1 / 18

Identifying and eliminating limiting noise sources of GEO600

ILIAS WG1Meeting, Hannover, September `04 Joshua Smith for the GEO team. Identifying and eliminating limiting noise sources of GEO600. P and Q. TFs depend on demod. phase, now P is close to optimal for diff. arm signals at DC GW content present in both: Q = h + nQ P = h + nP.

dorit
Download Presentation

Identifying and eliminating limiting noise sources of GEO600

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ILIAS WG1Meeting, Hannover, September `04 Joshua Smith for the GEO team Identifying and eliminating limiting noise sources of GEO600

  2. P and Q • TFs depend on demod. phase, now P is close to optimal for diff. arm signals at DC • GW content present in both: Q = h + nQ P = h + nP

  3. Sept ‘04 vs. Theoretical

  4. Noise Projections to hP

  5. Noise Projections hQ

  6. Oscillator Noises • PR & SR oscillator noise couplings to be determined • MID oscillator noises: • amplitude:2-3 below from 400Hz to 1.1kHz (at least)! • amplitude stab. loop (compare RF, stable DC ref) • first tests: no change in sensitivity • Phase: 8 below @ 1kHz • projection TBD next week • crystal oscillators (-165 dBc/rt(Hz) possible) • `2f’ local oscillator

  7. Dark noise + Shot noise • HPD dark noise measured with no light, may underestimate level • Shot noise (not shown, but at same level) calculated from current of HPD in lock, also slight underestimate • Monday: measure while MI locked on another diode (alignment quadrant cam), demodulate HPD signal at Fmi + 100kHz or so

  8. Modulation Depths Scanning FP measurement of light reflected from MPR C SR SR Normalized: Carrier: 1 SR-SB: 0.068 MI-SB: 0.020 PR-SB: 0.027 PR PR MI MI

  9. SR Mod. Depth • SR loop noise was not projected to be limiting • Possible explanation: coupling via large beats on Common-mode diode, or amplitude noise at dark port?

  10. PR mod depth • Dark noise of Quadrant diode drives frequency noise via loop gain • Switch to single-element diode allows (factor 2) reduction of mod depth without adding noise

  11. Other Bad Guys • acoustic & air current noises • Couples at some level, output bench and MIC breadboard strongest • Output telescope, HPD in vacuum now (some improvement@100Hz, cause?) • perhaps enclosure for MIC components • scattered light • visible problem, nonlinear, excited with e.g. stomps • Removed some clipping by shifting MSRr suspension • frequency noise • lower noise on main diode, improved gain distribution • complex loop not yet fully understood

  12. Dark Port Amplitude Noise Contrast Defect = P_carr_darkp / P_carr_BS = 1.5mW / 270W = 5e-6 Normalized C: 1 MI L: 16.4 SR: 4.8 MI H: 3.9 • Apparently reduced by SR modulation depth • Coupling may not be linear w/ eg. mod depth due to different relative SB ratios RIN @ dark port = V/rt(Hz) / V_DC = 5e-7/rt(Hz) @ 100 Hz = 5e-8/rt(Hz) @ 1kHz MI SR MI C

  13. Removing the Noise • MID fast path noise [factors are for that noise source, not h(t)] • bypass sqrt-circuit, reduces noise, increases nonlinearity • build noise reduction loop around HVA, factor 10 possible • Reduce Vbias in lock (controlled by LabView) • dark & shot noise • Increase laser power (2W to 10W into MC1) should give factor 2.2 • Increase PR factor by replacing T=1.35% MPR with T~0.1% mirror, circulating ~ 10kW rather than 1.4kW (270 W now), factor 2.7 • Laser amplitude noise • second loop commissioned, noise seems to be added after sensor, perhaps sensing at later point is required • identify the cause, coupling • Front end (signal) limited loops (SRC, MID < 200Hz) • Cross-projections : MID to SRC, etc. • coupled noise decrease with reduction of noise in other loops

  14. ESD Nonlinearity • Dielectric polariz. ~ V and F on dielectric ~ V, thus F ~ V**2 • Without sqrt-circuit this creates beats and 2f up-conversion • beat ratio = <A>/A(beat) = offset*(A(1)+A(2))/(A(1)*A(2)) • 2f upcon ratio =A(f1)/A(f2) = 4*offset/A(f1) • sets a limit on dynamic range

  15. Nonlinear ESD Noise

  16. Goals • Improve sensitivity before S4 • Improve understanding of couplings • Understand noise in PR loop (by next mtng!)

  17. MID loop (MI diff. arm-length) • LSC_MID_EP-P_HP • main GW signal, inphase • LSC_MID_FP-MCE-MCN  diff. FB sampled after MID electronics • LSC_MID_FP-MCE-HV, LSC_MID_FP-MCN-HV  diff. FB sampled after HVA

  18. MI Fast Path