1 / 37

The Impact of Regulations on Medical Device Design

The Impact of Regulations on Medical Device Design. Richard C. Fries, PE, CRE Manager, Reliability Engineering Datex-Ohmeda Madison, Wisconsin. Extra Activities for Regulated Industries. Develop and maintain a Quality System Product Documentation Design History File Technical File

eblack
Download Presentation

The Impact of Regulations on Medical Device Design

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Impact of Regulations onMedical Device Design Richard C. Fries, PE, CRE Manager, Reliability Engineering Datex-Ohmeda Madison, Wisconsin

  2. Extra Activities for Regulated Industries • Develop and maintain a Quality System • Product Documentation • Design History File • Technical File • Product submissions • Testing certifications • Extra time for: • Submissions • Answer questions from regulators • Re-submissions • Audits

  3. The Typical Road to Market for a Non-Medical Device • Generate a new idea for a product • Design the product • Test the product • Manufacture the product • Ship the product

  4. The Typical Road to Market for a Medical Device • Generate a new idea for a product • Design the product • Test the product • Submit data to the regulatory agency • Wait • Manufacture the product • Ship the product

  5. Timing of Product Development • Establish a window of opportunity to sell the product • Determine the amount of time to manufacture the product • Determine the amount of time for regulatory approval • Determine the amount of time to test the product • Determine the amount of time to design the product • Determine the amount of time to specify the product • Start the development cycle

  6. Types of Regulations • Process • ISO 9000 family • Audits by Notified Bodies • Product • Food and Drug Administration (FDA) • Medical Device Directive (MDD) • Individual country requirements (Canada, Australia, Japan, Russia) • City of Los Angeles • Other standards required for certain products • Environmental standards

  7. Process Regulations • Basis for product regulations • Requires the company to show an experienced quality system in place • ISO 9000 family used as the gold standard • For companies with design capabilities, ISO 9001 is the foundation • For medical device companies, ISO 13485 is beginning to be accepted

  8. ISO 9001 • Management responsibility • Quality system • Contract review • Design control • Document and data control • Purchasing • Control of customer supplied product • Product identification and traceability • Process control • Inspection and testing

  9. ISO 9001 • Control of inspection, measuring, and test equipment Inspection and test status • Control of non-conforming product • Corrective and preventive action • Handling, storage, packaging, preservation, and delivery • Control of quality records • Internal quality audits • Training • Servicing • Statistical techniques

  10. Design Control • Design and development planning • Organizational and technical interfaces • Design input • Design output • Design review • Verification • Validation • Design changes

  11. Product Regulations • United States • FDA • Europe • Medical Device Directive • Other Countries • Australia • Canada • Japan • Russia

  12. Food and Drug Administration • Quality system • Testing to prove the safety and efficacy of your product • Submission material dependent on the type of product you are making • Particular attention to software • MDRs • Recalls • Audits

  13. Food and Drug Administration • Safety and efficacy: • Requirement verification • Risk analysis • Environmental testing • Clinical testing

  14. Food and Drug Administration • Submissions: • Class I Little regulation • Class II 510(k) • Class III PMA

  15. FDA 2004 User Fees • Large business: • 510(k) $ 3,480 • PMA $206,811 • 180 day supplement $ 44,464 • Real-time supplement $ 14,890

  16. FDA 2004 User Fees • Small business: • 510(k) $ 2,784 • PMA $ 78,588 • 180 day supplement $ 16,896 • Real-time supplement $ 5,658

  17. Food and Drug Administration • Software: • Based on an bad experience in Canada • FDA doesn’t understand it • Therefore, they over-regulate it • All current regulations are in draft form • Software in a device is the same level as the device • Excess documentation required • Auditors free to regulate according to their own principles

  18. Food and Drug Administration • MDRs and Recalls: • MDR: a report sent to the FDA detailing the circumstances of your device killing or causing serious injury to a patient • The FDA also gets a report from the hospital or clinic where the situation occurred • Recall: a detailed plan for making design changes in all your devices currently in the field

  19. Food and Drug Administration • Audits: • General • Triggered by submissions • Triggered by field failures • Triggered by unsolicited information

  20. Medical Device Directive • Required for selling a product in Europe • Product must contain a CE mark • Must have a quality system • Product must meet a list of essential requirements • Certificates for all testing

  21. Medical Device Directive Process • Analyze the device to determine which directive is applicable • Identify the applicable Essentials Requirements List • Identify any corresponding Harmonized standards • Confirm that the device meets the Essential requirements/Harmonized Standards and document the evidence • Classify the device

  22. Medical Device Directive Process • Decide on the appropriate conformity assessment procedure • Identify and choose a notified body • Obtain conformity certifications for the device • Establish a Declaration of Conformity • Apply for the CE mark

  23. Medical Device Directive • Three directives: • Active Implantable Medical Devices Directive (AIMDD) • Medical Devices Directive (MDD) • In Vitro Diagnostic Medical Devices Directive (IVDMDD)

  24. Essentials Requirements List

  25. Declaration of Conformance • Every device, other than a custom-made or clinical investigation device, must be covered by a declaration of conformity • Document that states you have met all the essential requirements for your device • Must include the serial numbers or batch numbers of the products it covers • Signed by a member of Senior Management

  26. The CE Mark XXXX

  27. Difference Between FDA and MDD • FDA: • A submission must be sent to the FDA for each product to be marketed • Must wait for approval • MDD: • A company may qualify for self-certification to MDD for their products. These are checked during scheduled audits.

  28. Other Product Regulations • Countries • Japan • Australia • China • Russia • Type of Device • Alarms • Software • Environmental • EMC • Temperature/Humidity • Shipping

  29. Audits • 1-4 people in your spaces for 3 days to several months

  30. Audits • Will cover in detail your process and products • Auditors will “dig-in” in they find the hint of a problem • Major discrepancies will shut you down until they are fixed • Legal and/or punitive steps may be taken

  31. Newest of the Regulations • HIPAA • Health Insurance Portability and Accountability Act • Main components are Privacy and Security

  32. Protected Health Information (PHI) • PHI is health Information that: 1) is created or received by a health care provider, health plan, employer, or health care clearinghouse, and 2) relates to the past, present, or future physical or mental health or condition of an individual, the provisions of health care to an individual, or the past, present, or future payment for the provision of health care to an individual, and i) that identifies the individual or ii) with respect to which there is a reasonable basis to believe the information can be used to identify the individual.

  33. Protected Health Information (PHI) • Any health information that can be identified to a person • It includes information about treatment and care • PHI can include: • Name • Dates • Record number • Social security number • Full face photo • Any other unique identifying information

  34. De-Identification • Patient information from which identifiers have been deleted, redacted, or blocked, so that remaining information cannot reasonably be used to identify a person. Identifiers to be deleted include: • Name • Social security number • Address • Telephone number • Birth date • Admission date • FAX numbers • E-mail addresses • Medical record numbers • Health plan beneficiary numbers • Account numbers • Certification/license numbers • Full face photos.

  35. Civil Penalties for Non-Compliance • $100 for each violation • Total of $25,000 for all violations of an identical requirement in a calendar year

  36. Criminal Penalties for Wrongful Obtainment/Disclosure of PHI • Not more than $50,000 and/or not more than 1 year impisonment • Not more than $100,000 and/or not more than 5 years imprisonment if the offense is “under false pretenses” • Not more than $250,000 and/or not more than 10 years imprisonment for the intent to sell, use for commercial advantage, personal gain, or malicious harm Protected Health Information

  37. HIPAA Philosophy What I see here, What I hear here, When I leave here, Remains here!

More Related