/O,5AN,Ser,PWM, 8K x 14 words Flash,20MHZ…. Software: PicBasic Pro Compiler MicroCode Studio ,MPLAB Chip burner: PicStart+ -Visual basic 6.0 used as GUI - Communication protocol serial. -Stand alone platform (File.exe)

  5. 11C Yield Gas used 1%Oxygen+Nitrogen Average data graph over 100 runs Results

  6. Future Projects. 11C Target #2 New Layout Water Cool Grid Foil flange (one foil used) 500W power analysis. Grid Face 491K Conclusion: Projects has shown all the design goals. Cosmosflowork