The big picture I. Reinhard Hanuschik, QC Garching. Why do we take calibrations? SCIENCE observations contain: scientific information OB: question by astronomer to the Universe science observation: answer by the Universe. Calibrations. CALIBs contain:
E N D
The big picture I Reinhard Hanuschik, QC Garching • Why do we take calibrations? • SCIENCE observations contain: • scientific information • OB: question by astronomer to the Universe • science observation: answer by the Universe
Calibrations • CALIBs contain: • information about instrumental effects (ARC, FLAT, BIAS) • or atmospheric effects (STD flux, telluric) • SCIENCE reduction: • raw science calibrations with pipeline
Calibrations • CALIBs: • clean, specialized measurement of certain instrumental effect • principle: input known see what the instrument does with this input know about instrument
Assumptions • Instrument is stable between night and daytime • then: daytime calibrations sufficient • meaning: instrument conditions between day and night identical to acceptable level • if not (due to unexpected conditions): • science cannot be calibrated, lost
Criteria • Two important quality criteria on calibrations • 1. Completeness • If CALIB missing, or earlier/later CALIB exists but is inapplicable: • SCIENCE cannot be calibrated • CALIB completeness saving SCIENCE from last night • provided by calChecker
Criteria • 2. Quality • good quality valid measurement of instrumental conditions • valid: e.g. good flux and SNR • overexposed flat: always bad • well-exposed ARC, but grating not in same position as for science: bad • Also: • HC calibs check instrument status rather than individual science • CALIB quality ensures science of next night provided by HC monitor
The big picture II Where in the dataflow is QC and daytime calib checks? What is done on Paranal and in Garching?
Garching • automatic process for every instrument: • calChecker every 30 minutes (tellTracker 15 min) • HC monitor every hour • running 24/7 • output web pages identical for Garching (HQ) and Paranal (PL) • for convenience: mirror site ‘PL’ (replicated within 15 min)
Garching • QC scientists: • each is responsible for about 4 instruments • 5/7 schedule, daytime only • no shift system, no backups! If sick/travel: only automatic services • in charge of configuration and maintenance of these tools • e.g. calibration plan, scoring system • review of past day results • analysis of calChecker issues • analysis of red scores on HC monitor and AB monitor • produce master calibrations
Garching • If issues found: • fix them (if pipeline related), or ticket to developeror • understand them (based on trending, experience etc.)and • inform Paranal (mail, comment editors) • QC Garching has the first chance!
Paranal • Daytime 7/7: • review (double-check) calChecker and HC monitor, to fill gaps(weekends, issues in Garching evening/night) • react on Garching signals, coordinate actions, PPRS • pro-active: discover issues (never rely 100% on QC Garching, the QC scientist might be sick or on holiday) • but no need for duplication: if all scores are green, you can continue with the next instrumentno need for visual checks (unless HC system incomplete)! much better: check product monitors (P…) for details
Ideal workflow to check calChecker I • Daytime 7/7: • calScores green all ok, done, next instrument! • don’t forget cal4cal • don’t forget long-term calibrations • calScores red investigateMISS, NOK: check if calOb queue wasinterrupted/stopped restart • if yellow: wait for end of daytime calib, no need to check earlier! • nighttime CALIB missing check with OB grade, USD, QCG • provide your analysis text: very useful!
Ideal workflow to check calChecker II • analysis editor: • choose flag (MISS, NOK, OK) • enter comment • your email address • submit • to be taken into account upon next execution (within ≤ 30min) • contact QC
Ideal workflow to check HC monitor I • HC instrument score: • green/OK all ok, done, next instrument • red/NOK find reports (navigation or ‘score overview’)check scores and reportscheck comments, ‘Report news’ • understand origin, enter ‘Report news’, contact QC
Ideal workflow to check HC monitor II • useful info for understanding issues: • plots • ‘Report news’ • QC comments from the product pages • earlier HC plots (history) • tutorial (QC docu) Don’t hesitate to provide your own comments under ‘edit’!
What to do if … • red flags on calChecker? • analyze issue, take actions • e.g. science graded ‘C’, no issue: enter analysis, issue gets flagged ‘OK’ (upon next execution) • MISS or NOK calibration take it, issue turns ‘OK’ upon next execution • calChecker outdated? • inform QC • data transfer/NGAS access not working? • inform DHA • also check out ‘HELP’
What to do if … • red scores on HC monitor? • analyze issue • if you think the instrumental issue is gone: take another calibration; score will automatically turn green if true (within an hour or less) • provide comment ‘Report news’ • often: page still cached press ‘Ctrl+Shift+R’ to refresh page • HC monitor outdated (older than 24 hours)? • inform QC • Find more under ‘HELP’
What to do if … • red flag for ‘parser’ or ‘ftp’ on HC monitor? • page still cached? press ‘Ctrl+Shift+R’ to refresh page • if real issue: inform QC • comment editor used but comment not visible? • page still cached? press ‘Ctrl+Shift+R’ • might take up to 15 minutes until replication to ‘PL’