1 / 9

Dose Calibrator Quality Control

Dose Calibrator Quality Control. Alejandro Treto Keiser College. A Dose calibrator operation is based on the collection of ions formed as the ionizing radiation passes through the detecting medium (gas). Detecting medium: He, Ar. Basic Design of the Gaseous Detector. + +

eitan
Download Presentation

Dose Calibrator Quality Control

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Dose CalibratorQuality Control Alejandro Treto Keiser College

  2. A Dose calibrator operation is based on the collection of ions formed as the ionizing radiation passes through the detecting medium (gas). Detecting medium: He, Ar

  3. Basic Design of the Gaseous Detector + + +++ + - - - -- -- Gas - If the voltage across the detector electrodes is sufficiently high, ions formed will be collected.

  4. A dose calibrator is used to measure samples of radioactivity • Four quality control procedures required for the dose calibrator • Constancy • Linearity • Accuracy • Geometry

  5. Constancy • Each day • Long lived radionuclide (Cs 137) 4 1 3 5 7 9 11 13

  6. Linearity • At installation • Quarterly • Two methods of measurements • Tc99m vial • Lead attenuation leads

  7. Accuracy • Installation and annually • Two sealed reference standards • At least 50uCi ( 200uCi) • At least one with photon energy of 100keV – 500keV

  8. Geometry • At installation or whenever a change is made • Performed by increasing volume • 30ml vial – 1mCi of Tc99m in 1ml • 1,4,8,10,15,25

  9. Dose Calibrators • Records kept for 3 years including: • model, serial number, date of calibration, results and name of performer • Assurance that activity indicated is close to the truth is fundamental in the administration of radiopharmaceuticals

More Related